Agilent-technologies E4400-60597 Manual de usuario Pagina 3

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 18
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 2
Agilent Technologies Firmware Upgrade Kits E4400-60230 and E4400-60597 for PSG/ESG Signal Generators
Check Signal Generator Functionality
3
Firmware Upgrade Kits E4400-60230 and E4400-60597 for PSG/ESG Signal
Generators
Introduction
The instructions in this firmware upgrade guide enable you to install the PSG/ESG Upgrade Assistant 5.12 on your
computer from either the Agilent Technologies website or a CD-ROM and upgrade firmware on ESG-B signal generators
with <B.03.20.
NOTE For Agilent E44xxB ESG Signal Generators Models Only
When upgrading Agilent E44xxB ESG signal generators from firmware versions <B.03.20, see, “ESG-B
Firmware Installation For Versions <B.03.20” on page 14.
To upgrade firmware on PSG/ESG signal generators, perform the following procedures:
1. Check signal generator functionality on page 5.
2. Confirm required equipment on page 6.
3. Connect an RS-232, GPIB, or LAN interface to a PC on page 8.
4. Install PSG/ESG Upgrade Assistant 5.12 and firmware files on page 11.
5. Use the PSG/ESG Upgrade Assistant 5.12 with firmware files on page 12.
6. Check signal generator functionality on page 5.
Product Affected: . . . . . . . . . . . . . . . . . . . . . . . . . All
Serial Numbers: . . . . . . . . . . . . . . . . . . . . . . . . . . All
Options: . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . All except Option H98
To Be Performed By: . . . . . . . . . . . . . . . . . . . . . . (X) Agilent Technologies Service Center
(X) Personnel Qualified by Agilent Technologies
(X) Customer
Estimated Installation Time: . . . . . . . . . . . . . . . 75 minutes (RS-232), 30 Minutes (GPIB), 3 minutes
(LAN)
Estimated Verification Time: . . . . . . . . . . . . . . . 1 minute