User’s Guide 47
U3025AE10 Operational Check
1-Port Calibration Procedure
1. Connect the ECal or the mechanical cal kit to Port 1 or the port you are testing.
Torque to 8 in-lb. For more information press the Help button, see Figure 43.
Figure 43 ECal at Port 1
2. Perform a 1-Port Calibration on Port 1. On the PNA, press Calibration > Calibration
Wizard.
a. If using a mechanical cal kit, select SmartCal.
b. If using an ECal module, select ECal.
3. Continue following the Cal Wizard prompts. In the “Select Calibration Ports and ECal
Module” window, press Clear All and select Port 1, then press Next.
Figure 44 1-Port Calibration