Chapter 3: Testing Performance
To test delta time measurement accuracy
62
Equipment Required
Connections
Equipment Critical Specifications Recommended Model/Part
Sine wave source Output Frequency: 50 MHz to 7 GHz
Output Amplitude: -10 dBm to 10 dBm
Absolute Frequency Error: < +-1 ppm
Period Jitter: < 0.1 times the published typical
period jitter of the oscilloscope under test
when used with the specified band-pass
filter.
Agilent E8257 PSG
Microwave cable
assembly
50 ohm characteristic impedance (no
substitute)
3.5 mm (m) connectors
Max Frequency: >= 18 GHz
Agilent 8120-4948
Adapters, assorted 3.5 mm (f) to Precision BNC (m)
3.5 mm (f) to 3.5 mm (f)
SMA (f) to BNC (m)
Agilent 54855-67604
Agilent 83059B
Amphenol 901-165
500 MHz Band-pass
Filter
(for 8104 only)
Center Freq: 500 MHz
Pass 3dB BW: 50 MHz
Insertion Loss.: 2.1 dB
Attn @ 400 MHz: 35 dB
Attn @ 600 MHz: 35 dB
K&L Microwave
4LB40-500/T50-O/OP
300 MHz Band-pass
Filter
(for 8064 only)
Center Freq: 300 MHz
Pass 3dB BW: 20 MHz
Insertion Loss.: 3.3 dB
Attn @ 260 MHz: 36 dB
Attn @ 340 MHz: 36 dB
K&L Microwave
4LB40-300/T20-O/OP
Oscilloscope
Sine Wave Source
Filter
Adapter
(optional)
Cable
Adapter
(optional)
Adapter
(optional)