4-52 Chapter4
Troubleshooting
Greater than 1000: Personality Specific Error Messages
10172 Sweep Time too fast(<2sec)
The sweep time must be set to 2 seconds or longer for
the results to be valid.
10141 Sync word not found in frame (Burst Type)
One or more active GSM bursts that match the selected
Burst Type have been detected in the RF Input signal,
but none contain the selected Training Sequence Code
(TSC). The search was performed over the complete
GSM frame.
10143 Sync word not found in frame (Ref Burst)
One or more active GSM bursts that match the selected
Burst Type have been detected in the RF Input signal,
but none contain the selected Training Sequence Code
(TSC). The search was only performed using the
Reference Burst type and Reference TSC settings over
the complete GSM frame.
10142 Sync word not found in specified timeslot
(Burst Type)
One or more active GSM bursts that match the selected
Burst Type have been detected in the RF Input signal,
but none contain the selected Training Sequence Code
(TSC). The search was only performed over the
specified timeslot setting.
10259 Table could not be loaded.
When trying to load a table, the previous table has been
somehow corrupted. Use the
Save Table key to save a
valid table. Then edit the valid table, save it, and try to
load it again.
10259 Table could not be saved.
This message occurs if C: drive is full or corrupt. Check
the amount of space left on the drive.