Agilent Technologies N5230C Guía de usuario Pagina 42

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 54
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 41
36 User’s Guide U3025-90001
U3025AE10 Operational Check Procedures
9. From the Cal Sets drop-down menu, select 999.1 and check Enable. Select the
Reflection Tracking(1,1) term in the center drop-down menu and ensure that the
Enable and Error Terms are selected.
NOTE You may also create a table on the PNA and enter the limit line stimulus and
response values. Select Marker/Analysis > Analysis> Limit Test > Limit Test
ON and Limit Line ON > Global pass/fail display ON, select Show Table.
Figure 35 PNA-X Setting the Test Limits
Figure 36 Reflection Tracking Trace (Port 1-12)
Vista de pagina 41
1 2 ... 37 38 39 40 41 42 43 44 45 46 47 ... 53 54

Comentarios a estos manuales

Sin comentarios