Agilent 87050E
50 Ohm Multiport Test Sets
Data Sheet
This document describes the performance and fea-
tures of Agilent Technologies 87050E 50 ohm multi-
port test sets, both as standalone units and when
combined with 50 ohm 8712ET, 8712ES, 8714ET, or
8714ES network analyzers. The following options
are available:
• Option 004 (4 ports)
• Option 008 (8 ports)
• Option 012 (12 ports)
For more information about these test sets, please
read the following documents:
Agilent 87050E, brochure,
literature number, 5968-4763E
Agilent 87050E, configuration guide,
literature number, 5968-4765E
Agilent 87050E 3 MHz to 2.2 GHz
Vista de pagina 0
1 2 3 4 5 6 ... 19 20

Indice de contenidos

Pagina 1 - 50 Ohm Multiport Test Sets

Agilent 87050E50 Ohm Multiport Test SetsData Sheet This document describes the performance and fea-tures of Agilent Technologies 87050E 50 ohm multi-p

Pagina 2 - Introduction

10 Specification Supplemental information System bandwidth Network analyzer Fine Fine Med Wide Wide 15 Hz 15 Hz 4000 Hz 6500 Hz Receiver dy

Pagina 3 - Two-port calibration (user)

11Network analyzer Specification Characteristic (dBm) Output power (system maximum) <1 GHz >1 GHz8712ET, no attenuator — 10.5 78712E

Pagina 4

12Test set input/output performanceParameter Specification Supplemental Frequency range 3 MHz to 2.2 GHz — RF input powerMaximum input power —

Pagina 5

13Description Specification Supplemental information Front panel ports87050E, Option 004 4 Type-N, female 50Ω (nominal)87050E, Option 008 8

Pagina 6

14Physical dimensionsPhysical dimensions of the Agilent 87050E Option 004multiport test set1Physical dimensions of the Agilent 87050E Option 008 multi

Pagina 7 - Phase, ES models

15Block diagramsBlock diagram for the Agilent 87050E (only one test set port pair is shown)REAR PANELExternal DetectorsInput B*Input BYAUX InputInput

Pagina 8 - Uncorrected

16Test set controlControl of the switches inside the test set and cali-bration of the test system can be accomplished from the front panel of the netw

Pagina 9

17Disk driveTrace data, instrument states (including calibra-tion data), and IBASIC programs can be saved on floppy disks using the built-in 3.5 inch

Pagina 10 - General (continued)

18NFSThe analyzer’s built-in NFS (network file system) client provides access to remote files and directo-ries using the LAN. With NFS, remote files a

Pagina 11

19Other calibrations Besides using a Test Set Cal, individual instrument states and their corresponding calibrations can be saved and recalled for use

Pagina 12

2All specifications and characteristics apply over a 25 °C ± 5 °C range (unless otherwise stated) and 30 minutes after the instrument has been turned

Pagina 13 - Test set general information

Java is a U.S. trademark of Sun Microsystems, Inc.Microsoft, MS-DOS, Windows, Windows 95, Windows 98, Windows NT, Word 97, and Excel 97 are U.S. reg

Pagina 14 - Physical dimensions

3Agilent 85032B/E Type-N calibration kit Specification1 (dB) Typical2 (dB) Parameter 3 MHz 1.3 GHz 2.2 GHz to 1.3 GHz to 2.2 GHz to 3 GHzDir

Pagina 15 - Block diagrams

4System performance Two-port calibration (user) Specification1 (dB) Typical2 (dB) Parameter 3 MHz 1.3 GHz 2.2 GHz to 1.3 GHz to 2.2 GHz to 3

Pagina 16 - System features

5Agilent 85032B/E Type-N calibration kit1. These specifications and supplementals are valid for reflection (one-port) and transmission (enhanced-resp

Pagina 17 - Data hard copy

6System performance Transmission/reflection (T/R) calibration (user)1. These uncertainty curves only include the effects of the test port(s) within t

Pagina 18 - System features (continued)

7Agilent 85032B/E Type-N calibration kit 1. These uncertainty curves only include the effects of the test port(s) within the measurement path. The ef

Pagina 19 - Key network analyzer options

8Type-N Specification1 ( dB) Typical (dB)Parameter 3 MHz 1.3 GHz 3 MHz 1.3 GHz 2.2 GHz to 1.3 GHz to 2.2 GHz to 1.3 GHz to 2.2 GHz

Pagina 20 - Remove all doubt

9 Specification Typical System bandwidth Fine Fine Med Wide Wide 15 Hz 15 Hz 4000 Hz 6500 HzNetwork analyzer 3 MHz 1.3 GHz