Agilent-technologies Multiple Channel Controller 85330A Manual de usuario Pagina 26

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 166
  • Tabla de contenidos
  • SOLUCIÓN DE PROBLEMAS
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 25
1-8 85330A Multiple Channel Controller
Installation
Selecting Positive or Negative-Edge Event Triggers
Viewing or Changing
the GPIB Address of
the Multiple Channel
Controller
The factory default setting of the GPIB address is 9. To view the GPIB
address:
Press
[Select Instr] {SYSTEM} {GPIB} {READ}.
To change the current GPIB address:
1. Press
[Select Instr] {SYSTEM} {GPIB} {SET}.
2. Enter the new address and press
[Return].
Selecting Positive or Negative-Edge Event
Triggers
By default, the positive-edge of the EVENT TRIG line starts the
measurement. To change this so a negative-edge starts the measurement,
perform the following procedure.
CAUTION This device contains devices that are sensitive to static discharge. When you
remove the cover of this device, observe static safety precautions:
1. Place the unit on a grounded anti-static mat.
2. Wear a grounded wrist strap. Foot straps are only acceptable if you wear
one on each foot, and if you are standing on a grounded floor mat.
Refer to Figure 1-4 on page 1-9.
Vista de pagina 25
1 2 ... 21 22 23 24 25 26 27 28 29 30 31 ... 165 166

Comentarios a estos manuales

Sin comentarios