MegaZoom Concepts and Oscilloscope Operation
To recall traces and setups
To recall traces and setups
1Press the Save/Recall key to display the Save/Recall menu.
2Press the
Recall softkey to display the Recall menu.
3Press the
Recall softkey and select the type of information to recall.
You can recall a waveform Trace, an oscilloscope Setup, or Trace and Setup. Be
sure to recall both setup and trace if you want to measure the recalled trace
with cursors.
4 Select the file name to recall by turning the Entry knob while viewing
the
From: softkey.
INTERN_n are internal non-volatile oscilloscope memory files. All other files in
the list are stored on the floppy disk.
5 Recall the selected file by pressing the Press to Recall softkey.
The recalled trace will be displayed in minimum intensity.
6 To clear the display of any recalled trace, press the Clear Display softkey.
Recalling overwrites current settings
Because recalling a setup will overwrite the oscilloscopes current settings, you may
want to save the existing setup first.