Calibration Guide
Agilent Technologies ESA Spectrum Analyzers
This manual provides documentation for the following instruments:
Agilent ESA-E Series
E4402B (9 kHz - 3.0 GHz)
E4404B (9 kHz - 6.7 GHz)
E4405B (9 kHz - 13.2 GHz)
E4407B (9 kHz - 26.5 GHz)
and
Agilent ESA-L Series
E4411B (9 kHz - 1.5 GHz)
E4403B (9 kHz - 3.0 GHz)
E4408B (9 kHz - 26.5 GHz)
Manufacturing Part Number: E4401-90493
Supersedes E4401-90475
Printed in USA
March 2005
© Copyright 2000-2005 Agilent Technologies, Inc.
Vista de pagina 0
1 2 3 4 5 6 ... 641 642

Indice de contenidos

Pagina 1 - Calibration Guide

Calibration GuideAgilent Technologies ESA Spectrum Analyzers This manual provides documentation for the following instruments:Agilent ESA-E SeriesE44

Pagina 2

10 Chapter 1 Calibrating Calibration verifies that the analyzer performance is within all specifications. It is time consuming and requires extensi

Pagina 3

100 Chapter 2 Performance Verification Tests13. Input Attenuation Switching Uncertainty13. Input Attenuation Switching UncertaintyA 50 MHz CW signal i

Pagina 4

Chapter 2 101Performance Verification Tests13. Input Attenuation Switching UncertaintyFigure 2-13 Input Attenuator Switching Uncertainty Test Setup C

Pagina 5 - Contents

102 Chapter 2 Performance Verification Tests13. Input Attenuation Switching Uncertaintyusing the Agilent 8496G programmable attenuator, enter the cali

Pagina 6

Chapter 2 103Performance Verification Tests13. Input Attenuation Switching Uncertainty4. For each attenuation error value in Column 4 of Table 2-19, c

Pagina 7

104 Chapter 2 Performance Verification Tests13. Input Attenuation Switching Uncertainty7. Preset the synthesized signal generator (Blue Key, Special,

Pagina 8

Chapter 2 105Performance Verification Tests13. Input Attenuation Switching UncertaintyMeasure Switching UncertaintyPerform step 12 to step 15 for each

Pagina 9 - 1Calibrating

106 Chapter 2 Performance Verification Tests13. Input Attenuation Switching UncertaintyExample for 25 dB analyzer internal attenuation setting:Post-te

Pagina 10 - Calibrating

Chapter 2 107Performance Verification Tests14. Reference Level Accuracy: Agilent E4401B and E4411B14. Reference Level Accuracy: Agilent E4401B and E44

Pagina 11 - Calibration Cycle

108 Chapter 2 Performance Verification Tests14. Reference Level Accuracy: Agilent E4401B and E4411BProcedure Figure 2-14 Reference Level Accuracy Tes

Pagina 12 - 12 Chapter 1

Chapter 2 109Performance Verification Tests14. Reference Level Accuracy: Agilent E4401B and E4411BExample for 50 dB attenuator setting:Log Scale, Anal

Pagina 13 - (This test has been deleted.)

Chapter 1 11 CalibratingCalibration CycleCalibration Cycle The performance tests in chapter 2 should be used to check the analyzer against its spec

Pagina 14 - 14 Chapter 1

110 Chapter 2 Performance Verification Tests14. Reference Level Accuracy: Agilent E4401B and E4411Bamplitude and enter the result as the Test Record e

Pagina 15 - Chapter 1 15

Chapter 2 111Performance Verification Tests14. Reference Level Accuracy: Agilent E4401B and E4411BLog Scale, Digital Bandwidths, Option 1DRNOTE If the

Pagina 16 - 16 Chapter 1

112 Chapter 2 Performance Verification Tests14. Reference Level Accuracy: Agilent E4401B and E4411Ba. Press Single on the analyzer.b. Press Peak Searc

Pagina 17 - Chapter 1 17

Chapter 2 113Performance Verification Tests14. Reference Level Accuracy: Agilent E4401B and E4411BLinear Scale, Analog Bandwidths1. Set the 10 dB step

Pagina 18 - 18 Chapter 1

114 Chapter 2 Performance Verification Tests14. Reference Level Accuracy: Agilent E4401B and E4411BPeak Search (or Search)Marker, Delta6. Set the 10 d

Pagina 19 - Chapter 1 19

Chapter 2 115Performance Verification Tests14. Reference Level Accuracy: Agilent E4401B and E4411BLinear Scale, Digital Bandwidths, Option 1DRNOTE If

Pagina 20 - 20 Chapter 1

116 Chapter 2 Performance Verification Tests14. Reference Level Accuracy: Agilent E4401B and E4411Ba. Press Single on the analyzer.b. Press Peak Searc

Pagina 21

Chapter 2 117Performance Verification Tests15. Reference Level Accuracy: Agilent E4402B, E4403B, E4404B, E4405B, E4407B, andE4408B.15. Reference Level

Pagina 22 - 22 Chapter 1

118 Chapter 2 Performance Verification Tests15. Reference Level Accuracy: Agilent E4402B, E4403B, E4404B, E4405B, E4407B, and E4408B.Figure 2-15 Refe

Pagina 23 - Chapter 1 23

Chapter 2 119Performance Verification Tests15. Reference Level Accuracy: Agilent E4402B, E4403B, E4404B, E4405B, E4407B, andE4408B.Example for 50 dB a

Pagina 24 - Before You Start

12 Chapter 1 Calibrating Calibration CycleTable 1-1 Agilent E4401B Performance Verification TestsPerformance Test NameCalibration for Instrument O

Pagina 25 - Test equipment you will need

120 Chapter 2 Performance Verification Tests15. Reference Level Accuracy: Agilent E4402B, E4403B, E4404B, E4405B, E4407B, and E4408B.performance test

Pagina 26 - 8663A P,A

Chapter 2 121Performance Verification Tests15. Reference Level Accuracy: Agilent E4402B, E4403B, E4404B, E4405B, E4407B, andE4408B.9. On the analyzer,

Pagina 27

122 Chapter 2 Performance Verification Tests15. Reference Level Accuracy: Agilent E4402B, E4403B, E4404B, E4405B, E4407B, and E4408B.AMPLITUDE, Scale

Pagina 28

Chapter 2 123Performance Verification Tests15. Reference Level Accuracy: Agilent E4402B, E4403B, E4404B, E4405B, E4407B, andE4408B.Linear Scale, Digit

Pagina 29

124 Chapter 2 Performance Verification Tests15. Reference Level Accuracy: Agilent E4402B, E4403B, E4404B, E4405B, E4407B, and E4408B.b. Press Peak Sea

Pagina 30 - >42 dB

Chapter 2 125Performance Verification Tests16. Resolution Bandwidth Switching Uncertainty16. Resolution Bandwidth Switching UncertaintyTo measure the

Pagina 31 - Ω Minimum Loss

126 Chapter 2 Performance Verification Tests16. Resolution Bandwidth Switching UncertaintyInput/Output (or Input), Amptd Ref (On) (Agilent E4401B, E44

Pagina 32 - ≥91 cm (36 in)

Chapter 2 127Performance Verification Tests16. Resolution Bandwidth Switching Uncertainty7. If you are testing an analyzer equipped with Option 1DR, p

Pagina 33

128 Chapter 2 Performance Verification Tests16. Resolution Bandwidth Switching Uncertainty10 Hzb100 Hz 15)3 Hzc,b100 Hz 16)1 Hzc,b100 Hz 17)a. These R

Pagina 34 - 34 Chapter 2

Chapter 2 129Performance Verification Tests17. Absolute Amplitude Accuracy (Reference Settings): Agilent E4401B and E4411B17. Absolute Amplitude Accur

Pagina 35 - Chapter 2 35

Chapter 1 13 CalibratingCalibration CycleTable 1-2 Agilent E4402B Performance Verification TestsPerformance Test NameCalibration for Instrument Op

Pagina 36 - 36 Chapter 2

130 Chapter 2 Performance Verification Tests17. Absolute Amplitude Accuracy (Reference Settings): Agilent E4401B and E4411Bequipped with Option 1DS, P

Pagina 37 - Calibration

Chapter 2 131Performance Verification Tests17. Absolute Amplitude Accuracy (Reference Settings): Agilent E4401B and E4411B7. Adjust the synthesized si

Pagina 38

132 Chapter 2 Performance Verification Tests17. Absolute Amplitude Accuracy (Reference Settings): Agilent E4401B and E4411BFREQUENCY, 50 MHzAMPLITUDE,

Pagina 39

Chapter 2 133Performance Verification Tests17. Absolute Amplitude Accuracy (Reference Settings): Agilent E4401B and E4411BBW/ Avg, Resolution BW, 1 kH

Pagina 40

134 Chapter 2 Performance Verification Tests17. Absolute Amplitude Accuracy (Reference Settings): Agilent E4401B and E4411BAbsolute Amplitude Accuracy

Pagina 41

Chapter 2 135Performance Verification Tests17. Absolute Amplitude Accuracy (Reference Settings): Agilent E4401B and E4411BDet/Demod, Detector, Sample6

Pagina 42

136 Chapter 2 Performance Verification Tests18. Absolute Amplitude Accuracy (Reference Settings): Agilent E4402B, E4403B, E4404B, E4405B, E4407B, and

Pagina 43 - Ω/1 MΩ until the LED is lit

Chapter 2 137Performance Verification Tests18. Absolute Amplitude Accuracy (Reference Settings): Agilent E4402B, E4403B, E4404B,E4405B, E4407B, and E4

Pagina 44

138 Chapter 2 Performance Verification Tests18. Absolute Amplitude Accuracy (Reference Settings): Agilent E4402B, E4403B, E4404B, E4405B, E4407B, and

Pagina 45

Chapter 2 139Performance Verification Tests18. Absolute Amplitude Accuracy (Reference Settings): Agilent E4402B, E4403B, E4404B,E4405B, E4407B, and E4

Pagina 46

14 Chapter 1 Calibrating Calibration Cycle56.(This test has been deleted.)57.(This test has been deleted.)58.GSM Phase and Frequency Errorc•59. Com

Pagina 47

140 Chapter 2 Performance Verification Tests18. Absolute Amplitude Accuracy (Reference Settings): Agilent E4402B, E4403B, E4404B, E4405B, E4407B, and

Pagina 48 - E4403B, and E4411B

Chapter 2 141Performance Verification Tests18. Absolute Amplitude Accuracy (Reference Settings): Agilent E4402B, E4403B, E4404B,E4405B, E4407B, and E4

Pagina 49

142 Chapter 2 Performance Verification Tests18. Absolute Amplitude Accuracy (Reference Settings): Agilent E4402B, E4403B, E4404B, E4405B, E4407B, and

Pagina 50

Chapter 2 143Performance Verification Tests19. Overall Absolute Amplitude Accuracy: Agilent E4401B and E4411B19. Overall Absolute Amplitude Accuracy:

Pagina 51

144 Chapter 2 Performance Verification Tests19. Overall Absolute Amplitude Accuracy: Agilent E4401B and E4411B3. Zero and calibrate the power meter an

Pagina 52

Chapter 2 145Performance Verification Tests19. Overall Absolute Amplitude Accuracy: Agilent E4401B and E4411B8. Obtain the metrology data for the step

Pagina 53

146 Chapter 2 Performance Verification Tests19. Overall Absolute Amplitude Accuracy: Agilent E4401B and E4411BFigure 2-22 Amplitude Accuracy Test Set

Pagina 54

Chapter 2 147Performance Verification Tests19. Overall Absolute Amplitude Accuracy: Agilent E4401B and E4411Btest record as indicated in Table 2-30.Me

Pagina 55

148 Chapter 2 Performance Verification Tests19. Overall Absolute Amplitude Accuracy: Agilent E4401B and E4411B3. Perform the following steps for each

Pagina 56 - Accuracy

Chapter 2 149Performance Verification Tests19. Overall Absolute Amplitude Accuracy: Agilent E4401B and E4411Bc. Press Single and wait for the sweep to

Pagina 57

Chapter 1 15 CalibratingCalibration CycleTable 1-3 Agilent E4403B Performance Verification TestsPerformance Test NameCalibration for Instrument Op

Pagina 58

150 Chapter 2 Performance Verification Tests20. Overall Absolute Amplitude Accuracy: Agilent E4402B, E4403B, E4404B, E4405B, E4407B, and E4408B20. Ove

Pagina 59 - 6. Frequency Span Accuracy:

Chapter 2 151Performance Verification Tests20. Overall Absolute Amplitude Accuracy: Agilent E4402B, E4403B, E4404B, E4405B, E4407B,and E4408B4. Zero a

Pagina 60

152 Chapter 2 Performance Verification Tests20. Overall Absolute Amplitude Accuracy: Agilent E4402B, E4403B, E4404B, E4405B, E4407B, and E4408Bbe used

Pagina 61

Chapter 2 153Performance Verification Tests20. Overall Absolute Amplitude Accuracy: Agilent E4402B, E4403B, E4404B, E4405B, E4407B,and E4408BFigure 2-

Pagina 62

154 Chapter 2 Performance Verification Tests20. Overall Absolute Amplitude Accuracy: Agilent E4402B, E4403B, E4404B, E4405B, E4407B, and E4408BMeasuri

Pagina 63 - 7. Noise Sidebands

Chapter 2 155Performance Verification Tests20. Overall Absolute Amplitude Accuracy: Agilent E4402B, E4403B, E4404B, E4405B, E4407B,and E4408BTable 2-3

Pagina 64

156 Chapter 2 Performance Verification Tests20. Overall Absolute Amplitude Accuracy: Agilent E4402B, E4403B, E4404B, E4405B, E4407B, and E4408Ba. Set

Pagina 65 - −10 kHz

Chapter 2 157Performance Verification Tests21. Resolution Bandwidth Accuracy21. Resolution Bandwidth AccuracyThe output of a synthesized signal genera

Pagina 66

158 Chapter 2 Performance Verification Tests21. Resolution Bandwidth AccuracyFigure 2-25 Resolution Bandwidth Accuracy Test Setup CAUTION Use only 75

Pagina 67 - −30 kHz:

Chapter 2 159Performance Verification Tests21. Resolution Bandwidth AccuracyAttenuator Error (6 dB) ___________ dB3 dB Resolution Bandwidth Accuracy6.

Pagina 68 - −100 kHz:

16 Chapter 1 Calibrating Calibration CycleTable 1-4 Agilent E4404B Performance Verification TestsPerformance Test NameCalibration for Instrument O

Pagina 69

160 Chapter 2 Performance Verification Tests21. Resolution Bandwidth AccuracyFrequency6 dB Resolution Bandwidth Accuracy18.Set the analyzer Res BW to

Pagina 70

Chapter 2 161Performance Verification Tests21. Resolution Bandwidth Accuracy28.Press Marker, Normal on the analyzer. 29.Repeat step 19 through step 28

Pagina 71

162 Chapter 2 Performance Verification Tests22. Frequency Response: Agilent E4401B and E4411B22. Frequency Response: Agilent E4401B and E4411BThis tes

Pagina 72 - −1 MHz:

Chapter 2 163Performance Verification Tests22. Frequency Response: Agilent E4401B and E4411BAdditional Equipment for 75 Ω InputPower sensor, 75 ΩMinim

Pagina 73 - −10 MHz:

164 Chapter 2 Performance Verification Tests22. Frequency Response: Agilent E4401B and E4411Bcalibration factor of the reference sensor for 100 kHz.75

Pagina 74

Chapter 2 165Performance Verification Tests22. Frequency Response: Agilent E4401B and E4411BTable 2-40 Source/Splitter CharacterizationFrequencyPower

Pagina 75 - 9. System-Related Sidebands

166 Chapter 2 Performance Verification Tests22. Frequency Response: Agilent E4401B and E4411BMeasuring Frequency Response, 100 kHz to 1.5 GHz1. Refer

Pagina 76

Chapter 2 167Performance Verification Tests22. Frequency Response: Agilent E4401B and E4411BVideo BW, 3 kHz (Man)5. Adjust the source AMPLITUDE to obt

Pagina 77

168 Chapter 2 Performance Verification Tests22. Frequency Response: Agilent E4401B and E4411B0.15 dB, Flatness Relative to 50 MHz would be –0.45 dB.Ta

Pagina 78 - 10. Residual FM

Chapter 2 169Performance Verification Tests22. Frequency Response: Agilent E4401B and E4411BMeasuring Frequency Response, ≤100 kHz If the analyzer has

Pagina 79

Chapter 1 17 CalibratingCalibration Cycle56.(This test has been deleted.)57.(This test has been deleted.)58.GSM Phase and Frequency Errorc•59. Comm

Pagina 80 - →, Mkr → CF

170 Chapter 2 Performance Verification Tests22. Frequency Response: Agilent E4401B and E4411B10.On the analyzer, press Peak Search (or Search).11.Adju

Pagina 81

Chapter 2 171Performance Verification Tests22. Frequency Response: Agilent E4401B and E4411B4. Enter the most negative number from the Flatness Relati

Pagina 82 - ∆ Mkr1) amplitude, take its

172 Chapter 2 Performance Verification Tests23. Frequency Response, Agilent E4402B and E4403B23. Frequency Response, Agilent E4402B and E4403BThis tes

Pagina 83 - 11. Sweep Time Accuracy

Chapter 2 173Performance Verification Tests23. Frequency Response, Agilent E4402B and E4403BProcedureSource/Splitter Characterization1. Connect the eq

Pagina 84

174 Chapter 2 Performance Verification Tests23. Frequency Response, Agilent E4402B and E4403B10.Adjust the source amplitude to obtain a Channel A powe

Pagina 85 - ∆ Mkr1)

Chapter 2 175Performance Verification Tests23. Frequency Response, Agilent E4402B and E4403B475 MHz575 MHz675 MHz775 MHz825 MHz875 MHz925 MHz975 MHz10

Pagina 86

176 Chapter 2 Performance Verification Tests23. Frequency Response, Agilent E4402B and E4403BMeasuring Frequency Response, 100 kHz to 3.0 GHz1. Refer

Pagina 87

Chapter 2 177Performance Verification Tests23. Frequency Response, Agilent E4402B and E4403BCF Step, 100 MHzSPAN, 20 kHzInput/Output (or Input), Coupl

Pagina 88

178 Chapter 2 Performance Verification Tests23. Frequency Response, Agilent E4402B and E4403Bas marker (Mkr1) Amplitude.24.Repeat step 18 through step

Pagina 89

Chapter 2 179Performance Verification Tests23. Frequency Response, Agilent E4402B and E4403BFor example, if Flatness Error is –0.30 dB and 50 MHz Ref

Pagina 90

18 Chapter 1 Calibrating Calibration CycleTable 1-5 Agilent E4405B Performance Verification TestsPerformance Test NameCalibration for Instrument O

Pagina 91

180 Chapter 2 Performance Verification Tests23. Frequency Response, Agilent E4402B and E4403B575 675 775 825 875 925 975 1025 1075 1175 1275 1375 1500

Pagina 92 - – 0 dB from R

Chapter 2 181Performance Verification Tests23. Frequency Response, Agilent E4402B and E4403B2925 2999 a. This entry is for data taken with the functio

Pagina 93 - Bandwidths

182 Chapter 2 Performance Verification Tests23. Frequency Response, Agilent E4402B and E4403BMeasuring Frequency Response Less Than or Equal to 100 kH

Pagina 94 - 89.3 mV==

Chapter 2 183Performance Verification Tests23. Frequency Response, Agilent E4402B and E4403BTable 2-46. Frequencies less than 9 kHz only apply to anal

Pagina 95

184 Chapter 2 Performance Verification Tests23. Frequency Response, Agilent E4402B and E4403BTest Results Perform the following steps to verify the fr

Pagina 96 - ATmeas–ATref+()10⁄

Chapter 2 185Performance Verification Tests23. Frequency Response, Agilent E4402B and E4403B13.Note the most positive number from the Flatness Relativ

Pagina 97

186 Chapter 2 Performance Verification Tests23. Frequency Response, Agilent E4402B and E4403Band 2.0 GHz. Record this number in Table 2-48 as the Mini

Pagina 98 - ∆Mkr Reading – Total

Chapter 2 187Performance Verification Tests24. Frequency Response, Agilent E4404B, E4405B, E4407B, and E4408B24. Frequency Response, Agilent E4404B, E

Pagina 99 - Resolution Bandwidths

188 Chapter 2 Performance Verification Tests24. Frequency Response, Agilent E4404B, E4405B, E4407B, and E4408BAdapter, APC 3.5 (f) to APC 3.5 (f) Adap

Pagina 100 - Ω to 75 Ω minimum loss pad

Chapter 2 189Performance Verification Tests24. Frequency Response, Agilent E4404B, E4405B, E4407B, and E4408Bpower meter reading of –10 dBm ±0.1 dB.7.

Pagina 101 - Procedure

Chapter 1 19 CalibratingCalibration Cycle56.(This test has been deleted.)57.(This test has been deleted.)58.GSM Phase and Frequency Errorc•59. Comm

Pagina 102

190 Chapter 2 Performance Verification Tests24. Frequency Response, Agilent E4404B, E4405B, E4407B, and E4408B20 MHz 0 dB (Ref)50 MHz75 MHz175 MHz275

Pagina 103 - Ω Input only)

Chapter 2 191Performance Verification Tests24. Frequency Response, Agilent E4404B, E4405B, E4407B, and E4408BMeasuring Frequency Response, 100 kHz to

Pagina 104 - −8.2 dBmV ± 0.1 dB

192 Chapter 2 Performance Verification Tests24. Frequency Response, Agilent E4404B, E4405B, E4407B, and E4408B2. Set the source frequency to 10 MHz.3.

Pagina 105

Chapter 2 193Performance Verification Tests24. Frequency Response, Agilent E4404B, E4405B, E4407B, and E4408B18.Set the function generator amplitude t

Pagina 106 - 17.On the analyzer, press

194 Chapter 2 Performance Verification Tests24. Frequency Response, Agilent E4404B, E4405B, E4407B, and E4408BSetup Change Error =__________ dB30.For

Pagina 107 - −90 dBm (−41.25

Chapter 2 195Performance Verification Tests24. Frequency Response, Agilent E4404B, E4405B, E4407B, and E4408B10c 20 50 0 dB (ref)0 dB (ref)75 175 275

Pagina 108

196 Chapter 2 Performance Verification Tests24. Frequency Response, Agilent E4404B, E4405B, E4407B, and E4408B1825 1875 1925 1975 2025 2125 2325 2525

Pagina 109

Chapter 2 197Performance Verification Tests24. Frequency Response, Agilent E4404B, E4405B, E4407B, and E4408BMeasuring Frequency Response Less Than or

Pagina 110 - 110 Chapter 2

198 Chapter 2 Performance Verification Tests24. Frequency Response, Agilent E4404B, E4405B, E4407B, and E4408B9. On the analyzer, press Peak Search (o

Pagina 111 - Ω input

Chapter 2 199Performance Verification Tests24. Frequency Response, Agilent E4404B, E4405B, E4407B, and E4408B500 Hza200 Hza100 Hzaa. These frequencies

Pagina 112 - Peak Search (or Search)

2 NoticeThe information contained in this document is subject to change without notice.Agilent Technologies makes no warranty of any kind with regar

Pagina 113

20 Chapter 1 Calibrating Calibration CycleTable 1-6 Agilent E4407B Performance Verification TestsPerformance Test NameCalibration for Instrument O

Pagina 114

200 Chapter 2 Performance Verification Tests24. Frequency Response, Agilent E4404B, E4405B, E4407B, and E4408BMeasuring Frequency Response, >3 GHz1

Pagina 115

Chapter 2 201Performance Verification Tests24. Frequency Response, Agilent E4404B, E4405B, E4407B, and E4408BPresel Center.13.Adjust the synthesized s

Pagina 116

202 Chapter 2 Performance Verification Tests24. Frequency Response, Agilent E4404B, E4405B, E4407B, and E4408B26.Repeat step 10 through step 14 for fr

Pagina 117 - −90 dBm (with

Chapter 2 203Performance Verification Tests24. Frequency Response, Agilent E4404B, E4405B, E4407B, and E4408B5.25 GHz5.5 GHz5.75 GHz6.0 GHz6.25 GHz6.5

Pagina 118

204 Chapter 2 Performance Verification Tests24. Frequency Response, Agilent E4404B, E4405B, E4407B, and E4408B13.5 GHz14.0 GHz14.5 GHz15.0 GHz15.5 GHz

Pagina 119 - 30.23dB–=

Chapter 2 205Performance Verification Tests24. Frequency Response, Agilent E4404B, E4405B, E4407B, and E4408BAgilent E4404B/E4405B Test Results Perfor

Pagina 120

206 Chapter 2 Performance Verification Tests24. Frequency Response, Agilent E4404B, E4405B, E4407B, and E4408B50 MHz (dc) column of Table 2-52 for fre

Pagina 121

Chapter 2 207Performance Verification Tests24. Frequency Response, Agilent E4404B, E4405B, E4407B, and E4408B6.7 GHz. Record this number in Table 2-54

Pagina 122

208 Chapter 2 Performance Verification Tests24. Frequency Response, Agilent E4404B, E4405B, E4407B, and E4408BAgilent E4407B/E4408B Test Results Perfo

Pagina 123

Chapter 2 209Performance Verification Tests24. Frequency Response, Agilent E4404B, E4405B, E4407B, and E4408Bcontinue with step 15.9. Note the most po

Pagina 124

Chapter 1 21 CalibratingCalibration Cycle52. First LO OUTPUT Power Accuracy •53. IF Input Accuracy •54. Comms Frequency Response ••55.(This test ha

Pagina 125 - Uncertainty

210 Chapter 2 Performance Verification Tests24. Frequency Response, Agilent E4404B, E4405B, E4407B, and E4408B20.For Band 2 in Table 2-55, subtract th

Pagina 126 - →, Mkr →CF, Peak Search (or

Chapter 2 211Performance Verification Tests24. Frequency Response, Agilent E4404B, E4405B, E4407B, and E4408BMHz (ac) column of Table 2-50. Record thi

Pagina 127

212 Chapter 2 Performance Verification Tests24. Frequency Response, Agilent E4404B, E4405B, E4407B, and E4408B14.Note the most negative number from th

Pagina 128

Chapter 2 213Performance Verification Tests25. Frequency Response (Preamp On): Agilent E4401B25. Frequency Response (Preamp On): Agilent E4401BThis te

Pagina 129 - Ω to Type-N (m) 50 Ω

214 Chapter 2 Performance Verification Tests25. Frequency Response (Preamp On): Agilent E4401BProcedureSource/Splitter Characterization1. Refer to Fi

Pagina 130 - Ω analyzers)

Chapter 2 215Performance Verification Tests25. Frequency Response (Preamp On): Agilent E4401B75 Ω inputs: Connect the reference sensor to the power sp

Pagina 131 - Chapter 2 131

216 Chapter 2 Performance Verification Tests25. Frequency Response (Preamp On): Agilent E4401BMeasuring Frequency Response, Preamp On1. Refer to Figur

Pagina 132 - Ω−24 dBm

Chapter 2 217Performance Verification Tests25. Frequency Response (Preamp On): Agilent E4401BFigure 2-37 Frequency Response Test Setup, Preamp On75 Ω

Pagina 133 - −25 dBm (50Ω Input only)

218 Chapter 2 Performance Verification Tests25. Frequency Response (Preamp On): Agilent E4401B8. Record the marker (Mkr1) amplitude reading in Table 2

Pagina 134

Chapter 2 219Performance Verification Tests25. Frequency Response (Preamp On): Agilent E4401BTest Results1. Record the most positive number from the F

Pagina 135

22 Chapter 1 Calibrating Calibration CycleTable 1-7 Agilent E4408B Performance Verification TestsPerformance Test NameCalibration for Instrument O

Pagina 136 - E4405B, E4407B, and E4408B

220 Chapter 2 Performance Verification Tests25. Frequency Response (Preamp On): Agilent E4401B3. In Table 2-59, subtract the Minimum Response value fr

Pagina 137 - −14 dBm ±0.1 dB

Chapter 2 221Performance Verification Tests26. Frequency Response (Preamp On): Agilent E4402B26. Frequency Response (Preamp On): Agilent E4402BThis te

Pagina 138 - −20.30 dB

222 Chapter 2 Performance Verification Tests26. Frequency Response (Preamp On): Agilent E4402Bcalibration factor of the reference sensor at 1 MHz.4. O

Pagina 139

Chapter 2 223Performance Verification Tests26. Frequency Response (Preamp On): Agilent E4402B15.Set the synthesized sweeper CW frequency to 10 MHz and

Pagina 140

224 Chapter 2 Performance Verification Tests26. Frequency Response (Preamp On): Agilent E4402BMeasuring Frequency Response, Preamp On1. Refer to Figur

Pagina 141

Chapter 2 225Performance Verification Tests26. Frequency Response (Preamp On): Agilent E4402Buse a cable. 2. Set the source frequency to 10 MHz.3. Set

Pagina 142 - , Auto Align, All

226 Chapter 2 Performance Verification Tests26. Frequency Response (Preamp On): Agilent E4402B10.Set the analyzer center frequency to the next frequen

Pagina 143

Chapter 2 227Performance Verification Tests26. Frequency Response (Preamp On): Agilent E4402Bgenerator) as follows:a. Record the Flatness Error from T

Pagina 144 - 1 dB Actual

228 Chapter 2 Performance Verification Tests26. Frequency Response (Preamp On): Agilent E4402B175 MHz275 MHz375 MHz475 MHz575 MHz675 MHz775 MHz825 MHz

Pagina 145 - Ω adapter

Chapter 2 229Performance Verification Tests26. Frequency Response (Preamp On): Agilent E4402BTest Results1. Record the most positive number from the F

Pagina 146 - ActualTotalAtten RefAtten

Chapter 1 23 CalibratingCalibration CycleTable 1-8 Agilent E4411B Performance Verification TestsPerformance Test NameCalibration for Instrument Op

Pagina 147

230 Chapter 2 Performance Verification Tests26. Frequency Response (Preamp On): Agilent E4402B9. Note the most negative number from the Flatness Relat

Pagina 148

Chapter 2 231Performance Verification Tests27. Frequency Response (Preamp On): Agilent E4404B, E4405B, and E4407B27. Frequency Response (Preamp On): A

Pagina 149 - Chapter 2 149

232 Chapter 2 Performance Verification Tests27. Frequency Response (Preamp On): Agilent E4404B, E4405B, and E4407BChannel B of the power meter. This w

Pagina 150 - Ω Input

Chapter 2 233Performance Verification Tests27. Frequency Response (Preamp On): Agilent E4404B, E4405B, and E4407BTable 2-63.13.Repeat step 9 through s

Pagina 151

234 Chapter 2 Performance Verification Tests27. Frequency Response (Preamp On): Agilent E4404B, E4405B, and E4407B775 MHz825 MHz875 MHz925 MHz975 MHz1

Pagina 152

Chapter 2 235Performance Verification Tests27. Frequency Response (Preamp On): Agilent E4404B, E4405B, and E4407BMeasuring Frequency Response, Preamp

Pagina 153

236 Chapter 2 Performance Verification Tests27. Frequency Response (Preamp On): Agilent E4404B, E4405B, and E4407B5. On the analyzer, press Peak Searc

Pagina 154

Chapter 2 237Performance Verification Tests27. Frequency Response (Preamp On): Agilent E4404B, E4405B, and E4407B27.Calculate the Flatness Error for e

Pagina 155

238 Chapter 2 Performance Verification Tests27. Frequency Response (Preamp On): Agilent E4404B, E4405B, and E4407B33.If the analyzer is an Agilent E44

Pagina 156

Chapter 2 239Performance Verification Tests27. Frequency Response (Preamp On): Agilent E4404B, E4405B, and E4407BTable 2-64 Frequency Response Worksh

Pagina 157 - Equipment Required

24 Chapter 1 Calibrating Before You StartBefore You Start This brief procedure should be performed before starting the performance verification tes

Pagina 158

240 Chapter 2 Performance Verification Tests27. Frequency Response (Preamp On): Agilent E4404B, E4405B, and E4407BTest Results1. Record the most posit

Pagina 159 - − Lower Marker

Chapter 2 241Performance Verification Tests27. Frequency Response (Preamp On): Agilent E4404B, E4405B, and E4407B5. If the analyzer is an Agilent E440

Pagina 160 - −7dBm ±0.2 dB

242 Chapter 2 Performance Verification Tests27. Frequency Response (Preamp On): Agilent E4404B, E4405B, and E4407B15.The frequency response (preamp on

Pagina 161

Chapter 2 243Performance Verification Tests27. Frequency Response (Preamp On): Agilent E4404B, E4405B, and E4407B25.Transfer the values entered in Tab

Pagina 162 - Ω, BNC (m)

244 Chapter 2 Performance Verification Tests28. Other Input-Related Spurious Responses: Agilent E4401B and E4411B28. Other Input-Related Spurious Resp

Pagina 163

Chapter 2 245Performance Verification Tests28. Other Input-Related Spurious Responses: Agilent E4401B and E4411B75 Ω Input only: Use the minimum loss

Pagina 164 - ±0.1 dB

246 Chapter 2 Performance Verification Tests28. Other Input-Related Spurious Responses: Agilent E4401B and E4411B16.On the synthesized sweeper, press

Pagina 165 - Ω inputs

Chapter 2 247Performance Verification Tests28. Other Input-Related Spurious Responses: Agilent E4401B and E4411B24.On the synthesized sweeper, press R

Pagina 166

248 Chapter 2 Performance Verification Tests28. Other Input-Related Spurious Responses: Agilent E4401B and E4411BTable 2-67 Other Input-Related Spuri

Pagina 167 - ±0.1 dB for the current

Chapter 2 249Performance Verification Tests29. Other Input-Related Spurious Responses: Agilent E4402B, E4403B, E4404B, E4405B,E4407B, and E4408B29. Ot

Pagina 168 - Ω inputs (Option 1DP)

Chapter 1 25 CalibratingBefore You StartPeriodic verification of operation The analyzer requires periodic verification of operation. Under most con

Pagina 169 - ≤100 kHz

250 Chapter 2 Performance Verification Tests29. Other Input-Related Spurious Responses: Agilent E4402B, E4403B, E4404B, E4405B, E4407B, and E4408BFigu

Pagina 170

Chapter 2 251Performance Verification Tests29. Other Input-Related Spurious Responses: Agilent E4402B, E4403B, E4404B, E4405B,E4407B, and E4408B10.Rep

Pagina 171

252 Chapter 2 Performance Verification Tests29. Other Input-Related Spurious Responses: Agilent E4402B, E4403B, E4404B, E4405B, E4407B, and E4408Bfreq

Pagina 172 - Ω load replaces the power

Chapter 2 253Performance Verification Tests29. Other Input-Related Spurious Responses: Agilent E4402B, E4403B, E4404B, E4405B,E4407B, and E4408B30.On

Pagina 173

254 Chapter 2 Performance Verification Tests29. Other Input-Related Spurious Responses: Agilent E4402B, E4403B, E4404B, E4405B, E4407B, and E4408B15.0

Pagina 174

Chapter 2 255Performance Verification Tests30. Spurious Responses: Agilent E4401B and E4411B30. Spurious Responses: Agilent E4401B and E4411BThis test

Pagina 175

256 Chapter 2 Performance Verification Tests30. Spurious Responses: Agilent E4401B and E4411BAdapter, Type-N (m) to BNC (f) Adapter, Type-N (m) to Typ

Pagina 176

Chapter 2 257Performance Verification Tests30. Spurious Responses: Agilent E4401B and E4411BFigure 2-45 Third Order Intermodulation Distortion Test S

Pagina 177 - ±0.1 dB for 100 kHz

258 Chapter 2 Performance Verification Tests30. Spurious Responses: Agilent E4401B and E4411Bfrequency into the power meter. 7. Adjust the amplitude o

Pagina 178 - FlatError

Chapter 2 259Performance Verification Tests30. Spurious Responses: Agilent E4401B and E4411B17.Set the analyzer reference level to –15 dBm. 75 Ω Input

Pagina 179

26 Chapter 1 Calibrating Before You StartOscilloscope Bandwidth: dc to 100 MHz Vertical Scale Factor of 0.5 V to 5 V/DivTwo channelsMinimum Timebas

Pagina 180

260 Chapter 2 Performance Verification Tests30. Spurious Responses: Agilent E4401B and E4411B30.If the analyzer is an E4401B and is equipped with Opti

Pagina 181

Chapter 2 261Performance Verification Tests30. Spurious Responses: Agilent E4401B and E4411B36.Repeat step 10 through step 29. This is the TOI test fo

Pagina 182

262 Chapter 2 Performance Verification Tests30. Spurious Responses: Agilent E4401B and E4411B5. Set the analyzer control as follows: SPAN, 50 kHzBW/Av

Pagina 183

Chapter 2 263Performance Verification Tests31. Spurious Responses: Agilent E4402B and E4403B31. Spurious Responses: Agilent E4402B and E4403BThis test

Pagina 184 - 184 Chapter 2

264 Chapter 2 Performance Verification Tests31. Spurious Responses: Agilent E4402B and E4403BAdapter, Type-N (m) to BNC (f) Adapter, Type-N (m) to Typ

Pagina 185 - Chapter 2 185

Chapter 2 265Performance Verification Tests31. Spurious Responses: Agilent E4402B and E4403B3. Perform step 4 through step 29 using the information an

Pagina 186

266 Chapter 2 Performance Verification Tests31. Spurious Responses: Agilent E4402B and E4403BMarker →Mkr → CFDelta11.On the analyzer, press: FREQUENCY

Pagina 187

Chapter 2 267Performance Verification Tests31. Spurious Responses: Agilent E4402B and E4403BDistortion Amplitude (in dB) from the Mixer Level (in dBm)

Pagina 188

268 Chapter 2 Performance Verification Tests31. Spurious Responses: Agilent E4402B and E4403B36.Repeat step 10 through step 29. This is the TOI test f

Pagina 189

Chapter 2 269Performance Verification Tests31. Spurious Responses: Agilent E4402B and E4403B6. Record the power meter reading at 300 MHz in Table 2-73

Pagina 190

Chapter 1 27 CalibratingBefore You StartWide Offset Phase Noise Signal GeneratorFrequency Range: 1 GHz ±1 MHzAmplitude Range: 0 dBM ±5 dBPhase Nois

Pagina 191

270 Chapter 2 Performance Verification Tests31. Spurious Responses: Agilent E4402B and E4403B17.Record the power meter reading in Table 2-73 as the 90

Pagina 192

Chapter 2 271Performance Verification Tests31. Spurious Responses: Agilent E4402B and E4403B5. Adjust the synthesized sweeper power level until the po

Pagina 193

272 Chapter 2 Performance Verification Tests31. Spurious Responses: Agilent E4402B and E4403BAMPLITUDE, Ref Level, –20 dBmBW/Avg, Video BW, 30 Hz (Man

Pagina 194

Chapter 2 273Performance Verification Tests32. Spurious Responses: Agilent E4404B, E4405B, E4407B, and E4408B32. Spurious Responses: Agilent E4404B, E

Pagina 195

274 Chapter 2 Performance Verification Tests32. Spurious Responses: Agilent E4404B, E4405B, E4407B, and E4408B1.8 GHz low pass filter (2 required)4.4

Pagina 196

Chapter 2 275Performance Verification Tests32. Spurious Responses: Agilent E4404B, E4405B, E4407B, and E4408BFigure 2-49 Third Order Intermodulation

Pagina 197

276 Chapter 2 Performance Verification Tests32. Spurious Responses: Agilent E4404B, E4405B, E4407B, and E4408Bmeter reads –12 dBm ±0.1 dB.8. Disconnec

Pagina 198

Chapter 2 277Performance Verification Tests32. Spurious Responses: Agilent E4404B, E4405B, E4407B, and E4408B20.On the analyzer, press BW/Avg, Average

Pagina 199

278 Chapter 2 Performance Verification Tests32. Spurious Responses: Agilent E4404B, E4405B, E4407B, and E4408B31.If the analyzer is an E4404B, E4405B

Pagina 200

Chapter 2 279Performance Verification Tests32. Spurious Responses: Agilent E4404B, E4405B, E4407B, and E4408BPart 2: Second Harmonic Distortion 1. Zer

Pagina 201 - ↑ will allow you to step

28 Chapter 1 Calibrating Before You StartAttenuator, 1 dB StepAttenuation Range: 0 to 11 dB Frequency Range: 50 MHz ±1 MHzConnectors: Type-N female

Pagina 202

280 Chapter 2 Performance Verification Tests32. Spurious Responses: Agilent E4404B, E4405B, E4407B, and E4408BFigure 2-50 Second Harmonic Distortion

Pagina 203

Chapter 2 281Performance Verification Tests32. Spurious Responses: Agilent E4404B, E4405B, E4407B, and E4408B11.Subtract the power meter reading at 60

Pagina 204

282 Chapter 2 Performance Verification Tests32. Spurious Responses: Agilent E4404B, E4405B, E4407B, and E4408Bat 900 MHz is –7.05 dBm, the 900 MHz Fre

Pagina 205 - Chapter 2 205

Chapter 2 283Performance Verification Tests32. Spurious Responses: Agilent E4404B, E4405B, E4407B, and E4408Bwould be –1.2 dB:Measuring 3.1 GHz Freque

Pagina 206 - 206 Chapter 2

284 Chapter 2 Performance Verification Tests32. Spurious Responses: Agilent E4404B, E4405B, E4407B, and E4408Bbe –1.2 dB:Measuring 300 MHz Second Harm

Pagina 207

Chapter 2 285Performance Verification Tests32. Spurious Responses: Agilent E4404B, E4405B, E4407B, and E4408Breading is –10 dBm ±0.2 dB.6. On the anal

Pagina 208

286 Chapter 2 Performance Verification Tests32. Spurious Responses: Agilent E4404B, E4405B, E4407B, and E4408BAMPLITUDE, Ref Level, –20 dBmBW/Avg, Vid

Pagina 209 - Chapter 2 209

Chapter 2 287Performance Verification Tests32. Spurious Responses: Agilent E4404B, E4405B, E4407B, and E4408BMeasuring 1.55 GHz Second Harmonic Distor

Pagina 210

288 Chapter 2 Performance Verification Tests32. Spurious Responses: Agilent E4404B, E4405B, E4407B, and E4408B13.Calculate the 1.55 GHz Second Harmoni

Pagina 211 - Chapter 2 211

Chapter 2 289Performance Verification Tests32. Spurious Responses: Agilent E4404B, E4405B, E4407B, and E4408Bthe synthesized sweeper output directly t

Pagina 212

Chapter 1 29 CalibratingBefore You StartPower Splitter(for E4401B/ 02B/03B/11B)Frequency Range: 9 kHz to 13.2 GHz Insertion Loss: 6 dB nominal Outp

Pagina 213 - Ω to Type-N (f) 75 Ω

290 Chapter 2 Performance Verification Tests33. Gain Compression: Agilent E4401B, E4402B, E4403B, and E4411B33. Gain Compression: Agilent E4401B, E440

Pagina 214

Chapter 2 291Performance Verification Tests33. Gain Compression: Agilent E4401B, E4402B, E4403B, and E4411Bdirectional bridge connected to the power s

Pagina 215

292 Chapter 2 Performance Verification Tests33. Gain Compression: Agilent E4401B, E4402B, E4403B, and E4411B7. Record the actual synthesized sweeper p

Pagina 216

Chapter 2 293Performance Verification Tests33. Gain Compression: Agilent E4401B, E4402B, E4403B, and E4411B10.Set the synthesized sweeper amplitude Of

Pagina 217

294 Chapter 2 Performance Verification Tests34. Gain Compression: Agilent E4404B, E4405B, E4407B, and E4408B34. Gain Compression: Agilent E4404B, E440

Pagina 218 - ±0.2 dB

Chapter 2 295Performance Verification Tests34. Gain Compression: Agilent E4404B, E4405B, E4407B, and E4408Bless than or equal to 2.5 GHz. Use the dire

Pagina 219

296 Chapter 2 Performance Verification Tests34. Gain Compression: Agilent E4404B, E4405B, E4407B, and E4408BTable 2-80 for each frequency indicated.Ta

Pagina 220

Chapter 2 297Performance Verification Tests34. Gain Compression: Agilent E4404B, E4405B, E4407B, and E4408B8. Repeat step 3 through step 7 for each of

Pagina 221 - Agilent E4402B

298 Chapter 2 Performance Verification Tests34. Gain Compression: Agilent E4404B, E4405B, E4407B, and E4408B21.Replace the directional bridge with the

Pagina 222

Chapter 2 299Performance Verification Tests35. Displayed Average Noise Level: Agilent E4401B and E4411B35. Displayed Average Noise Level: Agilent E440

Pagina 223

3Where to Find the Latest InformationDocumentation is updated periodically. For the latest information about Agilent ESA Spectrum Analyzers, includi

Pagina 224

30 Chapter 1 Calibrating Before You Start1.8 GHz Low Pass Filter (for E4404/5/7/8B)Cutoff Frequency: 1.8 GHzRejection at >3 GHz: >45 dB 0955-

Pagina 225

300 Chapter 2 Performance Verification Tests35. Displayed Average Noise Level: Agilent E4401B and E4411BFigure 2-53 Displayed Average Noise Level Tes

Pagina 226

Chapter 2 301Performance Verification Tests35. Displayed Average Noise Level: Agilent E4401B and E4411BGenerator) installed and the minimum RBW is 100

Pagina 227

302 Chapter 2 Performance Verification Tests35. Displayed Average Noise Level: Agilent E4401B and E4411BRef Lvl Offset (100 Hz/10 Hz RBW) ____________

Pagina 228

Chapter 2 303Performance Verification Tests35. Displayed Average Noise Level: Agilent E4401B and E4411Bperform those procedures listed in Subset C in

Pagina 229

304 Chapter 2 Performance Verification Tests35. Displayed Average Noise Level: Agilent E4401B and E4411BTable 2-81 DANL Measurement Sequence, E4401B

Pagina 230

Chapter 2 305Performance Verification Tests35. Displayed Average Noise Level: Agilent E4401B and E4411BMeasure DANL 10 MHz 500 MHz 1 kHz Off 3)Measure

Pagina 231

306 Chapter 2 Performance Verification Tests35. Displayed Average Noise Level: Agilent E4401B and E4411BMeasuring Displayed Average Noise Level (DANL)

Pagina 232

Chapter 2 307Performance Verification Tests35. Displayed Average Noise Level: Agilent E4401B and E4411BHz) if test RBW = 100 Hz or 10 Hz)BW/Avg, Res B

Pagina 233

308 Chapter 2 Performance Verification Tests35. Displayed Average Noise Level: Agilent E4401B and E4411BAMPLITUDE, More, Ref Lvl Offst, (enter Ref Lvl

Pagina 234

Chapter 2 309Performance Verification Tests36. Displayed Average Noise Level: Agilent E4402B and E4403B36. Displayed Average Noise Level: Agilent E440

Pagina 235

Chapter 1 31 CalibratingBefore You StartTable 1-11 Recommended Adapters Critical Specifications for Adapter Substitution Recommended HP/Agilent Mo

Pagina 236

310 Chapter 2 Performance Verification Tests36. Displayed Average Noise Level: Agilent E4402B and E4403BFigure 2-54 Displayed Average Noise Level Tes

Pagina 237

Chapter 2 311Performance Verification Tests36. Displayed Average Noise Level: Agilent E4402B and E4403Bcontinue with step 11.6. If the analyzer has a

Pagina 238 - 238 Chapter 2

312 Chapter 2 Performance Verification Tests36. Displayed Average Noise Level: Agilent E4402B and E4403BFigure 2-54.17.Disconnect the BNC cable and ad

Pagina 239

Chapter 2 313Performance Verification Tests36. Displayed Average Noise Level: Agilent E4402B and E4403BSubset D in Table 2-85.8. After performing all

Pagina 240

314 Chapter 2 Performance Verification Tests36. Displayed Average Noise Level: Agilent E4402B and E4403Bcontinue with “Remove Reference Level Offset”.

Pagina 241 - Chapter 2 241

Chapter 2 315Performance Verification Tests36. Displayed Average Noise Level: Agilent E4402B and E4403BVideo BW, 1 Hz.8. On the analyzer, press Single

Pagina 242

316 Chapter 2 Performance Verification Tests37. Displayed Average Noise Level: Agilent E4404B and E4405B37. Displayed Average Noise Level: Agilent E44

Pagina 243

Chapter 2 317Performance Verification Tests37. Displayed Average Noise Level: Agilent E4404B and E4405BFigure 2-55 Displayed Average Noise Level Test

Pagina 244 - Agilent E4401B and E4411B

318 Chapter 2 Performance Verification Tests37. Displayed Average Noise Level: Agilent E4404B and E4405B6. If the analyzer has a minimum RBW of 1 kHz

Pagina 245

Chapter 2 319Performance Verification Tests37. Displayed Average Noise Level: Agilent E4404B and E4405B17.If the analyzer has Option 1DN, 50 Ω trackin

Pagina 246 - −10 dBm

32 Chapter 1 Calibrating Before You StartTable 1-12 Recommended Cables Critical Specifications for Cable SubstitutionRecommended HP/Agilent ModelU

Pagina 247 - Ω Input)

320 Chapter 2 Performance Verification Tests37. Displayed Average Noise Level: Agilent E4404B and E4405BMeasure DANL 1 GHz 2 GHz 1 kHz Off 2)Measure D

Pagina 248

Chapter 2 321Performance Verification Tests37. Displayed Average Noise Level: Agilent E4404B and E4405BMeasuring Displayed Average Noise Level (DANL)U

Pagina 249 - E4407B, and E4408B

322 Chapter 2 Performance Verification Tests37. Displayed Average Noise Level: Agilent E4404B and E4405BRBW = 1 kHz)AMPLITUDE, More, Ref Lvl Offst, (e

Pagina 250 - −20 dBm ±0.1 dB

Chapter 2 323Performance Verification Tests38. Displayed Average Noise Level: Agilent E4407B and E4408B38. Displayed Average Noise Level: Agilent E440

Pagina 251

324 Chapter 2 Performance Verification Tests38. Displayed Average Noise Level: Agilent E4407B and E4408BFigure 2-56 Displayed Average Noise Level Tes

Pagina 252

Chapter 2 325Performance Verification Tests38. Displayed Average Noise Level: Agilent E4407B and E4408B6. If the analyzer does not have Option 1DN (3.

Pagina 253 - ∆ Mkr1 or

326 Chapter 2 Performance Verification Tests38. Displayed Average Noise Level: Agilent E4407B and E4408B17.On the analyzer, press Input, Amptd Ref Out

Pagina 254

Chapter 2 327Performance Verification Tests38. Displayed Average Noise Level: Agilent E4407B and E4408BSubset D in Table 2-89.Table 2-89 DANL Measure

Pagina 255

328 Chapter 2 Performance Verification Tests38. Displayed Average Noise Level: Agilent E4407B and E4408B8. After performing all applicable DANL measur

Pagina 256

Chapter 2 329Performance Verification Tests38. Displayed Average Noise Level: Agilent E4407B and E4408BRBW = 1 kHz)AMPLITUDE, More, Ref Lvl Offst, (en

Pagina 257

332 Performance Verification Tests These tests verify the electrical performance of the analyzer. Allow the analyzer to warm up in accordance with the

Pagina 258 - → CF, Delta

330 Chapter 2 Performance Verification Tests39. Residual Responses39. Residual ResponsesThe analyzer input is terminated and the analyzer is swept fro

Pagina 259 - ↑, ↑, ↑. The center

Chapter 2 331Performance Verification Tests39. Residual ResponsesFigure 2-57 Residual Response Test Setup 2. Press Preset on the analyzer. Press the

Pagina 260

332 Chapter 2 Performance Verification Tests39. Residual Responsesbeen checked for the model and frequency ranges below.6. Press Single and wait for a

Pagina 261

Chapter 2 333Performance Verification Tests40. Fast Time Domain Amplitude Accuracy: Agilent E4401B, E4402B, E4404B, E4405B,and E4407B (Option AYX)40.

Pagina 262 - ∆ Mkr1) amplitude

334 Chapter 2 Performance Verification Tests40. Fast Time Domain Amplitude Accuracy: Agilent E4401B, E4402B, E4404B, E4405B, and E4407B (Option AYX)In

Pagina 263

Chapter 2 335Performance Verification Tests41. Tracking Generator Absolute Amplitude and Vernier Accuracy: Agilent E4401B andE4411B (Option 1DN or 1DQ

Pagina 264

336 Chapter 2 Performance Verification Tests41. Tracking Generator Absolute Amplitude and Vernier Accuracy: Agilent E4401B and E4411B (Option 1DN or 1

Pagina 265

Chapter 2 337Performance Verification Tests41. Tracking Generator Absolute Amplitude and Vernier Accuracy: Agilent E4401B andE4411B (Option 1DN or 1DQ

Pagina 266 - ↓, ↓. The center frequency

338 Chapter 2 Performance Verification Tests41. Tracking Generator Absolute Amplitude and Vernier Accuracy: Agilent E4401B and E4411B (Option 1DN or 1

Pagina 267

Chapter 2 339Performance Verification Tests41. Tracking Generator Absolute Amplitude and Vernier Accuracy: Agilent E4401B andE4411B (Option 1DN or 1DQ

Pagina 268

34 Chapter 2 Performance Verification TestsTests included in this section:Tests included in this section:1. 10 MHz Reference Output Accuracy2. 10 MHz

Pagina 269 - ±0.10 dB

340 Chapter 2 Performance Verification Tests42. Tracking Generator Absolute Amplitude and Vernier Accuracy: Agilent E4402B, E4403B, E4404B, E4405B, E4

Pagina 270

Chapter 2 341Performance Verification Tests42. Tracking Generator Absolute Amplitude and Vernier Accuracy: Agilent E4402B, E4403B,E4404B, E4405B, E440

Pagina 271

342 Chapter 2 Performance Verification Tests42. Tracking Generator Absolute Amplitude and Vernier Accuracy: Agilent E4402B, E4403B, E4404B, E4405B, E4

Pagina 272

Chapter 2 343Performance Verification Tests43. Tracking Generator Level Flatness: Agilent E4401B and E4411B (Option 1DN or 1DQ)43. Tracking Generator

Pagina 273

344 Chapter 2 Performance Verification Tests43. Tracking Generator Level Flatness: Agilent E4401B and E4411B (Option 1DN or 1DQ)Additional Equipment f

Pagina 274

Chapter 2 345Performance Verification Tests43. Tracking Generator Level Flatness: Agilent E4401B and E4411B (Option 1DN or 1DQ)relative to the power l

Pagina 275

346 Chapter 2 Performance Verification Tests43. Tracking Generator Level Flatness: Agilent E4401B and E4411B (Option 1DN or 1DQ)5. Disconnect the powe

Pagina 276 - ≥1 kHz, press SPAN, 4 kHz

Chapter 2 347Performance Verification Tests43. Tracking Generator Level Flatness: Agilent E4401B and E4411B (Option 1DN or 1DQ)Tracking Generator Leve

Pagina 277

348 Chapter 2 Performance Verification Tests43. Tracking Generator Level Flatness: Agilent E4401B and E4411B (Option 1DN or 1DQ)4. Press Single.5. Rec

Pagina 278

Chapter 2 349Performance Verification Tests43. Tracking Generator Level Flatness: Agilent E4401B and E4411B (Option 1DN or 1DQ)performance verificatio

Pagina 279

Chapter 2 35Performance Verification TestsTests included in this section:23. Frequency Response, Agilent E4402B and E4403B24. Frequency Response, Agil

Pagina 280

350 Chapter 2 Performance Verification Tests44. Tracking Generator Level Flatness: Agilent E4402B, E4403B, E4404B, E4405B, E4407B, and E4408B (Option

Pagina 281

Chapter 2 351Performance Verification Tests44. Tracking Generator Level Flatness: Agilent E4402B, E4403B, E4404B, E4405B, E4407B,and E4408B (Option 1D

Pagina 282

352 Chapter 2 Performance Verification Tests44. Tracking Generator Level Flatness: Agilent E4402B, E4403B, E4404B, E4405B, E4407B, and E4408B (Option

Pagina 283

Chapter 2 353Performance Verification Tests44. Tracking Generator Level Flatness: Agilent E4402B, E4403B, E4404B, E4405B, E4407B,and E4408B (Option 1D

Pagina 284

354 Chapter 2 Performance Verification Tests44. Tracking Generator Level Flatness: Agilent E4402B, E4403B, E4404B, E4405B, E4407B, and E4408B (Option

Pagina 285

Chapter 2 355Performance Verification Tests44. Tracking Generator Level Flatness: Agilent E4402B, E4403B, E4404B, E4405B, E4407B,and E4408B (Option 1D

Pagina 286

356 Chapter 2 Performance Verification Tests45. Tracking Generator Harmonic Spurious Outputs: Agilent E4401B and E4411B (Option 1DN or 1DQ)45. Tracki

Pagina 287

Chapter 2 357Performance Verification Tests45. Tracking Generator Harmonic Spurious Outputs: Agilent E4401B and E4411B (Option1DN or 1DQ)Figure 2-65

Pagina 288

358 Chapter 2 Performance Verification Tests45. Tracking Generator Harmonic Spurious Outputs: Agilent E4401B and E4411B (Option 1DN or 1DQ)Source, Am

Pagina 289

Chapter 2 359Performance Verification Tests45. Tracking Generator Harmonic Spurious Outputs: Agilent E4401B and E4411B (Option1DN or 1DQ)f. Press MKR

Pagina 290 - Ω to Type-N (f), 75 Ω

36 Chapter 2 Performance Verification TestsTests included in this section:46. Tracking Generator Harmonic Spurious Outputs: Agilent E4402B, E4403B, E4

Pagina 291 - Ω analyzer

360 Chapter 2 Performance Verification Tests46. Tracking Generator Harmonic Spurious Outputs: Agilent E4402B, E4403B, E4404B, E4405B, E4407B, and E440

Pagina 292

Chapter 2 361Performance Verification Tests46. Tracking Generator Harmonic Spurious Outputs: Agilent E4402B, E4403B, E4404B,E4405B, E4407B, and E4408B

Pagina 293 - ∆ Mkr1) amplitude is the

362 Chapter 2 Performance Verification Tests46. Tracking Generator Harmonic Spurious Outputs: Agilent E4402B, E4403B, E4404B, E4405B, E4407B, and E440

Pagina 294

Chapter 2 363Performance Verification Tests46. Tracking Generator Harmonic Spurious Outputs: Agilent E4402B, E4403B, E4404B,E4405B, E4407B, and E4408B

Pagina 295

364 Chapter 2 Performance Verification Tests47. Tracking Generator Non-Harmonic Spurious Outputs: Agilent E4401B and E4411B (Option 1DN or 1DQ)47. Tra

Pagina 296

Chapter 2 365Performance Verification Tests47. Tracking Generator Non-Harmonic Spurious Outputs: Agilent E4401B and E4411B (Option1DN or 1DQ)Figure 2-

Pagina 297

366 Chapter 2 Performance Verification Tests47. Tracking Generator Non-Harmonic Spurious Outputs: Agilent E4401B and E4411B (Option 1DN or 1DQ)FREQUEN

Pagina 298 - Table 2-79

Chapter 2 367Performance Verification Tests47. Tracking Generator Non-Harmonic Spurious Outputs: Agilent E4401B and E4411B (Option1DN or 1DQ)75 Ω outp

Pagina 299

368 Chapter 2 Performance Verification Tests47. Tracking Generator Non-Harmonic Spurious Outputs: Agilent E4401B and E4411B (Option 1DN or 1DQ)analyze

Pagina 300

Chapter 2 369Performance Verification Tests47. Tracking Generator Non-Harmonic Spurious Outputs: Agilent E4401B and E4411B (Option1DN or 1DQ)Determini

Pagina 301

Chapter 2 37Performance Verification TestsTests included in this section:CalibrationTo perform calibration:1. Run all performance verification tests l

Pagina 302 - _____ dB

370 Chapter 2 Performance Verification Tests48. Tracking Generator Non-Harmonic Spurious Outputs: Agilent E4402B, E4403B, E4404B, E4405B, E4407B, and

Pagina 303 - Chapter 2 303

Chapter 2 371Performance Verification Tests48. Tracking Generator Non-Harmonic Spurious Outputs: Agilent E4402B, E4403B, E4404B,E4405B, E4407B, and E4

Pagina 304 - Ω Inputs

372 Chapter 2 Performance Verification Tests48. Tracking Generator Non-Harmonic Spurious Outputs: Agilent E4402B, E4403B, E4404B, E4405B, E4407B, and

Pagina 305

Chapter 2 373Performance Verification Tests48. Tracking Generator Non-Harmonic Spurious Outputs: Agilent E4402B, E4403B, E4404B,E4405B, E4407B, and E4

Pagina 306

374 Chapter 2 Performance Verification Tests48. Tracking Generator Non-Harmonic Spurious Outputs: Agilent E4402B, E4403B, E4404B, E4405B, E4407B, and

Pagina 307 - Ω input only)

Chapter 2 375Performance Verification Tests48. Tracking Generator Non-Harmonic Spurious Outputs: Agilent E4402B, E4403B, E4404B,E4405B, E4407B, and E4

Pagina 308 - 1. Press

376 Chapter 2 Performance Verification Tests49. Tracking Generator LO Feedthrough: Agilent E4402B, E4403B, E4404B, E4405B, E4407B, and E4408B(Option 1

Pagina 309 - Ω Input using a BNC cable

Chapter 2 377Performance Verification Tests49. Tracking Generator LO Feedthrough: Agilent E4402B, E4403B, E4404B, E4405B, E4407B,and E4408B(Option 1DN

Pagina 310

378 Chapter 2 Performance Verification Tests49. Tracking Generator LO Feedthrough: Agilent E4402B, E4403B, E4404B, E4405B, E4407B, and E4408B(Option 1

Pagina 311

Chapter 2 379Performance Verification Tests50. Gate Delay Accuracy and Gate Length Accuracy: Agilent E4401B, E4402B, E4404B,E4405B, and E4407B (Option

Pagina 312 - Ω tracking generator, do the

38 Chapter 2 Performance Verification TestsTests included in this section:Table 2-1 Calibration RequirementsTest #Performance Verification Tests Cali

Pagina 313 - ° to 30° C range

380 Chapter 2 Performance Verification Tests50. Gate Delay Accuracy and Gate Length Accuracy: Agilent E4401B, E4402B, E4404B, E4405B, and E4407B (Opti

Pagina 314 - → , Mkr→ CF

Chapter 2 381Performance Verification Tests50. Gate Delay Accuracy and Gate Length Accuracy: Agilent E4401B, E4402B, E4404B,E4405B, and E4407B (Option

Pagina 315

382 Chapter 2 Performance Verification Tests50. Gate Delay Accuracy and Gate Length Accuracy: Agilent E4401B, E4402B, E4404B, E4405B, and E4407B (Opti

Pagina 316

Chapter 2 383Performance Verification Tests50. Gate Delay Accuracy and Gate Length Accuracy: Agilent E4401B, E4402B, E4404B,E4405B, and E4407B (Option

Pagina 317

384 Chapter 2 Performance Verification Tests51. Gate Mode Additional Amplitude Error: Agilent E4401B, E4402B, E4404B, E4405B, and E4407B (Option 1D6)5

Pagina 318

Chapter 2 385Performance Verification Tests51. Gate Mode Additional Amplitude Error: Agilent E4401B, E4402B, E4404B, E4405B, andE4407B (Option 1D6)Fig

Pagina 319

386 Chapter 2 Performance Verification Tests51. Gate Mode Additional Amplitude Error: Agilent E4401B, E4402B, E4404B, E4405B, and E4407B (Option 1D6)P

Pagina 320

Chapter 2 387Performance Verification Tests52. First LO OUTPUT Power Accuracy (Option AYZ only)52. First LO OUTPUT Power Accuracy (Option AYZ only)Thi

Pagina 321 - ° to 30° C range, but

388 Chapter 2 Performance Verification Tests52. First LO OUTPUT Power Accuracy (Option AYZ only)Procedure1. Zero and calibrate the power sensor and po

Pagina 322 - →, Μκρ→ ΧΦ

Chapter 2 389Performance Verification Tests52. First LO OUTPUT Power Accuracy (Option AYZ only)Post-Test Instrument Restoration12.Disconnect the power

Pagina 323

Chapter 2 39Performance Verification TestsTests included in this section:24. Frequency Response Frequency Response25. Frequency Response (Preamp On) F

Pagina 324

390 Chapter 2 Performance Verification Tests53. IF INPUT Accuracy (Option AYZ only)53. IF INPUT Accuracy (Option AYZ only)This test only applies to an

Pagina 325

Chapter 2 391Performance Verification Tests53. IF INPUT Accuracy (Option AYZ only)Figure 2-73 IF INPUT Accuracy Test SetupProcedureDetermining the IF

Pagina 326 - Ω tracking

392 Chapter 2 Performance Verification Tests53. IF INPUT Accuracy (Option AYZ only)6. On the analyzer, press Single and wait for the sweep to finish.7

Pagina 327

Chapter 2 393Performance Verification Tests53. IF INPUT Accuracy (Option AYZ only)18.Subtract the Input Power (step 14) from the Measured Power (step

Pagina 328 - if test

394 Chapter 2 Performance Verification Tests54. Comms Frequency Response (Option BAC or BAH)54. Comms Frequency Response (Option BAC or BAH)This test

Pagina 329 - →, Mkr→ CF

Chapter 2 395Performance Verification Tests54. Comms Frequency Response (Option BAC or BAH)Cable, Type-N, 183-cm (72-in) Cable, APC 3.5Termination, 50

Pagina 330 - 39. Residual Responses

396 Chapter 2 Performance Verification Tests54. Comms Frequency Response (Option BAC or BAH)is being measured. 6. Set the source frequency to 50 MHz a

Pagina 331

Chapter 2 397Performance Verification Tests54. Comms Frequency Response (Option BAC or BAH)810 MHz820 MHz830 MHz840 MHz850 MHz860 MHz880 MHz890 MHz900

Pagina 332

398 Chapter 2 Performance Verification Tests54. Comms Frequency Response (Option BAC or BAH)1790 MHz1800 MHz1810 MHz1820 MHz 1830 MHz1840 MHz1850 MHz1

Pagina 333 - ≥5 ms) to a fast sweep time

Chapter 2 399Performance Verification Tests54. Comms Frequency Response (Option BAC or BAH)Part 2: Measuring Frequency Response, 800 MHz to 1000 MHz,

Pagina 335 - E4411B (Option 1DN or 1DQ)

40 Chapter 2 Performance Verification TestsTests included in this section:Manual use of Agilent 8494G and Agilent 8496G Attenuators with the Agilent 1

Pagina 336

400 Chapter 2 Performance Verification Tests54. Comms Frequency Response (Option BAC or BAH)splitter. Connect the power splitter to the analyzer 50 Ω

Pagina 337 - Chapter 2 337

Chapter 2 401Performance Verification Tests54. Comms Frequency Response (Option BAC or BAH)17.Perform a peak search on the analyzer. Manually press Se

Pagina 338

402 Chapter 2 Performance Verification Tests54. Comms Frequency Response (Option BAC or BAH)Table 2-110 Frequency Response WorksheetTest:_______ Atte

Pagina 339 - Chapter 2 339

Chapter 2 403Performance Verification Tests54. Comms Frequency Response (Option BAC or BAH)1740 MHz1750 MHz1760 MHz1770 MHz1780 MHz1790 MHz1800 MHz181

Pagina 340 - (Option 1DN)

404 Chapter 2 Performance Verification Tests54. Comms Frequency Response (Option BAC or BAH)Part 3: Test Results 1. Repeat the following steps using t

Pagina 341 - −18 -2 2)

Chapter 2 405Performance Verification Tests54. Comms Frequency Response (Option BAC or BAH)b. Subtract the Minimum Response for the 1700 MHz to 2000 M

Pagina 342 - −0.15 = 0.08 − 0.23)

406 Chapter 2 Performance Verification Tests58. GSM Phase and Frequency Error (Options BAH and B7E)58. GSM Phase and Frequency Error (Options BAH and

Pagina 343 - (Option 1DN or 1DQ)

Chapter 2 407Performance Verification Tests58. GSM Phase and Frequency Error (Options BAH and B7E)1. Set up the equipment as shown in Figure 2-76.2. P

Pagina 344 - ≥100 kHz

408 Chapter 2 Performance Verification Tests58. GSM Phase and Frequency Error (Options BAH and B7E)14. Repeat step 7 through step 13 for each of the f

Pagina 345

Chapter 2 409Performance Verification Tests59. Comms Absolute Power Accuracy (Options BAC or BAH)59. Comms Absolute Power Accuracy (Options BAC or BAH

Pagina 346 - Ω RF Out

Chapter 2 41Performance Verification TestsTests included in this section:Table 2-2 Agilent 11713A Settings for Agilent 8494G and Agilent 8496G1 dB St

Pagina 347 - Ω tracking generators only

410 Chapter 2 Performance Verification Tests59. Comms Absolute Power Accuracy (Options BAC or BAH)Additional Equipment for Option BABAdapter, Type N (

Pagina 348 - DVM Readout= Level Flatness–

Chapter 2 411Performance Verification Tests59. Comms Absolute Power Accuracy (Options BAC or BAH)1. Press Preset on the analyzer. Press the Factory Pr

Pagina 349 - Chapter 2 349

412 Chapter 2 Performance Verification Tests59. Comms Absolute Power Accuracy (Options BAC or BAH)0.02 dB, then 1 dB Actual0dB is 0.02 dB. In this cas

Pagina 350

Chapter 2 413Performance Verification Tests59. Comms Absolute Power Accuracy (Options BAC or BAH)Figure 2-78 Amplitude Accuracy Test Setup 12.Set the

Pagina 351

414 Chapter 2 Performance Verification Tests59. Comms Absolute Power Accuracy (Options BAC or BAH)14.Set the 10 dB and 1 dB step attenuators to 0 dB.1

Pagina 352

Chapter 2 415Performance Verification Tests59. Comms Absolute Power Accuracy (Options BAC or BAH)Table 2-120 GSM Transmit Power (Option BAH), Preamp

Pagina 353

416 Chapter 2 Performance Verification Tests59. Comms Absolute Power Accuracy (Options BAC or BAH)19.Table 2-117 and Table 2-121 have two Amplitude Ac

Pagina 354 - −1.01 dB

Chapter 2 417Performance Verification Tests59. Comms Absolute Power Accuracy (Options BAC or BAH)numbers in parenthesis.25.Continue with Part 2.Table

Pagina 355

418 Chapter 2 Performance Verification Tests59. Comms Absolute Power Accuracy (Options BAC or BAH)Table 2-125 GSM Transmit Power Accuracy (Cellular B

Pagina 356 - Ω to 75 Ω Minimum loss pad

Chapter 2 419Performance Verification Tests59. Comms Absolute Power Accuracy (Options BAC or BAH)Part 2: cdmaOne Receive Channel Power (Preamp Off)Tab

Pagina 357 - −10 dBm. Press

42 Chapter 2 Performance Verification Tests1. 10 MHz Reference Output Accuracy1. 10 MHz Reference Output AccuracyThe setability is measured by changin

Pagina 358 - ⇑ (step-up key) to

420 Chapter 2 Performance Verification Tests59. Comms Absolute Power Accuracy (Options BAC or BAH)NOTE The external attenuators and cables are now par

Pagina 359

Chapter 2 421Performance Verification Tests59. Comms Absolute Power Accuracy (Options BAC or BAH)h. Calculate the amplitude accuracy as follows:7. Tab

Pagina 360

422 Chapter 2 Performance Verification Tests59. Comms Absolute Power Accuracy (Options BAC or BAH)to 1 GHz Cellular Band, and the 1.7 GHz to 2 GHz PCS

Pagina 361

Chapter 2 423Performance Verification Tests59. Comms Absolute Power Accuracy (Options BAC or BAH)Part 3: cdmaOne Receive Channel Power (Preamplifier O

Pagina 362 - ⇑ (step up key) to

424 Chapter 2 Performance Verification Tests59. Comms Absolute Power Accuracy (Options BAC or BAH)4. Calculate the actual total attenuation by adding

Pagina 363

Chapter 2 425Performance Verification Tests59. Comms Absolute Power Accuracy (Options BAC or BAH)f. Press Peak Search (or Search). The marker can stil

Pagina 364

426 Chapter 2 Performance Verification Tests59. Comms Absolute Power Accuracy (Options BAC or BAH)−90 0 0.24 39)−100 0 0.24 40)Table 2-134 cdmaOne Re

Pagina 365

4273 Performance Verification Test Records

Pagina 366 - Ω Option only)

428 Chapter 3 Performance Verification Test RecordsAgilent E4401B Performance Verification Test RecordAgilent E4401B Performance Verification Test Re

Pagina 367 - ±750 kHz

Chapter 3 429Performance Verification Test RecordsAgilent E4401B Performance Verification Test Record50 Ω Termination ___________ ___________ ______

Pagina 368

Chapter 2 43Performance Verification Tests1. 10 MHz Reference Output Accuracy10 s.d. Press Enter.e. On Channel 1, press 50 Ω/1 MΩ until the LED is lit

Pagina 369 - Response

430 Chapter 3 Performance Verification Test RecordsAgilent E4401B Performance Verification Test Record 1490 MHz 10 MHz 1489.91 MHz (2)________

Pagina 370

Chapter 3 431Performance Verification Test RecordsAgilent E4401B Performance Verification Test RecordOffset from 500 MHz signal 30 kHz to 230 kHz (1

Pagina 371 - −10 dBm. Press STORE

432 Chapter 3 Performance Verification Test RecordsAgilent E4401B Performance Verification Test Record–12 –0.40 dB (3)________ 0.40 dB ±0.075 dB–16

Pagina 372

Chapter 3 433Performance Verification Test RecordsAgilent E4401B Performance Verification Test Record–20 –0.4 dB (26)________ 0.4 dB ±0.078 dB–24 –0.

Pagina 373

434 Chapter 3 Performance Verification Test RecordsAgilent E4401B Performance Verification Test Record–24 –0.54 dB (48)________ 0.54 dB ±0.074 dB–28

Pagina 374

Chapter 3 435Performance Verification Test RecordsAgilent E4401B Performance Verification Test Record–12 –0.4 dB (70)________ 0.4 dB ±0.075 dB–16

Pagina 375 - Chapter 2 375

436 Chapter 3 Performance Verification Test RecordsAgilent E4401B Performance Verification Test Record–20 –2.0% (97)________ 2.0% ±0.063%Linear Fidel

Pagina 376

Chapter 3 437Performance Verification Test RecordsAgilent E4401B Performance Verification Test Record–56 –1.14 dB (116)________ 1.14 dB ±0.078 dB–60

Pagina 377

438 Chapter 3 Performance Verification Test RecordsAgilent E4401B Performance Verification Test Record–5 43.75 –0.3 dB (2)________ 0.3 dB ±0.100 dB–3

Pagina 378

Chapter 3 439Performance Verification Test RecordsAgilent E4401B Performance Verification Test Record–75 –26.25 –0.7 dB (21)________ 0.7 dB ±0.099 dB

Pagina 379 - ∆t markers

44 Chapter 2 Performance Verification Tests1. 10 MHz Reference Output Accuracythe maximum frequency change.16.Divide the maximum frequency change by t

Pagina 380 - 4. Press

440 Chapter 3 Performance Verification Test RecordsAgilent E4401B Performance Verification Test Record 200 Hz (Option 1DR only) –0.3 dB (12)________

Pagina 381 - −500 ns and −650 ns

Chapter 3 441Performance Verification Test RecordsAgilent E4401B Performance Verification Test Record –20 dBm input –0.54 dB (7)________ 0.54 dB

Pagina 382

442 Chapter 3 Performance Verification Test RecordsAgilent E4401B Performance Verification Test Record20 to 30° C50 Ω Input (Non-Option 1DP)(9 kHz to

Pagina 383

Chapter 3 443Performance Verification Test RecordsAgilent E4401B Performance Verification Test Record25. Frequency Response (Preamp On)(Option 1DS on

Pagina 384 - Ω Inputs only)

444 Chapter 3 Performance Verification Test RecordsAgilent E4401B Performance Verification Test Record Peak-to-Peak Response (3)________ 4.0 dB ±0.

Pagina 385

Chapter 3 445Performance Verification Test RecordsAgilent E4401B Performance Verification Test Record 53 MHz (1)________ 1.0 dB ±0.123 dB 50.004 MHz

Pagina 386 - 9. Record the marker delta (

446 Chapter 3 Performance Verification Test RecordsAgilent E4401B Performance Verification Test Record 50 Ω, 10 Hz RBW, Preamp On 400 kHz (16)_

Pagina 387 - AYZ only)

Chapter 3 447Performance Verification Test RecordsAgilent E4401B Performance Verification Test Record 1 MHz to 10 MHz (33)________ –99 dBmV ±1

Pagina 388

448 Chapter 3 Performance Verification Test RecordsAgilent E4401B Performance Verification Test RecordPower Sweep Accuracy (4)________ 1.8 dB ±0.008

Pagina 389

Chapter 3 449Performance Verification Test RecordsAgilent E4401B Performance Verification Test Record2nd Harmonic, ≥20 MHz(2)________ –25 dBc±2.6 dB3

Pagina 390

Chapter 2 45Performance Verification Tests2. 10 MHz High-Stability Frequency Reference Output Accuracy: Agilent E4401B, E4402B,E4404B, E4405B, and E44

Pagina 391

450 Chapter 3 Performance Verification Test RecordsAgilent E4402B Performance Verification Test RecordAgilent E4402B Performance Verification Test Re

Pagina 392 - ∆ Mkr1) frequency

Chapter 3 451Performance Verification Test RecordsAgilent E4402B Performance Verification Test Record10 dB Step Attenuator ___________ ___________ __

Pagina 393

452 Chapter 3 Performance Verification Test RecordsAgilent E4402B Performance Verification Test Record Center Freq Span 1500 MHz 10 MHz 14

Pagina 394

Chapter 3 453Performance Verification Test RecordsAgilent E4402B Performance Verification Test Record (Serial Prefix ≥ US4510, SG4510 or MY4510, Non

Pagina 395

454 Chapter 3 Performance Verification Test RecordsAgilent E4402B Performance Verification Test Record 1 kHz Res BW,(Non-Option 1D5) (1)_________ 15

Pagina 396 - 396 Chapter 2

Chapter 3 455Performance Verification Test RecordsAgilent E4402B Performance Verification Test Record–24 –0.50 dB (6)________ 0.50 dB ±0.074 dB–28

Pagina 397 - Chapter 2 397

456 Chapter 3 Performance Verification Test RecordsAgilent E4402B Performance Verification Test Record–32 –0.4 dB (29)________ 0.4 dB ±0.077 dB–36

Pagina 398 - 398 Chapter 2

Chapter 3 457Performance Verification Test RecordsAgilent E4402B Performance Verification Test Record–40 –0.70 dB (52)________ 0.70 dB ±0.081 dB–44

Pagina 399

458 Chapter 3 Performance Verification Test RecordsAgilent E4402B Performance Verification Test Record–28 –0.4 dB (74)________ 0.4 dB ±0.073 dB–32 –0

Pagina 400 - Ω Input using

Chapter 3 459Performance Verification Test RecordsAgilent E4402B Performance Verification Test Record–8 –2.0% (99)________ 2.0% ±0.064%–12 –2.0% (100

Pagina 401 - Chapter 2 401

46 Chapter 2 Performance Verification Tests2. 10 MHz High-Stability Frequency Reference Output Accuracy: Agilent E4401B, E4402B, E4404B, E4405B, and E

Pagina 402

460 Chapter 3 Performance Verification Test RecordsAgilent E4402B Performance Verification Test Record13. Input Attenuation Switching UncertaintyInpu

Pagina 403

Chapter 3 461Performance Verification Test RecordsAgilent E4402B Performance Verification Test Record–70 dBm –0.5 dB (7)________ 0.5 dB ±0.099 dB–80

Pagina 404

462 Chapter 3 Performance Verification Test RecordsAgilent E4402B Performance Verification Test Record–10 dBm –0.3 dB (25)________ 0.3 dB ±0.091 dB 0

Pagina 405

Chapter 3 463Performance Verification Test RecordsAgilent E4402B Performance Verification Test Record 3 Hz (Option 1DR and 1D5 only) Firmware Revisio

Pagina 406 - BAH and B7E)

464 Chapter 3 Performance Verification Test RecordsAgilent E4402B Performance Verification Test Record−40 dBm Reference Level –40 dBm input –0.54

Pagina 407

Chapter 3 465Performance Verification Test RecordsAgilent E4402B Performance Verification Test Record Minimum Response –0.46 dB (2)________ ±0.19 d

Pagina 408

466 Chapter 3 Performance Verification Test RecordsAgilent E4402B Performance Verification Test RecordOption UKBac CoupledBand 0(100 kHz to 3.0 GHz)

Pagina 409 - °C temperature

Chapter 3 467Performance Verification Test RecordsAgilent E4402B Performance Verification Test Record Peak-to-Peak Response (3)________ 1.52 dB ±0.

Pagina 410 - ≥–25 dBm

468 Chapter 3 Performance Verification Test RecordsAgilent E4402B Performance Verification Test RecordOption UKBac CoupledBand 0(100 kHz to 3.0 GHz)

Pagina 411

Chapter 3 469Performance Verification Test RecordsAgilent E4402B Performance Verification Test Record Minimum Response –1.5 dB (2)________ ±0.28 dB

Pagina 412 - 412 Chapter 2

Chapter 2 47Performance Verification Tests2. 10 MHz High-Stability Frequency Reference Output Accuracy: Agilent E4401B, E4402B,E4404B, E4405B, and E44

Pagina 413

470 Chapter 3 Performance Verification Test RecordsAgilent E4402B Performance Verification Test RecordBand 0A(800 MHz to 1.0 GHz)(serial number US394

Pagina 414 - <−25 dBm

Chapter 3 471Performance Verification Test RecordsAgilent E4402B Performance Verification Test Record33. Gain CompressionTest Frequency 53 MHz (1)__

Pagina 415 - ≤–20 dBm

472 Chapter 3 Performance Verification Test RecordsAgilent E4402B Performance Verification Test Record1kHz RBW, Preamp On,20 to 30° C 10 MHz

Pagina 416

Chapter 3 473Performance Verification Test RecordsAgilent E4402B Performance Verification Test Record–8 dB –0.5 dB (7)________ 0.5 dB ±0.008 dB–9 dB

Pagina 417 - 25.Continue with Part 2

474 Chapter 3 Performance Verification Test RecordsAgilent E4402B Performance Verification Test Record Positive Level Flatness, >1.5 GHz(7)_____

Pagina 418

Chapter 3 475Performance Verification Test RecordsAgilent E4402B Performance Verification Test Record50. Gate Delay Accuracy and Gate Length Accuracy

Pagina 419

476 Chapter 3 Performance Verification Test RecordsAgilent E4402B Performance Verification Test Record−45 dBm –0.70 dB (4)________ 0.70 dB ±0.152 dB−

Pagina 420

Chapter 3 477Performance Verification Test RecordsAgilent E4402B Performance Verification Test Record−30 dBm–0.75 dBa(21)________0.75 dBb±0.156 dB−40

Pagina 421 - +()– 40dB+

478 Chapter 3 Performance Verification Test RecordsAgilent E4402B Performance Verification Test Record−40 dBm –1.77 dB (35)________ 1.77 dB ±0.161 dB

Pagina 422 - 422 Chapter 2

Chapter 3 479Performance Verification Test RecordsAgilent E4403B Performance Verification Test RecordAgilent E4403B Performance Verification Test Rec

Pagina 423

48 Chapter 2 Performance Verification Tests3. Frequency Readout and Marker Frequency Count Accuracy: Agilent E4401B, E4402B, E4403B, and E4411B3. Freq

Pagina 424

480 Chapter 3 Performance Verification Test RecordsAgilent E4403B Performance Verification Test RecordMicrowave Spectrum Analyzer (Option 1DN only) _

Pagina 425 - ActualTotalAtten–RefAtten

Chapter 3 481Performance Verification Test RecordsAgilent E4403B Performance Verification Test Record6. Frequency Span Accuracy Span Start Fr

Pagina 426

482 Chapter 3 Performance Verification Test RecordsAgilent E4403B Performance Verification Test Record 20 ms –1.0% (2)________ 1.0% ±0.16% 100 ms

Pagina 427

Chapter 3 483Performance Verification Test RecordsAgilent E4403B Performance Verification Test Record–76 –1.06 dB (19)________ 1.06 dB ±0.083 dB–80

Pagina 428 - Ω Power Sensor

484 Chapter 3 Performance Verification Test RecordsAgilent E4403B Performance Verification Test RecordCumulative Log Fidelity,Res BW ≤300 Hz(Option 1

Pagina 429 - ±0.000070 ppm

Chapter 3 485Performance Verification Test RecordsAgilent E4403B Performance Verification Test Record–92 –1.22 dB (65)________ 1.22 dB ±0.100 dB–96

Pagina 430

486 Chapter 3 Performance Verification Test RecordsAgilent E4403B Performance Verification Test Record–80 –0.4 dB (87)________ 0.4 dB ±0.089 dBLinear

Pagina 431

Chapter 3 487Performance Verification Test RecordsAgilent E4403B Performance Verification Test Record–28 –0.72 dB (109)________ 0.72 dB ±0.073 dB–32

Pagina 432

488 Chapter 3 Performance Verification Test RecordsAgilent E4403B Performance Verification Test Record60 dB –0.70 dB (12)________ 0.70 dB ±0.100 dB65

Pagina 433

Chapter 3 489Performance Verification Test RecordsAgilent E4403B Performance Verification Test RecordResolution Bandwidth 3 kHz –0.3 dB (1)________ 0

Pagina 434

Chapter 2 49Performance Verification Tests3. Frequency Readout and Marker Frequency Count Accuracy: Agilent E4401B, E4402B,E4403B, and E4411B“Part 2:

Pagina 435

490 Chapter 3 Performance Verification Test RecordsAgilent E4403B Performance Verification Test Record –40 dBm input –0.6 dB (5)________ 0.6 dB ±

Pagina 436

Chapter 3 491Performance Verification Test RecordsAgilent E4403B Performance Verification Test Record23. Frequency Response Note: Enter data in the

Pagina 437

492 Chapter 3 Performance Verification Test RecordsAgilent E4403B Performance Verification Test Record36. Displayed Average Noise Level1 kHz RBW

Pagina 438

Chapter 3 493Performance Verification Test RecordsAgilent E4403B Performance Verification Test RecordIncremental Vernier Accuracy–2 dB –0.2 dB (10)__

Pagina 439

494 Chapter 3 Performance Verification Test RecordsAgilent E4403B Performance Verification Test Record46. Tracking Generator Harmonic Spurious Output

Pagina 440

Chapter 3 495Performance Verification Test RecordsAgilent E4404B Performance Verification Test RecordAgilent E4404B Performance Verification Test Rec

Pagina 441

496 Chapter 3 Performance Verification Test RecordsAgilent E4404B Performance Verification Test RecordPower Splitter ___________ ___________ ________

Pagina 442

Chapter 3 497Performance Verification Test RecordsAgilent E4404B Performance Verification Test Record 1500 MHz 1 MHz 1499.991 MHz(3)________ 1

Pagina 443

498 Chapter 3 Performance Verification Test RecordsAgilent E4404B Performance Verification Test Record7. Noise Sidebands Offset from 1 GHz signal (

Pagina 444

Chapter 3 499Performance Verification Test RecordsAgilent E4404B Performance Verification Test Record Option 120 Offset from 1 GHz signal 100

Pagina 445

Contents 51. CalibratingCalibration Cycle . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

Pagina 446

50 Chapter 2 Performance Verification Tests3. Frequency Readout and Marker Frequency Count Accuracy: Agilent E4401B, E4402B, E4403B, and E4411Bdisplay

Pagina 447

500 Chapter 3 Performance Verification Test RecordsAgilent E4404B Performance Verification Test Record 1 ms (Option AYX or B7D only)–1.0% (6)______

Pagina 448

Chapter 3 501Performance Verification Test RecordsAgilent E4404B Performance Verification Test Record–72 –0.80 dB (18)________ 0.80 dB ±0.084 dB–76

Pagina 449

502 Chapter 3 Performance Verification Test RecordsAgilent E4404B Performance Verification Test Record–80 –0.4 dB (41)________ 0.4 dB ±0.089 dBCumula

Pagina 450

Chapter 3 503Performance Verification Test RecordsAgilent E4404B Performance Verification Test Record–88 –1.18 dB (64)________ 1.18 dB ±0.085 dB–92

Pagina 451

504 Chapter 3 Performance Verification Test RecordsAgilent E4404B Performance Verification Test Record–76 –0.4 dB (86)________ 0.4 dB ±0.083

Pagina 452

Chapter 3 505Performance Verification Test RecordsAgilent E4404B Performance Verification Test Record–24 –0.66 dB (108)________ 0.66 dB ±0.074 dB–28

Pagina 453 - GHz signal

506 Chapter 3 Performance Verification Test RecordsAgilent E4404B Performance Verification Test Record55 dB –0.65 dB (11)________ 0.65 dB ±0.099 dB60

Pagina 454

Chapter 3 507Performance Verification Test RecordsAgilent E4404B Performance Verification Test RecordLog, Res BW ≤300 Hz(Option 1DR only)Reference Le

Pagina 455

508 Chapter 3 Performance Verification Test RecordsAgilent E4404B Performance Verification Test Record 10 kHz –0.3 dB (3)________ 0.3 dB ±0.048 dB 30

Pagina 456

Chapter 3 509Performance Verification Test RecordsAgilent E4404B Performance Verification Test Record0 dBm Reference Level 0 dBm input –0.54 dB (

Pagina 457

Chapter 2 51Performance Verification Tests4. Frequency Readout and Marker Frequency Count Accuracy: Agilent E4404B, E4405B,E4407B, and E4408B4. Freque

Pagina 458

510 Chapter 3 Performance Verification Test RecordsAgilent E4404B Performance Verification Test Record10 kHz 8.5 kHz (7)________ 11.5 kHz ±220 Hz3 kH

Pagina 459

Chapter 3 511Performance Verification Test RecordsAgilent E4404B Performance Verification Test Record Peak-to-Peak Response (6)________ 2.6 dB ±0.3

Pagina 460

512 Chapter 3 Performance Verification Test RecordsAgilent E4404B Performance Verification Test Record Maximum Response (19)________ 1.5 dB ±0.39 d

Pagina 461

Chapter 3 513Performance Verification Test RecordsAgilent E4404B Performance Verification Test Recordac CoupledBand 0(100 kHz to 3.0 GHz) Maximum R

Pagina 462

514 Chapter 3 Performance Verification Test RecordsAgilent E4404B Performance Verification Test Record Peak-to-Peak Response (21)________ 3.0 dB ±0

Pagina 463

Chapter 3 515Performance Verification Test RecordsAgilent E4404B Performance Verification Test Record0 to 55° CBand 0(1 MHz to 3.0 GHz) Maximum Res

Pagina 464

516 Chapter 3 Performance Verification Test RecordsAgilent E4404B Performance Verification Test Record 2.0 GHz 5600.0 MHz (5)________ –80 dBc ±.8

Pagina 465

Chapter 3 517Performance Verification Test RecordsAgilent E4404B Performance Verification Test Record 10 MHz to 1 GHz (1)________ –116 dBm ±

Pagina 466

518 Chapter 3 Performance Verification Test RecordsAgilent E4404B Performance Verification Test Record10 Hz RBW, Preamp On,20 to 30° C 10 MH

Pagina 467

Chapter 3 519Performance Verification Test RecordsAgilent E4404B Performance Verification Test Record–2 dB –0.2 dB (10)________ 0.2 dB ±0.008 dB–3 dB

Pagina 468

52 Chapter 2 Performance Verification Tests4. Frequency Readout and Marker Frequency Count Accuracy: Agilent E4404B, E4405B, E4407B, and E4408BProcedu

Pagina 469

520 Chapter 3 Performance Verification Test RecordsAgilent E4404B Performance Verification Test Record46. Tracking Generator Harmonic Spurious Output

Pagina 470

Chapter 3 521Performance Verification Test RecordsAgilent E4404B Performance Verification Test Record65 ms Gate Length 64.993 ms (4)________ 65.007 m

Pagina 471

522 Chapter 3 Performance Verification Test RecordsAgilent E4404B Performance Verification Test Record−25 dBm –0.74 dB (9)________ 0.74 dB ±0.161 dB−

Pagina 472

Chapter 3 523Performance Verification Test RecordsAgilent E4404B Performance Verification Test Record−40 dBm –1.46 dB (25)________ 1.46 dB ±0.161 dB−

Pagina 473

524 Chapter 3 Performance Verification Test RecordsAgilent E4404B Performance Verification Test Record−70 dBm –1.88 dB (43)________ 1.88 dB ±0.161 dB

Pagina 474

Chapter 3 525Performance Verification Test RecordsAgilent E4405B Performance Verification Test RecordAgilent E4405B Performance Verification Test Rec

Pagina 475

526 Chapter 3 Performance Verification Test RecordsAgilent E4405B Performance Verification Test RecordPower Splitter ___________ ___________ ________

Pagina 476

Chapter 3 527Performance Verification Test RecordsAgilent E4405B Performance Verification Test Record 1500 MHz 1 MHz 1499.991 MHz(3)________ 1

Pagina 477

528 Chapter 3 Performance Verification Test RecordsAgilent E4405B Performance Verification Test Record Center Freq Span 1500 MHz 20 MHz 149

Pagina 478

Chapter 3 529Performance Verification Test RecordsAgilent E4405B Performance Verification Test Record 20 kHz (2)________ –100 dBc/Hz ±2.44 dB

Pagina 479 - Test Record

Chapter 2 53Performance Verification Tests4. Frequency Readout and Marker Frequency Count Accuracy: Agilent E4404B, E4405B,E4407B, and E4408BPart 2: M

Pagina 480

530 Chapter 3 Performance Verification Test RecordsAgilent E4405B Performance Verification Test Record 10 MHz (4)________ –137 dBc/Hz ±2.44 dB9

Pagina 481

Chapter 3 531Performance Verification Test RecordsAgilent E4405B Performance Verification Test RecordCumulative Log Fidelity,Res BW ≥1 kHzdB from Ref

Pagina 482

532 Chapter 3 Performance Verification Test RecordsAgilent E4405B Performance Verification Test RecordIncremental Log Fidelity,Res BW ≥1 kHzdB from R

Pagina 483

Chapter 3 533Performance Verification Test RecordsAgilent E4405B Performance Verification Test Record–4 –0.34 dB (43)________ 0.34 dB ±0.082 dB–8 –0

Pagina 484

534 Chapter 3 Performance Verification Test RecordsAgilent E4405B Performance Verification Test RecordIncremental Log Fidelity,Res BW ≤300 Hz(Option

Pagina 485

Chapter 3 535Performance Verification Test RecordsAgilent E4405B Performance Verification Test Record–4 –2.0% (93)________ 2.0% ±0.064%–8 –2.0% (94)_

Pagina 486

536 Chapter 3 Performance Verification Test RecordsAgilent E4405B Performance Verification Test Record–40 –0.90 dB (112)________ 0.90 dB ±0.081 dB–44

Pagina 487

Chapter 3 537Performance Verification Test RecordsAgilent E4405B Performance Verification Test Record15. Reference Level AccuracyLog, Res BW ≥1 kHzRe

Pagina 488

538 Chapter 3 Performance Verification Test RecordsAgilent E4405B Performance Verification Test Record 0 dBm –0.3 dB (18)________ 0.3 dB ±0.100 dB–30

Pagina 489

Chapter 3 539Performance Verification Test RecordsAgilent E4405B Performance Verification Test Record 120 kHz –0.3 dB (6)________ 0.3 dB ±0.048 dB 30

Pagina 490

54 Chapter 2 Performance Verification Tests4. Frequency Readout and Marker Frequency Count Accuracy: Agilent E4404B, E4405B, E4407B, and E4408Bverific

Pagina 491

540 Chapter 3 Performance Verification Test RecordsAgilent E4405B Performance Verification Test Record –10 dBm input –0.54 dB (2)________ 0.54 dB

Pagina 492

Chapter 3 541Performance Verification Test RecordsAgilent E4405B Performance Verification Test Record1 kHz 850 Hz (9)________ 1.15 kHz ±22 Hz120 kHz

Pagina 493

542 Chapter 3 Performance Verification Test RecordsAgilent E4405B Performance Verification Test Recorddc CoupledBand 2(6.7 GHz to 13.2 GHz) Maximum

Pagina 494

Chapter 3 543Performance Verification Test RecordsAgilent E4405B Performance Verification Test Record Minimum Response –0.5 dB (17)________ ±0.11 d

Pagina 495

544 Chapter 3 Performance Verification Test RecordsAgilent E4405B Performance Verification Test RecordOption UKBdc CoupledBand 0(100 Hz to 3.0 GHz)

Pagina 496

Chapter 3 545Performance Verification Test RecordsAgilent E4405B Performance Verification Test Recordac CoupledBand 0A(800 MHz to 1.0 GHz)(serial num

Pagina 497

546 Chapter 3 Performance Verification Test RecordsAgilent E4405B Performance Verification Test Record Peak-to-Peak Response (24)________ 4.0 dB ±0

Pagina 498

Chapter 3 547Performance Verification Test RecordsAgilent E4405B Performance Verification Test Record0 to 55° CBand 0(1 MHz to 3.0 GHz) Maximum Res

Pagina 499

548 Chapter 3 Performance Verification Test RecordsAgilent E4405B Performance Verification Test Record 2.0 GHz 5600.0 MHz (5)________ –80 dBc ±.8

Pagina 500

Chapter 3 549Performance Verification Test RecordsAgilent E4405B Performance Verification Test Record 1403 MHz (3)________ 1.0 dB ±0.123 dB 2503 MHz

Pagina 501

Chapter 2 55Performance Verification Tests5. Frequency Span Accuracy: Agilent E4401B and E4411B5. Frequency Span Accuracy: Agilent E4401B and E4411BFo

Pagina 502

550 Chapter 3 Performance Verification Test RecordsAgilent E4405B Performance Verification Test Record10 Hz RBW, Preamp On,0 to 55° C 10 MHz

Pagina 503

Chapter 3 551Performance Verification Test RecordsAgilent E4405B Performance Verification Test Record–2 dB –0.5 dB (2)________ 0.5 dB ±0.008 dB–3 dB

Pagina 504

552 Chapter 3 Performance Verification Test RecordsAgilent E4405B Performance Verification Test Record Negative Level Flatness, 1 MHz to 10 MHz–3.0

Pagina 505

Chapter 3 553Performance Verification Test RecordsAgilent E4405B Performance Verification Test Record 9 kHz to 2.9 GHz (1)________ –16 dBm ±2.58 dB 2

Pagina 506

554 Chapter 3 Performance Verification Test RecordsAgilent E4405B Performance Verification Test Record−5 dBm –0.82 dB (2)________ 0.82 dB ±0.161 dB−2

Pagina 507

Chapter 3 555Performance Verification Test RecordsAgilent E4405B Performance Verification Test Record−20 dBm –0.77 dB (20)________ 0.77 dB ±0.161 dB−

Pagina 508

556 Chapter 3 Performance Verification Test RecordsAgilent E4405B Performance Verification Test Record−40 dBm –1.88 dB (35)________ 1.88 dB ±0.161 dB

Pagina 509

Chapter 3 557Performance Verification Test RecordsAgilent E4407B Performance Verification Test RecordAgilent E4407B Performance Verification Test Rec

Pagina 510

558 Chapter 3 Performance Verification Test RecordsAgilent E4407B Performance Verification Test RecordPower Splitter ___________ ___________ ________

Pagina 511

Chapter 3 559Performance Verification Test RecordsAgilent E4407B Performance Verification Test Record 1500 MHz 1 MHz 1499.991 MHz(3)________ 1

Pagina 512

56 Chapter 2 Performance Verification Tests5. Frequency Span Accuracy: Agilent E4401B and E4411BFigure 2-5 Frequency Span Readout Accuracy Test Setu

Pagina 513

560 Chapter 3 Performance Verification Test RecordsAgilent E4407B Performance Verification Test Record 1500 MHz 1 MHz 1499.999998 MHz(17)_____

Pagina 514

Chapter 3 561Performance Verification Test RecordsAgilent E4407B Performance Verification Test Record 9000 MHz 1 MHz 8999.999999 MHz(21)______

Pagina 515

562 Chapter 3 Performance Verification Test RecordsAgilent E4407B Performance Verification Test Record (Serial Prefix ≥ US4510, SG4510 or MY4510, Op

Pagina 516

Chapter 3 563Performance Verification Test RecordsAgilent E4407B Performance Verification Test Record9. System-Related Sidebands Offset from 500 MHz

Pagina 517

564 Chapter 3 Performance Verification Test RecordsAgilent E4407B Performance Verification Test RecordCumulative Log Fidelity,Res BW ≥1 kHzdB from Re

Pagina 518

Chapter 3 565Performance Verification Test RecordsAgilent E4407B Performance Verification Test RecordIncremental Log Fidelity,Res BW ≥1 kHzdB from Re

Pagina 519

566 Chapter 3 Performance Verification Test RecordsAgilent E4407B Performance Verification Test Record–4 –0.34 dB (43)________ 0.34 dB ±0.082 dB–8 –0

Pagina 520

Chapter 3 567Performance Verification Test RecordsAgilent E4407B Performance Verification Test RecordIncremental Log Fidelity,Res BW ≤300 Hz(Option 1

Pagina 521

568 Chapter 3 Performance Verification Test RecordsAgilent E4407B Performance Verification Test Record–4 –2.0% (93)________ 2.0% ±0.064%–8 –2.0% (94)

Pagina 522

Chapter 3 569Performance Verification Test RecordsAgilent E4407B Performance Verification Test Record–40 –0.90 dB (112)________ 0.90 dB ±0.081 dB–44

Pagina 523

Chapter 2 57Performance Verification Tests5. Frequency Span Accuracy: Agilent E4401B and E4411BCW, 100 MHzPOWER LEVEL, –5 dBm3. Set the synthesized si

Pagina 524

570 Chapter 3 Performance Verification Test RecordsAgilent E4407B Performance Verification Test Record15. Reference Level AccuracyLog, Res BW ≥1 kHzR

Pagina 525

Chapter 3 571Performance Verification Test RecordsAgilent E4407B Performance Verification Test Record 0 dBm –0.3 dB (18)________ 0.3 dB ±0.100 dB–30

Pagina 526

572 Chapter 3 Performance Verification Test RecordsAgilent E4407B Performance Verification Test Record 120 kHz –0.3 dB (6)________ 0.3 dB ±0.048 dB 3

Pagina 527

Chapter 3 573Performance Verification Test RecordsAgilent E4407B Performance Verification Test Record –10 dBm input –0.54 dB (2)________ 0.54 dB

Pagina 528

574 Chapter 3 Performance Verification Test RecordsAgilent E4407B Performance Verification Test Record1 kHz 850 Hz (9)________ 1.15 kHz ±22 Hz120 kHz

Pagina 529

Chapter 3 575Performance Verification Test RecordsAgilent E4407B Performance Verification Test Record Peak-to-Peak Response (6)________ 0.92 dB ±0.

Pagina 530

576 Chapter 3 Performance Verification Test RecordsAgilent E4407B Performance Verification Test Record Minimum Response –2.0 dB (17)________ ±0.59

Pagina 531

Chapter 3 577Performance Verification Test RecordsAgilent E4407B Performance Verification Test Recordac CoupledBand 1(3.0 GHz to 6.7 GHz) Maximum R

Pagina 532

578 Chapter 3 Performance Verification Test RecordsAgilent E4407B Performance Verification Test RecordOption UKBdc CoupledBand 0(100 Hz to 3.0 GHz)

Pagina 533

Chapter 3 579Performance Verification Test RecordsAgilent E4407B Performance Verification Test Record Maximum Response (10)________ 2.5 dB ±0.39 dB

Pagina 534

58 Chapter 2 Performance Verification Tests5. Frequency Span Accuracy: Agilent E4401B and E4411B10 110 20 100 2)10010 10.1 10.01 10.09 3)0.1800 900 81

Pagina 535

580 Chapter 3 Performance Verification Test RecordsAgilent E4407B Performance Verification Test RecordOption UKBac CoupledBand 0A(800 MHz to 1.0 GHz)

Pagina 536

Chapter 3 581Performance Verification Test RecordsAgilent E4407B Performance Verification Test Record Minimum Response –3.0 dB (32)________ ±0.40 d

Pagina 537

582 Chapter 3 Performance Verification Test RecordsAgilent E4407B Performance Verification Test RecordNon-Option UKBdc CoupledBand 0B(1.7 GHz to 2.0

Pagina 538

Chapter 3 583Performance Verification Test RecordsAgilent E4407B Performance Verification Test RecordOption UKBac CoupledBand 0B(1.7 GHz to 2.0 GHz)(

Pagina 539

584 Chapter 3 Performance Verification Test RecordsAgilent E4407B Performance Verification Test RecordNon-Option UKBdc CoupledBand 0B(1.7 GHz to 2.0

Pagina 540

Chapter 3 585Performance Verification Test RecordsAgilent E4407B Performance Verification Test RecordOption UKBac CoupledBand 0B(1.7 GHz to 2.0 GHz)(

Pagina 541

586 Chapter 3 Performance Verification Test RecordsAgilent E4407B Performance Verification Test Record 15.0 GHz 18830.35 MHz (17)________ –65 dBc ±

Pagina 542

Chapter 3 587Performance Verification Test RecordsAgilent E4407B Performance Verification Test Record 14003 MHz (7)________ 1.0 dB ±0.123 dB38. Disp

Pagina 543

588 Chapter 3 Performance Verification Test RecordsAgilent E4407B Performance Verification Test Record 10 MHz to 1 GHz (18)________ –150 dBm

Pagina 544

Chapter 3 589Performance Verification Test RecordsAgilent E4407B Performance Verification Test Record–5 dB –0.5 dB (4)________ 0.5 dB ±0.008 dB–6 dB

Pagina 545

Chapter 2 59Performance Verification Tests6. Frequency Span Accuracy: Agilent E4402B, E4403B, E4404B, E4405B, E4407B, and E4408B6. Frequency Span Accu

Pagina 546

590 Chapter 3 Performance Verification Test RecordsAgilent E4407B Performance Verification Test Record Positive Level Flatness, >10 MHz to 1.5 G

Pagina 547

Chapter 3 591Performance Verification Test RecordsAgilent E4407B Performance Verification Test Record50. Gate Delay and Gate Length Accuracy (Option

Pagina 548

592 Chapter 3 Performance Verification Test RecordsAgilent E4407B Performance Verification Test Record6.9 GHz 13.0 dBm (13)________ 17.5 dBm ±0.16 dB

Pagina 549

Chapter 3 593Performance Verification Test RecordsAgilent E4407B Performance Verification Test Record58. GSM Phase and Frequency Error (Option BAH an

Pagina 550

594 Chapter 3 Performance Verification Test RecordsAgilent E4407B Performance Verification Test RecordGSM Transmit Power Accuracy (Option BAH only)GS

Pagina 551

Chapter 3 595Performance Verification Test RecordsAgilent E4407B Performance Verification Test Record−85 dBm –1.56 dB (29)________ 1.56 dB ±0.152 dBP

Pagina 552

596 Chapter 3 Performance Verification Test RecordsAgilent E4408B Performance Verification Test RecordAgilent E4408B Performance Verification Test Re

Pagina 553

Chapter 3 597Performance Verification Test RecordsAgilent E4408B Performance Verification Test Record1 dB Step Attenuator ___________ ___________ ___

Pagina 554

598 Chapter 3 Performance Verification Test RecordsAgilent E4408B Performance Verification Test Record 16000 MHz 20 MHz 15999.83 MHz(10)______

Pagina 555

Chapter 3 599Performance Verification Test RecordsAgilent E4408B Performance Verification Test Record 16000 MHz 1 MHz 20999.999995 MHz(23)____

Pagina 556

6 ContentsE4405B, E4407B, and E4408B . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .13619. Overall Absol

Pagina 557

60 Chapter 2 Performance Verification Tests6. Frequency Span Accuracy: Agilent E4402B, E4403B, E4404B, E4405B, E4407B, and E4408BFigure 2-6 Frequenc

Pagina 558

600 Chapter 3 Performance Verification Test RecordsAgilent E4408B Performance Verification Test Record Span Start Freq 3000 MHz 0 Hz 2370

Pagina 559

Chapter 3 601Performance Verification Test RecordsAgilent E4408B Performance Verification Test Record 1 s –1.0% (4)________ ±1.0% ±0.16% 10 s –1.

Pagina 560

602 Chapter 3 Performance Verification Test RecordsAgilent E4408B Performance Verification Test Record–84 –1.14 dB (21)________ 1.14 dB ±0.086 dBIn

Pagina 561

Chapter 3 603Performance Verification Test RecordsAgilent E4408B Performance Verification Test RecordCumulative Log Fidelity,Res BW ≤300 Hz(Option 1D

Pagina 562

604 Chapter 3 Performance Verification Test RecordsAgilent E4408B Performance Verification Test Record–92 –1.22 dB (65)________ 1.22 dB ±0.100 dB–96

Pagina 563

Chapter 3 605Performance Verification Test RecordsAgilent E4408B Performance Verification Test Record–80 –0.4 dB (87)________ 0.4 dB ±0.089 dBLinear

Pagina 564

606 Chapter 3 Performance Verification Test RecordsAgilent E4408B Performance Verification Test Record–28 –0.72 dB (109)________ 0.72 dB ±0.073 dB–32

Pagina 565

Chapter 3 607Performance Verification Test RecordsAgilent E4408B Performance Verification Test Record60 dB –0.70 dB (12)________ 0.70 dB ±0.100 dB65

Pagina 566

608 Chapter 3 Performance Verification Test RecordsAgilent E4408B Performance Verification Test RecordResolution Bandwidth 3 kHz –0.3 dB (1)________

Pagina 567

Chapter 3 609Performance Verification Test RecordsAgilent E4408B Performance Verification Test Record –40 dBm input –0.6 dB (5)________ 0.6 dB ±

Pagina 568

Chapter 2 61Performance Verification Tests6. Frequency Span Accuracy: Agilent E4402B, E4403B, E4404B, E4405B, E4407B, and E4408BPOWER LEVEL, –5 dBm3.

Pagina 569

610 Chapter 3 Performance Verification Test RecordsAgilent E4408B Performance Verification Test Record24. Frequency Response Note: Enter data in the

Pagina 570

Chapter 3 611Performance Verification Test RecordsAgilent E4408B Performance Verification Test Record0 to 55° CBand 0(9 kHz to 3.0 GHz Maximum Resp

Pagina 571

612 Chapter 3 Performance Verification Test RecordsAgilent E4408B Performance Verification Test Record 2.0 GHz 2042.8 MHz (1)________ –65 dBc ±.8

Pagina 572

Chapter 3 613Performance Verification Test RecordsAgilent E4408B Performance Verification Test Record 8 GHz TOI 5 dBm (4)________ ±0.93 dB 300 MHz

Pagina 573

614 Chapter 3 Performance Verification Test RecordsAgilent E4408B Performance Verification Test Record 2 GHz to 3 GHz (13)________ –120 dBm

Pagina 574

Chapter 3 615Performance Verification Test RecordsAgilent E4408B Performance Verification Test Record–6 dB –0.2 dB (13)________ 0.2 dB ±0.008 dB–7 dB

Pagina 575

616 Chapter 3 Performance Verification Test RecordsAgilent E4408B Performance Verification Test Record3rd Harmonic, <20 kHz(3)________ –15 dBc±2.6

Pagina 576

Chapter 3 617Performance Verification Test RecordsAgilent E4411B Performance Verification Test RecordAgilent E4411B Performance Verification Test Rec

Pagina 577

618 Chapter 3 Performance Verification Test RecordsAgilent E4411B Performance Verification Test Record50Ω Termination ___________ ___________ ______

Pagina 578

Chapter 3 619Performance Verification Test RecordsAgilent E4411B Performance Verification Test Record 1490 MHz 1 MHz 1489.999998 MHz(5)_______

Pagina 579

62 Chapter 2 Performance Verification Tests6. Frequency Span Accuracy: Agilent E4402B, E4403B, E4404B, E4405B, E4407B, and E4408BTable 2-9 Frequency

Pagina 580

620 Chapter 3 Performance Verification Test RecordsAgilent E4411B Performance Verification Test Record Sweep Time 5 ms –1.0% (1)________ 1.0%±0.

Pagina 581

Chapter 3 621Performance Verification Test RecordsAgilent E4411B Performance Verification Test Record–68 –0.98 dB (17)________ 0.98 dB ±0.080 dB–72

Pagina 582

622 Chapter 3 Performance Verification Test RecordsAgilent E4411B Performance Verification Test Record–76 –0.4 dB (40)________ 0.4 dB ±0.083

Pagina 583

Chapter 3 623Performance Verification Test RecordsAgilent E4411B Performance Verification Test Record–84 –1.14 dB (63)________ 1.14 dB ±0.086 dB–88

Pagina 584

624 Chapter 3 Performance Verification Test RecordsAgilent E4411B Performance Verification Test Record–72 –0.4 dB (85)________ 0.4 dB ±0.084 dB–76

Pagina 585

Chapter 3 625Performance Verification Test RecordsAgilent E4411B Performance Verification Test Record–20 –0.60 dB (107)________ 0.60 dB ±0.078 dB–24

Pagina 586

626 Chapter 3 Performance Verification Test RecordsAgilent E4411B Performance Verification Test Record50 dB –0.60 dB (10)________ 0.60 dB ±0.095 dB5

Pagina 587

Chapter 3 627Performance Verification Test RecordsAgilent E4411B Performance Verification Test Record16. Resolution Bandwidth Switching UncertaintyRe

Pagina 588

628 Chapter 3 Performance Verification Test RecordsAgilent E4411B Performance Verification Test Record –30 dBm input –0.6 dB (4)________ 0.6 dB

Pagina 589

Chapter 3 629Performance Verification Test RecordsAgilent E4411B Performance Verification Test Record9 kHz 7.2 kHz (11)________ 10.8 kHz ±163 Hz22. F

Pagina 590

Chapter 2 63Performance Verification Tests7. Noise Sidebands7. Noise SidebandsA 1 GHz CW signal is applied to the input of the analyzer. The marker fu

Pagina 591

630 Chapter 3 Performance Verification Test RecordsAgilent E4411B Performance Verification Test Record Minimum Response –1.0 dB (2)________ ±0.11 d

Pagina 592

Chapter 3 631Performance Verification Test RecordsAgilent E4411B Performance Verification Test Record 50 Ω, 100 Hz RBW 400 kHz (11)________ –12

Pagina 593

632 Chapter 3 Performance Verification Test RecordsAgilent E4411B Performance Verification Test RecordNegative Vernier Accuracy–0.75 dB (3)________ ±

Pagina 594

Chapter 3 633Performance Verification Test RecordsAgilent E4411B Performance Verification Test Record Positive Level Flatness, >10 MHz(5)_______

Pagina 595

634 Chapter 3 Performance Verification Test RecordsAgilent E4411B Performance Verification Test Record

Pagina 596

6354 If You Have a ProblemThis chapter includes information on how to check for a problem with your Agilent ESA Series Spectrum Analyzer, and how to r

Pagina 597

636 Chapter 4 If You Have a ProblemWhat You’ll Find in This ChapterWhat You’ll Find in This Chapter This chapter includes information on how to check

Pagina 598

Chapter 4 637If You Have a ProblemBefore You Call Agilent TechnologiesBefore You Call Agilent Technologies Check the Basics A problem can often be res

Pagina 599

638 Chapter 4 If You Have a ProblemBefore You Call Agilent Technologieschapter.❏ If the equipment to perform the performance verification tests is not

Pagina 600

Chapter 4 639If You Have a ProblemBefore You Call Agilent TechnologiesTable 4-1 Agilent Technologies Sales and Service OfficesUNITED STATESInstrument

Pagina 601

64 Chapter 2 Performance Verification Tests7. Noise SidebandsProcedure This performance test consists of five parts: Part 1: Noise Sideband Suppressio

Pagina 602

640 Chapter 4 If You Have a ProblemHow to Return Your Analyzer for ServiceHow to Return Your Analyzer for Service Service Tag If you are returning you

Pagina 603

Chapter 4 641If You Have a ProblemHow to Return Your Analyzer for ServiceFigure 4-1

Pagina 604

642 Chapter 4 If You Have a ProblemHow to Return Your Analyzer for ServiceOther Packaging CAUTION Analyzer damage can result from using packaging mate

Pagina 605

Chapter 2 65Performance Verification Tests7. Noise SidebandsDet/Demod, Detector, SampleSweep, Sweep Time, 5 secSingle Wait for the completion of a swe

Pagina 606

66 Chapter 2 Performance Verification Tests7. Noise Sidebands4. Press the following analyzer keys to measure the carrier amplitude:Peak Search (or Sea

Pagina 607

Chapter 2 67Performance Verification Tests7. Noise SidebandsSweep, Sweep Time, 5 secSingleWait for the completion of a sweep, then press Peak Search (

Pagina 608

68 Chapter 2 Performance Verification Tests7. Noise SidebandsFREQUENCY, ↓, ↓Single Record the marker amplitude noise reading in Table 2-10 as the Lowe

Pagina 609

Chapter 2 69Performance Verification Tests7. Noise Sidebandsverification test record in locations 1, 2, 3 and 4 respectively.6. This completes the Noi

Pagina 610

Contents 742. Tracking Generator Absolute Amplitude and Vernier Accuracy: Agilent E4402B, E4403B, E4404B, E4405B, E4407B, and E4408B (Option 1DN) . .

Pagina 611

70 Chapter 2 Performance Verification Tests8. Noise Sidebands - Wide Offsets: Agilent E4402B, E4404B, E4405B and E4407B8. Noise Sidebands - Wide Offse

Pagina 612

Chapter 2 71Performance Verification Tests8. Noise Sidebands - Wide Offsets: Agilent E4402B, E4404B, E4405B and E4407Bnoise sideband suppression. Part

Pagina 613

72 Chapter 2 Performance Verification Tests8. Noise Sidebands - Wide Offsets: Agilent E4402B, E4404B, E4405B and E4407BNOTE A spur may exist at 100 kH

Pagina 614

Chapter 2 73Performance Verification Tests8. Noise Sidebands - Wide Offsets: Agilent E4402B, E4404B, E4405B and E4407BPart 3: Noise Sideband Suppressi

Pagina 615

74 Chapter 2 Performance Verification Tests8. Noise Sidebands - Wide Offsets: Agilent E4402B, E4404B, E4405B and E4407Breading in Table 2-11 as the Lo

Pagina 616

Chapter 2 75Performance Verification Tests9. System-Related Sidebands9. System-Related SidebandsA 500 MHz CW signal is applied to the input of the ana

Pagina 617 - Ω Input (No Option 1DP)

76 Chapter 2 Performance Verification Tests9. System-Related Sidebandsa. Set the synthesized signal generator controls as follows: FREQUENCY, 500 MHz

Pagina 618

Chapter 2 77Performance Verification Tests9. System-Related SidebandsSingle. Wait for the completion of a new sweep, then press Peak Search (or Search

Pagina 619

78 Chapter 2 Performance Verification Tests10. Residual FM10. Residual FMThis test measures the inherent short-term instability of the analyzer LO sys

Pagina 620

Chapter 2 79Performance Verification Tests10. Residual FMCAUTION Use only 75 Ω cables, connectors, or adapters on instruments with 75 Ω connectors, or

Pagina 621

8 ContentsAgilent E4403B Performance Verification Test Record . . . . . . . . . . . . . . . . . . . . . . . . . . .479Agilent E4404B Performance Ver

Pagina 622

80 Chapter 2 Performance Verification Tests10. Residual FMdelta (∆ Mkr1) amplitude reads −8dB ±0.3 dB. 7. Press Delta, then rotate the knob counterclo

Pagina 623

Chapter 2 81Performance Verification Tests10. Residual FMstep 8 by the Deviation recorded in step 12. Record this value as Test Record entry 1 (Residu

Pagina 624

82 Chapter 2 Performance Verification Tests10. Residual FMreference level. If it is not, press Sweep, Sweep (Cont), FREQUENCY, and use the knob to pla

Pagina 625

Chapter 2 83Performance Verification Tests11. Sweep Time Accuracy11. Sweep Time AccuracyThis test uses a function generator to amplitude-modulate a 50

Pagina 626

84 Chapter 2 Performance Verification Tests11. Sweep Time Accuracyconnectors, or the connectors will be damaged.Procedure 1. Set the synthesized signa

Pagina 627

Chapter 2 85Performance Verification Tests11. Sweep Time Accuracyaccuracy as follows:NOTE The sweep time accuracy is defined as a percentage of the in

Pagina 628

86 Chapter 2 Performance Verification Tests12. Display Scale Fidelity12. Display Scale FidelityA 50 MHz CW signal is applied to the input of the analy

Pagina 629

Chapter 2 87Performance Verification Tests12. Display Scale FidelityFigure 2-12 Scale Fidelity Test Setup CAUTION Use only 75 Ω cables, connectors, a

Pagina 630

88 Chapter 2 Performance Verification Tests12. Display Scale FidelityIf the calibration data does not indicate an actual attenuation value for the 0 d

Pagina 631

Chapter 2 89Performance Verification Tests12. Display Scale Fidelity6. Press Peak Search (or Search) on the analyzer.7. Adjust the synthesized signal

Pagina 632

91Calibrating

Pagina 633

90 Chapter 2 Performance Verification Tests12. Display Scale FidelityCalculate Incremental Log Fidelity1. Calculate the Incremental Log Fidelity Error

Pagina 634 - 634 Chapter 3

Chapter 2 91Performance Verification Tests12. Display Scale FidelityExample Calculation for ILFE at –20 dB from REFLVL setting:2. Record the result in

Pagina 635 - If You Have a Problem

92 Chapter 2 Performance Verification Tests12. Display Scale FidelityBW/Avg, Average (On)5, Enter3. Press Peak Search (or Search) on the analyzer and

Pagina 636 - 636 Chapter 4

Chapter 2 93Performance Verification Tests12. Display Scale FidelityCalculate incremental log fidelity:Calculate the Incremental Log Fidelity Error (I

Pagina 637 - Check the Basics

94 Chapter 2 Performance Verification Tests12. Display Scale FidelitySPAN, 10 kHzBW/Avg, 3 kHzAMPLITUDE, Scale Type (Lin)Input/Output (or Input), Inpu

Pagina 638 - Service Options

Chapter 2 95Performance Verification Tests12. Display Scale FidelityTable 2-15.11.Calculate the Linear Fidelity Error (LFE) as a percentage of referen

Pagina 639 - Chapter 4 639

96 Chapter 2 Performance Verification Tests12. Display Scale FidelitySetup for linear scale measurement:1. Set the following parameters on the analyze

Pagina 640 - Original Packaging

Chapter 2 97Performance Verification Tests12. Display Scale Fidelity10.If the analyzer has a 75W Input press:Input/Output (or Input)Input Z Corr (75)Z

Pagina 641 - Figure 4-1

98 Chapter 2 Performance Verification Tests12. Display Scale Fidelity16. Press SPAN, Zero Span on the analyzer.17.Adjust the synthesized signal genera

Pagina 642 - Other Packaging

Chapter 2 99Performance Verification Tests12. Display Scale FidelityPost-Test Instrument Restoration 25. Remove the RF cable from the analyzer input