Additional 53200 Series capabilities
In addition to the built-in math and statistic features, the 53230A model provides new measurement capability not avail-
able in previous generation counters.
• Continuous, gap-free measurements for true Allan Deviation computations. In this mode, all samples are taken within a
single gate open/gate close sequence and computed back-to-back. There is no gap between readings that otherwise
occurs with the per sample gate open/gate close sequence.
Time stamp measurements to record events (edges) as they occur on the counter input channels. Timestamp measure-
ment speed is up to one million samples per second.
Microwave burst/pulse measurements are available on optional channel 3 with the pulse/burst measurement software
(Option 150). Measurements include: Burst carrier frequency, pulse repetition interval (PRI), pulse repetition frequency
(PRF), positive (on) width, and negative (off) width.