53210A 53220A 53230A
Optional microwave inputs (nom)
Frequency range
Option 106 100 MHz - 6 GHz
Option 115 300 MHz - 15 GHz
Input
Connector Front panel precision Type-N(f)
Option 203 moves the input connector to a rear panel SMA(f)
Input impedance (typ) 50 Ω ± 1.5% (SWR < 2.5)
Input coupling AC
Continuous wave
amplitude range
Option 106 Autoranged to +19 dBm max. (2 Vrms)
Option 115 Autoranged to +13 dBm max. (1.0 Vrms)
Sensitivity (typ)
5
6 GHz (Opt 106): -27 dBm (10 mVrms)
15 GHz (Opt 115):
< 3 GHz: -23 dBm
3 – 11 GHz: -27 dBm
> 11 GHz: -21 dBm
Input event thresholds
Level range Auto-ranged for optimum sensitivity and bandwidth
AM tolerance
6
50% modulation depth
Maximum input
Damage level 6 GHz (Opt 106): > +27 dBm (5 Vrms)
15 GHz (Opt 115): > +19 dBm (2 Vrms)
Input Channel Characteristics continued
1. AC coupling occurs after 50 Ω termination.
2. When ordered with optional rear terminals, the standard/baseband channel inputs are active on both the front and rear of the
universal counter though the specifications provided only apply to the rear terminals. Performance for the front terminals with rear
terminals installed is not specified.
3. Multiply value(s) by 10 for the 50 V range.
4. Stated specification assumes Noise Reject OFF. Noise Reject ON doubles the sensitivity minimum voltage levels.
5. Assumes sine wave.
6. CW only. Assumes AM Rate > 10/gate. For Option 106, spec applies for input powers > -20 dBm; use a tolerance of 15% modulation
depth for frequencies less than 900 MHz. For Option 115, spec applies for input powers > -10 dBm.