Agilent Technologies DisplayPort Manual de servicio Pagina 218

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 246
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 217
218 Agilent U7232A DisplayPort Electrical Performance Compliance Test Application
24 Calibrating the Infiniium Oscilloscope and Probe
Figure 83 Time Scale Calibration Dialog box
d Click on the Std + Dflt button to continue the calibration, using the
Factory default calibration factors.
e When the calibration procedure is complete, you will be prompted
with a Calibration Complete message window. Click the OK button to
close this window.
f Confirm that the Vertical and Trigger Calibration Status for all
Channels passed.
g Click the Close button to close the calibration window.
h The internal calibration is completed.
i Read NOTE below.
NOTE
These steps do not need to be performed every time a test is run. However, if the ambient
temperature changes more than 5 degrees Celsius from the calibration temperature, this
calibration should be performed again. The delta between the calibration temperature and
the present operating temperature is shown in the Utilities>Calibration menu.
Vista de pagina 217
1 2 ... 213 214 215 216 217 218 219 220 221 222 223 ... 245 246

Comentarios a estos manuales

Sin comentarios