Service
Chapter 348
Figure 3-3 Example of an Unacceptable Waveform
TIP You will find that positioning the wire for switching transients affects the offset. Go
back and forth between the two wires, positioning and repositioning, until both
adjustments are deemed acceptable.
Step 7. Reassemble the Power Sensor, ensuring that the waveform shown on the
oscilloscope does not change. If the waveform has changed, remove the cover and
readjust the black/white and brown/white wires again.