Vista de pagina 14
Agilent B2900A Configuration and Connection Guide, Edition 1 2- 7
B2900A Accessories
Adapters & Connectors
CAUTION Maximum voltage and current
The test fixture and adapters must be used under the following limitations to prevent damage to them.
Agilent N1294A-001 and 002 Banana to Triax adapter:
±250 V maximum, ±42 V maximum for connecting N1295A Small test fixture.
Agilent N1295A Small Test Fixture: ±42 V, 1.05 A maximum
Figure 2-3 Connection Diagram of N1294A-001 Banana to Triaxial adapter for 2 wire
(non-kelvin) connection
Figure 2-4 Connection Diagram of N1294A-002 Banana to Triaxial adapter for 4 wire (kelvin)
connection