6
Graph 3.
–25 dBm input level
at 25°C.
Typical* power measurement uncertainty at –25 dBm
input level
1.8
1.6
1.4
1.2
1
0.8
0.6
0.4
0.2
0
0.25 1 2 4 6 8 10 12 14 16 18 20 22 26.5
± dB Uncertainty
Frequency in GHz
HP 53150A/151A
HP 53152A
Graph 4.
–25 dBm input level
from 0 to 55°C.
1
0.9
0.8
0.7
0.6
0.5
0.5
0.3
0.2
0.1
0
0.25 1 2 4 6 8 10 12 14 16 18 20 22 26.5
± dB Uncertainty
Frequency in GHz
HP 53150A/151A
HP 53152A
*Typical means approximately 2/3 of all
units will meet these characteristics.