Ch 93 HI Measurement 1. Make Hardware Connections
Turn mainframe power OFF
Connect DMM leads as shown in Figure 2-6
Turn mainframe power ON
2. Measure Channel 93 HI Resistance
Send *RST to multiplexer
Send CLOS (@nn93) to close chan 93, where nn = card # (omit
leading zeroes in nn )
Trigger the DMM with TRIG SGL and note reading
Send OPEN (@nn93) to open channel 93
Subtract measured value of (R38 + R43) from DMM reading
Enter the result in Table 2-2 for Channel 93 HI
Figure 2-6. Ch 93 HI Measurement Connections
2-8 Verification Tests