56 N2820/1A User’s Guide
3 Characteristics and Specifications
Zoomed-In Input Voltage Range Versus Supply Voltage
Zoomed-In Input Voltage Range Versus Supply Voltage
For the zoomed- in probe input, the following graph shows
the typical input voltage range allowed based on the supply
voltage that you are measuring. For example, when
measuring a 5V input, the input voltage range is \45 mV.
Figure 31 Zoomed-In Input Voltage Range Versus Supply Voltage
NOTE This plot can be used with N2825A heads (user-defined R
SENSE
) to verify
current range versus supply voltage.