Agilent-technologies 8753ET Manual de usuario

Busca en linea o descarga Manual de usuario para Hardware Agilent-technologies 8753ET. Agilent Technologies 8753ET User Manual Manual de usuario

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 28
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 0
This document describes the performance
and features of the following products:
Agilent 8753ES S-parameter vector network
analyzer
Agilent 8753ES Option 011 vector network
analyzer without test set
Agilent 8753ET Transmission/reflection vector
network analyzer
For more information about these analyzers,
please see the following documents:
Pub. Number
Agilent 8753ET and 8753ES 5968-5159E
Network Analyzers
Overview
Agilent 8753ET and 8753ES 5968-5158E
Network Analyzers
Configuration Guide
Agilent 8753ET/ES
8753ET and 8753ES Network
Analyzers, 30 kHz to 3 or 6 GHz
Data Sheet
Vista de pagina 0
1 2 3 4 5 6 ... 27 28

Indice de contenidos

Pagina 1 - 8753ET and 8753ES Network

This document describes the performance and features of the following products:• Agilent 8753ES S-parameter vector network analyzer• Agilent 8753ES Op

Pagina 2 - Table of Contents

10Test-Port Output Characteristics7Frequency characteristicsRange 30 kHz to 3 GHz Option 006 30 kHz to 6 GHzResolution 1 HzStability9±7.5 ppm (0° to 5

Pagina 3 - Measurement Uncertainty

11External source mode1(CW time sweep only) Frequency range 300 kHz to 6 GHz R-channel input requirements (typical)Power level 0 to -25 dBmSpectral

Pagina 4 - System Performance Summaries

12Test Port Output CharacteristicsFrequency characteristicsRange 300 kHz to 3 GHz Option 006 30 kHz to 6 GHz Resolution 1 Hz Stability typically ±7.

Pagina 5 - 2. Typical performance

13Frequency offset mode3Frequency range 300 kHz to 3 GHzOption 006 300 kHz to 6 GHz R-channel input requirements Power level300 kHz to 3 GHz 0 to -35

Pagina 6

14Test-Port Output CharacteristicsFrequency characteristicsRange 300 kHz to 3 GHzOption 006 300 kHz to 6 GHzResolution 1 HzStability4±7.5 ppm (0° to

Pagina 7

15External source mode1(CW time sweep only)Frequency range 300 kHz to 6 GHz R-channel input requirements2Power level 0 to -25 dBmSpectral purity Ma

Pagina 8

MeasurementNumber of display channelsFour display channels available.Number of measurement channels Two primary and two auxiliary measurementchannels

Pagina 9 - 3. 2 dB less with Option 004

17Sweep triggerSet to either continuous, single, group sweep, orexternal trigger. Set external trigger to take a com-plete sweep or to measure individ

Pagina 10 - Agilent 8753ES Specifications

18Trace FunctionsDisplay data Display current measurement data, memory data,or current measurement with measurement andmemory data simultaneously. Tr

Pagina 11 - Phase Characteristics

19Frequency offset on/offSets the RF source to be swept at a fixed offset frequency above the receiver as required in a sweptRF/IF, fixed LO, mixer te

Pagina 12

341021212224Table of ContentsDefinitions and Test ConditionsSystem Performance SummariesSpecifications and CharacteristicsMeasurement Throughput Summa

Pagina 13

20Front Panel Connectors 8753ES test ports (without Option 011) Connector type 7 mm, precision Impedance 50 ohms (nominal) 8753ES Option 011 test po

Pagina 14 - Agilent 8753ET Specifications

21General ConditionsRFI and EMI susceptibility: defined by VDE 0730,CISPR Publication 11, and FCC Class B Standards.ESD (electrostatic discharge): mus

Pagina 15 - Magnitude characteristics

22Agilent 85046A/B S-Parameter Test SetsThe 85046A/B S-parameter test sets provide thecapability to measure reflection and transmissioncharacteristics

Pagina 16 - Group Delay Characteristics

23Agilent 85047A S-Parameter Test SetThe 85047A S-parameter test set provides thecapability to simultaneously measure the reflec-tion and transmission

Pagina 17 - Display Control

24Calibration KitsVector accuracy enhancement procedures requirethat the systematic errors of the measurement sys-tem be characterized by measuring kn

Pagina 18 - System Capabilities

25Agilent 85038 Family of Calibration KitsThe 85038A calibration kit contains open circuits,short circuits, and terminations with both maleand female

Pagina 19 - Remote Programming

26Verification KitsMeasuring known devices other than the standardsused in calibration is an easy way to verify the proper operation of an 8753ES meas

Pagina 20 - General Characteristics

27Connectors 50-ohm type-N (f) to75-ohm type-N (m)standard Option 004, 50-ohm type-N (m) to 75-ohm type-N (f)Dimensions 14-mm D x 70-mm L(0.56 in x 2

Pagina 21 - General Conditions

Agilent 8347A RF AmplifierThe 8347A RF amplifier delivers increased power across a 300 kHz to 3 GHz frequency range. The 8347A provides leveled output

Pagina 22

3Specifications describe the instrument’s warrantedperformance after a half-hour warm-up and overthe temperature range of 25° ± 5° C, unless other-wis

Pagina 23 - Specifications

4Agilent 8753ES (50-ohm systems) 7-mm test portsConfigurationNetwork analyzer 8753ES Standard and Option 006Calibration kit 85031BTest-port cables 118

Pagina 24 - Accessories

5Agilent 8753ES (50-ohm systems)Type-N test portsConfigurationNetwork analyzer 8753ESStandard and Option 006Calibration kit 85032BTest-port cables

Pagina 25

6Agilent 8753ES (50-ohm systems)3.5-mm test portsConfigurationNetwork analyzer 8753ESStandard and Option 006Calibration kit 85033DTest-port cables 118

Pagina 26

Agilent 8753ES (75-Ohm Systems)Type-N test portsConfigurationNetwork analyzer 8753ES Option 075Calibration kit 85036BTest-port cables 11857BSystem Dyn

Pagina 27 - 85024A High-Frequency Probe

8Agilent 8753ES (75-ohm systems)Type-F test portsConfigurationNetwork analyzer 8753E Option 075Calibration kit 85039BTest-port cables 11857BSystem dyn

Pagina 28

91. One-port or enhanced-response calibration.2. Enhanced-response or response-only calibration.3. 2 dB less with Option 004.4. Option 004 may degrade