Agilent Technologies InfiniiVision 5000 Series Guía de usuario Pagina 286

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 296
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 285
286 5000 Series Oscilloscope User’s Guide
8 Power and Environmental Conditions
Transient Withstand Capability
CAUTION
Maximum input voltage for analog inputs
CAT I 300 Vrms, 400 Vpk; transient overvoltage 1.6 kVpk
CAT II 100 Vrms, 400 Vpk
with N2863A 10:1 probe: CAT I 600 V, CAT II 300 V (DC + peak AC)
with 10073C 10:1 probe: CAT I 500 Vpk, CAT II 400 Vpk
with 50 input: 5 Vrms
CAUTION
Do not exceed 5 Vrms in 50 mode on the 2-channel models. Input
protection is enabled in 50 mode, and the 50 load will disconnect
if greater than 5 Vrms is detected. However, the input could still be
damaged, depending on the time constant of the signal. The 50 input
protection mode only functions when the oscilloscope is powered on.
Vista de pagina 285
1 2 ... 281 282 283 284 285 286 287 288 289 290 291 ... 295 296

Comentarios a estos manuales

Sin comentarios