5
High Frequency Active Differential Probe
Agilent 1130A InfiniiMax High-Performance Active Probe System
• 1.5 GHz InfiniiMax probe system
• InfiniiMax probe amplifier supports
both differential- and single-ended
measurements for a more
cost-effective solution
• Unrivaled InfiniiMax probing
accessories support browsing,
solder-in, and socketed use models
at the maximum performance
available
• Compatible with 6000 Series
oscilloscopes (300 MHz -
1 GHz only)
The 1.5 GHz 1130A InfiniiMax
probe amplifier is a perfect com-
plement to the 6000 Series 1 GHz
models. Its 1.5 GHz bandwidth,
extremely low input capacitance
(0.32 pF), high common mode
rejection and the patented resis-
tor probe tip technology provide
ultra low loading of the DUT and
superior signal fidelity. Agilent’s
innovative InfiniiMax 1130A
differential probe is the easiest-
to-use, and highest performance
probing system available for high-
speed digital design, and repre-
sent a new industry standard for
accuracy, flexibility and reliability.
Designers can achieve 1 GHz sys-
tem bandwidth in conjunction
with 1 GHz 6000 Series oscillo-
scopes even when manually
browsing with the probe or mak-
ing hands-off measurements.
Optional solder-in probe heads
and solder-in sockets as well as
browser configuration provide
full bandwidth at the probe tip.
Specifications
Operating Characteristics
Probe Bandwidth (–3dB) > 1.5 GHz
Rise and Fall Time (10% to 90%) 233 psec
System Bandwidth (–3dB) 1130A with MSO/DSO610xA: 1 GHz
Input Capacitance Cm = 0.1 pF Cm is between tips
Cg = 0.34 pF Cg is ground for each tip
Cdiff = 0.27 pF Differential mode capacitance = Cm + Cg/2
Cse = 0.44 pF Single-ended mode capacitance = Cm + Cg
Input Resistance Differential mode resistance = 50 kΩ ±1%
Single-ended mode resistance = 25 kΩ ±1%
Input Dynamic Range ±2.5 V
Input Common Mode Range ±6.75 V dc to 100 Hz; ±1.25 V > 100 Hz
Maximum Signal Slew Rate 18 V/ns when probing a single-ended signal
30 V/ns when probing a differential signal
DC Attenuation 10:1 ± 3% before calibration on oscilloscope
10:1 ± 1% after calibration on oscilloscope
Offset Range ± 12.0 V when probing single-ended
Maximum Input Voltage 30 Vpeak, CAT I
ESD Tolerance >8 kV from 100 pF, 300 Ω HBM
Maximum Number of Probes 2
Supported by 6000 Series
Agilent 1130A InfiniiMax probe offers you the highest performance available for
measuring differential and single-ended signals.