3-13
Testing Performance
To verify bandwidth
To verify bandwidth
This test verifies bandwidth. In this test you will use an oscilloscope calibrator
with a level sinewave output.
You will use the peak-to-peak volt age both at 1 MHz and at bandwidth frequency
to verify the bandwidth response of the oscilloscope.
54622A, 54622D, and 54624A
Test limits at 1 mV/div to 5 V/div:
All channels (±3 dB)
dc to 100 MHz
ac coupled 10 Hz to 100 MHz
54621A and 54621D
Test limits at 1 mV/div to 5 V/div:
All channels (±3 dB)
dc to 60 MHz
ac coupled 10 Hz to 60 MHz
Table 3-6
Equipment Required to Verify Bandwidth
Equipment Critical Specifications Recommended Model/Part
Oscilloscope Calibrator Fluke 5820A
Cable * Type N (m), 24-inch Agilent 11500B
Feedthrough 50
, BNC (m) and (f) Agilent 11048C
* The oscilloscope calibrator is supplied with 2 or more coaxial cables N (m), BNC (m),
1 meter long, Fluke P/N 686318.
service.book Page 13 Wednesday, December 18, 2002 8:35 AM