Following is a description of the Select Start Mode screen options:
APP 3/7 Options Screen
Following is a description of the APP 3/7 Options screen options:
Table 2 3 Select Start Mode Options (Apparatus 3 / 7)
Option Function
1 Start Now Select option 1, Start Now. The APP 3/7 Options screen displays.
2 Delayed Start Select option 2, Delayed Start. The APP 3/7 Options screen displays.
3 Remote Start This option is designed for use with the 8000 coupled with non-Agilent
instruments. Contact the Dissolution Systems Service Department for information
about this function.
4 Modify Program Select option 4, Modify Program, to modify or review the selected program.
“Creating a Sample Program” on page 61 to review programming options. All
program variable items can be viewed or changed.
Press ESC to return to the Select Start Mode screen.
Selecting option 4 again repeats the modify program operation.
When the program is verified or changed to the correct settings, select the desired
start mode (option 1 - 3) to start the program.
***APP 3/7 OPTIONS***
1 LOAD & RUN APP 3/7 2 RUN WO/ LOADING
3 RUN 8000 ONLY 4 SET APP 3/7
TIME: 01/01/12 08:09:51 ELAPSE: 000:00:00