Agilent Technologies E8364C Manual de usuario Pagina 41

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 65
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 40
40
E8361C
Test port input continued
Description Specification Option UNL Supplemental
Standard or Option 014 and Option 014 information
Third Order Intercept
1
– Tone spacing from 100 kHz to 5 MHz
Typical:
10 to 500 MHz +30 dBm
500 MHz to 24 GHz +24 dBm
24 to 40 GHz +23 dBm
40 to 50 GHz +24 dBm
50 to 67 GHz +26 dBm
Third Order Intercept
1
– Tone spacing from 5 to 20 MHz
Typical:
10 to 500 MHz
Not applicable
500 MHz to 24 GHz +20 dBm
24 to 40 GHz +20 dBm
40 to 50 GHz +22 dBm
50 to 67 GHz +24 dBm
Third Order Intercept
1
– Tone spacing from 20 to 50 MHz
Typical:
10 to 500 MHz
Not applicable
500 MHz to 24 GHz +26 dBm
24 to 40 GHz +24 dBm
40 to 50 GHz +25 dBm
50 to 67 GHz +27 dBm
System compression level – at maximum leveled output power
See Dynamic Accuracy Chart
Trace noise magnitude
10 to 45 MHz
2
< 0.150 dB rms < 0.150 dB rms
45 to 500 MHz
4, 5
< 0.010 dB rms < 0.010 dB rms 1 kHz IF bandwidth
500 MHz to 24 GHz < 0.006 dB rms < 0.006 dB rms ratio measurement, nominal
24 to 67 GHz < 0.006 dB rms < 0.009 dB rms power at test port
67 to 70 GHz
2
< 0.006 dB rms < 0.009 dB rms
Trace noise magnitude
2
– Option 080 enabled
3
10 to 45 MHz
< 0.150 dB rms < 0.150 dB rms
45 to 500 MHz
4, 5
< 0.010 dB rms < 0.010 dB rms 1 kHz IF bandwidth
500 MHz to 24 GHz < 0.006 dB rms < 0.006 dB rms ratio measurement, nominal
24 to 67 GHz < 0.009 dB rms < 0.012 dB rms power at test port
67 to 70 GHz < 0.009 dB rms < 0.012 dB rms
Trace noise phase
10 to 45 MHz
2
< 0.800° rms < 0.800° rms
45 to 500 MHz
5
< 0.100° rms < 0.100° rms 1 kHz IF bandwidth
500 MHz to 24 GHz < 0.060° rms < 0.060° rms ratio measurement, nominal
24 to 67 GHz < 0.100° rms < 0.100° rms power at test port
67 to 70 GHz
2
< 0.100° rms < 0.100° rms
Trace noise phase
2
– Option 080 enabled
3
10 to 45 MHz
< 0.800° rms < 0.800° rms 1 kHz IF bandwidth
45 to 500 MHz
5
< 0.100° rms < 0.100° rms ratio measurement, nominal
500 MHz to 24 GHz < 0.060° rms < 0.060° rms power at test port
24 to 67 GHz < 0.100° rms < 0.100° rms
67 to 70 GHz < 0.100° rms < 0.100° rms
1. Third order intercept is a typical specification that applies while the network
analyzer receiver is in its linear range.
2. Typical performance.
3. 0 Hz offset.
4. Trace noise magnitude may be degraded to 20 mdB rms at harmonic frequencies
of the first IF (8.33 MHz) below 80 MHz.
5. Specified value is for worst-case noise floor at 45 MHz.