Analog power sweep
Range: –20 dBm to maximum available power, can
be offset using step attenuator.
External leveling
Range:
At external 33330D/E detector: –36 to +4 dBm
At external leveling input: –200 µV to –0.5 volts
Bandwidth
External detector mode: 10 or 100 kHz (sweep
speed and modulation mode dependent),
nominal
Power meter mode: 0.7 Hz, nominal
Source match
(internally leveled), typical
6
<20 GHz 1.6:1 SWR
<40 GHz 1.8:1 SWR
<50 GHz 2.0:1 SWR
Spectral purity
Specifications apply in CW, step, list, and manual
sweep modes of operation. Specifications for
harmonics beyond maximum instrument frequen-
cies are typical.
Spurious signals (dBc)
Harmonics
Agilent model numbers
Output 83620B 83623B 83623L 83630B/L 83640B/L
frequencies 83622B 83624B 83650B/L
<2.0 GHz
Standard –30 –25
7
–25
7
–30 –30
7
Option 006 –30
7
–25
7
–30
7
–30
7
³2.0 GHz and
<26.5 GHz
Standard –50 –25 –45 –50 –50
Option 006 –60 –60 –60 –50
³26.5 GHz
Standard –40
Option 006 –40
Subharmonics
Output 83620B 83623B 83623L 83630B/L 83640B/L
frequencies 83622B 83624B 83650B/L
<7 GHz None None None None None
³7 and
²20 GHz –50 –50 –50 –50 –50
>20 GHz and
²40 GHz –50 –40
8
>40 GHz –35
8
6. Typically 2.0:1 SWR at frequencies below 50 MHz.
7. Specification is –20 dBc below 50 MHz.
8. Specifications typical below 0 dBm.