Disassembly
A2 Disassembly
A2A80A1 and A2A80A2 Filter Assembly Removal
To remove either of the filter option assemblies on the A2A80 Audio
Analyzer 1 assembly:
1. Remove the A2A80 assembly from the Test Set.
2. Turn the plastic hex nut on each standoff a quarter turn and push
the standoffs through their holes to release the filter board from the
host board, see Figure 5-10. It may be necessary to compress or
squeeze the expandable flanges to slide them through the mounting
holes.
Figure 5-10 A2A80 Audio Analyzer 1 Assembly