Performance Tests Test Records
168 Agilent 81600B Tunable Laser Source Family, Fourth Edition
Agilent 81600B #160 Performance Test Page 6 of 14
Model Agilent 81600B #160 Tunable Laser Report No. ________ Date_______
Wavelength Repeatability
Measurement Uncertainty: ± 0.1 pm
Repeatability of
1495.000 nm (= reference)
Measurement Result
Repeatability of
1580.000 nm (= reference)
Measurement Result
Initial Setting REF = nm Initial Setting REF = nm
from 1520.000 nm to REF nm from 1495.000 nm to REF nm
from 1550.000 nm to REF nm from 1520.000 nm to REF nm
from 1580.000 nm to REF nm from 1550.000 nm to REF nm
from 1610.000 nm to REF nm from 1610.000 nm to REF nm
from 1640.000 nm to REF nm from 1640.000 nm to REF nm
largest measured wavelength nm largest measured wavelength nm
smallest measured
wavelength
nm smallest measured
wavelength
nm
Wavelength Repeatability nm Wavelength Repeatability nm
= largest measured wavelength - smallest measured
wavelength
= largest measured wavelength - smallest measured
wavelength
Upper Test Limit 0.0016 nm Upper Test Limit 0.0016 nm
Performance Characteristic 0.0010 nm typical Performance Characteristic 0.0010 nm typical
Repeatability of
1640.000 nm (= reference)
Measurement
Result
Initial Setting REF =
nm
from 1495.000 nm to REF nm
from 1520.000 nm to REF nm
from 1550.000 nm to REF nm
from 1580.000 nm to REF nm
from 1610.000 nm to REF nm
from 1615.000 nm to REF nm
largest measured wavelength nm
smallest measured
wavelength
nm
Wavelength Repeatability nm
= largest measured wavelength - smallest measured
wavelength
Upper Test Limit 0.0016 nm
Performance Characteristic 0.0010 nm typical