22 34410A/11A/L4411A User’s Guide
1Quick Start
To M ea s u re C apacita nce
Press to select the capacitance function.
Ranges: 1 nF, 10 nF, 100 nF, 1 mF, 10 mF
Configurable parameters: RANGE, NULL, and NULL VALUE
To nullout the test lead capacitance:
1 Disconnect the + lead’s probe end from the test circuit, and leave open.
2 Press null.
3 Reconnect the + lead’s probe end to the test circuit, and measure the
corrected capacitance value.
Connect test leads as shown:
Capacitance
UG_ED5.book Page 22 Thursday, March 1, 2012 11:28 AM