Agilent Technologies N4962A Guía de usuario Pagina 70

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 88
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 69
Remote GPIB Interface
70
Serial BERT 12.5 Gb/s User Guide
5.5.3 Measurement Commands
Command
:MODe?
Type
Query
Description
Returns the current data path.
Results
{NORM | THRU | OFF}
Example
See below
Command
:MODe {NORMal | THRU | OFF}
Type
Command
Description
Sets the instrument data path:
NORMal
Normal operation, IN/IN¯¯connected to error detector, PRBS
generator enabled.
THRU
Input through, IN/IN¯¯ connected to DataThru/DataThru¯¯¯¯¯¯ PRBS
generator enabled.
OFF
Disabled, IN/IN¯¯connected to the error detector, only the PRBS
generator disabled.
Parameters
{NORM | THRU | OFF}
Example
OUTPUT 725; ":MOD THRU"
OUTPUT 725; ":MOD?"
ENTER 725; value$
PRINT value$ ! should be “THRU”
Vista de pagina 69
1 2 ... 65 66 67 68 69 70 71 72 73 74 75 ... 87 88

Comentarios a estos manuales

Sin comentarios