Agilent Technologies N5230C Manual de servicio Pagina 213

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 296
  • Tabla de contenidos
  • SOLUCIÓN DE PROBLEMAS
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 212
Service Guide N5230-90025 7-11
PNA Series Microwave Network Analyzers Repair and Replacement Procedures
N5230C Removing and Replacing Front Panel Subassemblies
Replacing the A2 Display Assembly
1. Remove the inverter board. Refer to procedure on page 7-10. Remove the inverter cable from the cable
clamp (item
).
1. Remove the touch screen controller board. Refer to above procedure.
2. Remove the six retaining screws (item
) from the A2 display assembly.
3. Remove the A2 display assembly from the front panel assembly.
4. Install the new A2 display assembly by reversing the removal instructions.
5. Perform the post-repair adjustments, verifications, and performance tests that pertain to this
replacement procedure. Refer to Table 7-2 on page 7-58.
Figure 7-3 Front Panel Subassemblies Removal
Vista de pagina 212
1 2 ... 208 209 210 211 212 213 214 215 216 217 218 ... 295 296

Comentarios a estos manuales

Sin comentarios