4
Specifi cations are for 90 minutes warm- up time, slow
mode, and calibration temperature within 18 °C - 28 °C
DC Specifications
Table 1. DC Accuracy ± (% of reading + % of range)
Function Range
1
Test Current or
Burden Voltage
Input
Impedance
1 Year
23 °C ± 5 °C
Temperature Coefficient
0 °C - 18 °C
28 °C - 55 °C
DC Voltage 100.000 mV — 10 MΩ or > 10 GΩ 0.018 + 0.008 0.0020 + 0.0008
1.00000 V — 10 MΩ or > 10 GΩ 0.015 + 0.005 0.0015 + 0.0008
10.0000 V — 10 MΩ 0.015 + 0.005 0.0020 + 0.0008
100.000 V — 10 MΩ 0.015 + 0.005 0.0020 + 0.0008
1000.00 V — 10 MΩ 0.015 + 0.005 0.0020 + 0.0008
Resistance
(2-wire)
2
Resistance
(4-wire)
2
100.000 Ω 1 mA — 0.050 + 0.008 0.0060 + 0.0008
1.00000 kΩ 500 μA — 0.050 + 0.008 0.0060 + 0.0005
10.0000 kΩ 100 μA — 0.050 + 0.005 0.0060 + 0.0005
100.000 kΩ 10 μA — 0.050 + 0.005 0.0060 + 0.0005
1.00000 MΩ 1 μA — 0.060 + 0.005 0.0060 + 0.0005
10.0000 MΩ 100 nA — 0.250 + 0.005 0.0250 + 0.0005
100.000 MΩ 100 nA/10 MΩ — 2.000 + 0.005 0.3000 + 0.0005
DC Current 100.000 μA < 0.02 V — 0.05 + 0.015 0.007 + 0.0015
1.00000 mA < 0.2 V — 0.05 + 0.007 0.007 + 0.0010
10.0000mA < 0.02 V — 0.05 + 0.015 0.008 + 0.0015
100.000mA < 0.2 V — 0.05 + 0.007 0.008 + 0.0010
1.00000A < 0.2 V — 0.10 + 0.015 0.012 + 0.0015
10.0000A < 0.6 V — 0.25 + 0.007 0.015 + 0.0010
Continuity
3
1000 Ω 0.5 mA — 0.05 + 0.03 0.005 + 0.005
Diode Test
4
1.0000V 0.5 mA — 0.05 + 0.03 0.005 + 0.005
Notes on Specifications:
1. 20% over range on all ranges except 1000 VDC and 10 A range.
2. Specifications are for 4-wire Ω or 2-wire Ω using NULL function. If without NULL function, add 0.2 Ω additional error.
3. Continuity thresholds is fixed at less than 10 Ω
4. Specifications are for the voltage measured at the input terminals only.