Interface Capabilities
The baud rate, stop bits, parity, handshake protocol, and data bits must be
configured exactly the same for both the controller and the oscilloscope to
properly communicate over the RS-232-C bus. The oscilloscope’s RS-232-C
interface capabilities are as follows:
Baud Rate: 1200, 2400, 9600, or 19,200
Stop Bits: 1
Parity: None
Protocol: DTR or XON/XOFF
Data Bits: 8
Protocol
DTR (Data Terminal Ready) With a three-wire interface, selecting
DTR for the handshake protocol does not allow the sending or receiving
device to control data flow. No control over the data flow increases the
possibility of missing data or transferring incomplete data.
With an extended hardwire interface, selecting DTR allows a hardware
handshake to occur. With hardware handshake, hardware signals control data
flow.
XON/XOFF XON/XOFF stands for Transmit On/Transmit Off. With this
mode the receiver (controller or oscilloscope) controls data flow and can
request that the sender (oscilloscope or controller) stop data flow. By
sending XOFF (ASCII 17) over its transmit data line, the receiver
requests that the sender disables data transmission. A subsequent XON
(ASCII 19) allows the sending device to resume data transmission.
A controller sending data to the oscilloscope should send no more than 32
bytes of data after an XOFF.
The oscilloscope will not send any data after an XOFF is received until an
XON is received.
Programming over RS-232-C
Interface Capabilities
4–8