Agilent Technologies N5230C Guía de usuario Pagina 78

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 102
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 77
70 User’s Guide
U3042AE12 Operational Check
N5230A 1-Port Calibration and Verification Procedure
1. Connect the ECal or the mechanical cal kit to Port 1 or the port you are testing.
Torque to 8 in-lb. For more information press the Help button, see Figure 51.
Figure 51 ECal at N5230A Port 1
2. Perform a 1-Port Calibration on Port 1. On the PNA, press Calibration > Calibration
Wizard.
a. If using a mechanical cal kit, select SmartCal.
b. If using an ECal module, select ECal.
3. Continue following the Cal Wizard prompts. In the “Select Calibration Ports and ECal
Module” window, press Clear All and select Port 1, then press Next. See Figure 52.
Figure 52 1-Port Calibration
4. Continue to follow the prompts until the “Calibration Completed” window appears.
Help
Vista de pagina 77
1 2 ... 73 74 75 76 77 78 79 80 81 82 83 ... 101 102

Comentarios a estos manuales

Sin comentarios