Sequence: 1 to 200 segments (optional
8000)
Dead time:
-001: < 1.1 μs
-002: < 800 ns
Trigger time interpolator
5 ps resolution
Internal and external trigger
Internal trigger input
Threshold adjust range: same as
vertical FSR
Sensitivity: > 15% FSR
Frequency range:
-001: DC to 500 MHz
-002: DC to 1 GHz
External trigger input
Threshold adjust range: (-FS/2, FS/2)
for FS = 500 mV, 1 V, 2 V, and 5 V
Impedance: 50 Ω
Maximum input voltage: ±5 V DC
Sensitivity: > 15% FS
Frequency range: DC to 1 GHz
Coupling
DC, AC (50 Hz LF reject), 50 kHz HF
reject
Modes
Edge, positive and negative
Window
HF divide by 4
Pre-trigger
Adjustable to 100% of horizontal full
scale
Post-trigger
Adjustable up to 200 MSamples
Control I/O (MMCX)
Ctrl I/O A and B signals
TTL & CMOS compatible (3.3 V)
Ctrl I/O A and B output
10 MHz reference clock out with 50 Ω
impedance
Acquisition active
Acquisition skipping to next segment
Trigger ready
Ctrl I/O A and B input
Trigger enable
Trigger OUT
Offset: ± 2.5 V (no load)
Amplitude ± 0.8 V (no load), ±15 mA max
Rise/fall time: 2.5 ns into 50 Ω
Coupling: DC
Output impedance: 50 Ω
CLK IN ext. clock/ref
Amplitude: > 500 mV pk-pk into 50 Ω
Threshold: variable between -2 V and +2 V
Maximum input voltage: ±2 V DC
CLK IN ext. clock input
10 MHz to 2 GHz
SR may be refi ned with sparsing
CLK IN ext. reference frequency
9-10.2 MHz
General
Host computer and operating system:
PC compatible (x86) systems running
Microsoft Windows XP, Windows
2003 Server, Windows 2000, Wind
River VxWorks, National Instruments
LabVIEW RT, or Linux.
PowerPC systems running Wind River
VxWorks.
For more information on which spe-
cifi c processors and operating system
versions are supported, please contact
us.
Transfer speed:
High-speed PCI bus transfers data
at sustained rates to host computer:
Up to 100 Mbytes/s for 32-bit/33 MHz
operation
Power consumption
With standard memory
-001: <13 W
-002: <14 W
With maximum memory option
-001: <14 W
-002: <16 W
Current requirements
U1061A-001:
12 V 0.46 A
5 V 0.46 A
3.3 V 1.1 A
-12 V 40 mA
With maximum memory option
3.3 V 1.5 A
U1061A-002:
12 V 0.46 A
5 V 0.46 A
3.3 V 1.4 A
-12 V 40 mA
With maximum memory option
3.3 V 2.2 A
Warranty
1 year
Environmental and Physical
Operating temperature
0° to 40°C
Relative humidity
5 to 95% (non-condensing)
Dimensions
3U PXI/CompactPCI® standard
100 mm x 160 mm x 20 mm
Safety
Complies with EN61010-1
EMC immunity
Complies with EN61326-1
Industrial Environment
EMC emissions
Complies with EN61326-1 Class A for
radiated emissions
Required airfl ow
> 2 m/s in situ
7
Comentarios a estos manuales