6 Oscilloscope Utility Settings
114 Agilent 3000 Series Oscilloscopes User’s Guide
To set up masks
You can create masks by adding horizontal and vertical margins to a
signal. You can save and load masks from internal memory.
To adjust a mask’s horizontal failure margin
1 In the Mask menu (UtilityMask Test), select XMask.
2 Turn the Entry
knob to adjust the horizontal failure margin.
The margin can be set from 0.04 div to 4.00 div.
To adjust a mask’s vertical failure margin
1 In the Mask menu (UtilityMask Test), select YMask.
2 Turn the Entry
knob to adjust the vertical failure margin.
The margin can be set from 0.04 div to 4.00 div.
To create a mask using the failure margin settings
1 In the Mask menu (UtilityMask Test), select Create Mask.
To save a mask
1 In the Mask menu (UtilityMask Test), select Save.
To load a mask
1 In the Mask menu (UtilityMask Test), select Load.