Agilent Technologies 8487D Especificaciones

Busca en linea o descarga Especificaciones para Multímetros Agilent Technologies 8487D. Agilent Technologies 8487D Specifications Manual de usuario

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 144
  • Tabla de contenidos
  • SOLUCIÓN DE PROBLEMAS
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente

Indice de contenidos

Pagina 1 - Service Guide

Agilent Technologies Part no. E4418-90059Revision CSeptember 2003Service GuideAgilent TechnologiesE4418B/E4419B Power Meters

Pagina 2 - Printed in the UK

Sales and Service Officesx Agilent E4418B/E4419B Service GuideEurope:Agilent TechnologiesTest & MeasurementEuropean Marketing OrganizationP.O. Box

Pagina 3 - Legal Information

Replaceable PartsAssembly and Disassembly Guidelines5-16 Agilent E4418B/E4419B Service GuideReplacing the Front Panel Power Sensor Cable Assemblies1.

Pagina 4 - Exclusive Remedies

Replaceable PartsAssembly and Disassembly GuidelinesAgilent E4418B/E4419B Service Guide 5-17

Pagina 5 - Equipment Operation

Replaceable PartsAssembly and Disassembly Guidelines5-18 Agilent E4418B/E4419B Service GuideReplacing the Front Panel Power Reference Cable Assembly1.

Pagina 6 - Markings

Replaceable PartsAssembly and Disassembly GuidelinesAgilent E4418B/E4419B Service Guide 5-19Replacing the Rear Panel Power Sensor Cable Assemblies (Op

Pagina 7 - Installation Instructions

Replaceable PartsAssembly and Disassembly Guidelines5-20 Agilent E4418B/E4419B Service Guide

Pagina 8 - List of Related Publications

Replaceable PartsAssembly and Disassembly GuidelinesAgilent E4418B/E4419B Service Guide 5-21Replacing the Rear Panel Power Reference Cable Assembly (O

Pagina 9

Replaceable PartsAssembly and Disassembly Guidelines5-22 Agilent E4418B/E4419B Service GuideReplacing the Rear Panel Recorder Output(s) Cable Assembly

Pagina 10 - United States:

Replaceable PartsAssembly and Disassembly GuidelinesAgilent E4418B/E4419B Service Guide 5-23Replacing The Chassis Assembly1. Remove the Power Meter Co

Pagina 11 - Sales and Service Offices

Replaceable PartsAssembly and Disassembly Guidelines5-24 Agilent E4418B/E4419B Service Guide

Pagina 13 - Table of Contents

Sales and Service OfficesAgilent E4418B/E4419B Service Guide xiReturning Your Power Meter for ServiceUse the information in this section if you need t

Pagina 14

TroubleshootingIntroduction6-2 Agilent E4418B/E4419B Service GuideIntroductionThis chapter enables qualified service personnel to diagnose suspected f

Pagina 15

TroubleshootingIntroductionAgilent E4418B/E4419B Service Guide 6-3Figure 6-1: Troubleshooting Guide -Upper Window TTL OutputStartPower Up meterPOSTPa

Pagina 16

TroubleshootingIntroduction6-4 Agilent E4418B/E4419B Service GuideASelectPreset/localConfirmMeas/SetupLimitsLimits OnTTL OutputTTL Output On, Fail O/P

Pagina 17 - Specifications

TroubleshootingIntroductionAgilent E4418B/E4419B Service Guide 6-5Figure 6-2: Troubleshooting Guide - Lower Window TTL OutputStartPower Up meterPOSTP

Pagina 18

TroubleshootingIntroduction6-6 Agilent E4418B/E4419B Service GuideASelectPreset/localConfirmLower Window - usingMeas/SetupLimitsLimits OnTTL OutputTTL

Pagina 19 - Power Meter Specifications

TroubleshootingIntroductionAgilent E4418B/E4419B Service Guide 6-7Figure 6-3: Troubleshooting Guide - TTL InputsStartPower Up meterPOSTPass?Connect S

Pagina 20 - Accuracy

TroubleshootingIntroduction6-8 Agilent E4418B/E4419B Service GuideABDid powermeter channelszero?Apply TTL logic 1 level to RJ45 connector. pin 5 wrtpi

Pagina 21 - 1 mW Power Reference

TroubleshootingIntroductionAgilent E4418B/E4419B Service Guide 6-9BRemove A2 Processor Assy (see chapter 5) to gain accessto the Communications Assemb

Pagina 22 - Zero Drift of Sensors

TroubleshootingIntroduction6-10 Agilent E4418B/E4419B Service GuideFigure 6-4: Troubleshooting Guide - RS232/422 InterfaceStartPower upmeterPOSTPass?

Pagina 23 - Measurement Noise

Measurement Uncertainty Analysis – Instrument Accuracy TestAgilent E4416A/E4417A Service Guide A-1Appendix AMeasurement Uncertainty Analysis – Instrum

Pagina 24

About this Guidexii Agilent E4418B/E4419B Service GuideAbout this GuideChapter 1: SpecificationsThis chapter lists the power meter’s specifications an

Pagina 25 - Settling Time

Measurement Uncertainty Analysis – Instrument Accuracy TestMeasurement IntroductionA-2 Agilent E4416A/E4417A Service GuideMeasurement IntroductionIn t

Pagina 26 - 12481632641282565121024

Measurement Uncertainty Analysis – Instrument Accuracy TestMeasurement IntroductionAgilent E4416A/E4417A Service Guide A-3Uncertainties1. There is no

Pagina 27

Measurement Uncertainty Analysis – Instrument Accuracy TestMeasurement IntroductionA-4 Agilent E4416A/E4417A Service GuideThe Measurement Uncertainty

Pagina 28 - Power Sensor Specifications

Measurement Uncertainty Analysis – Instrument Accuracy TestMeasurement IntroductionAgilent E4416A/E4417A Service Guide A-5Standard Uncertainty u(xi):T

Pagina 29

Measurement Uncertainty Analysis – Instrument Accuracy TestMeasurement IntroductionA-6 Agilent E4416A/E4417A Service GuideWorst case error on 0.1V ran

Pagina 30 - General Characteristics

Measurement Uncertainty Analysis – Instrument Accuracy TestMeasurement IntroductionAgilent E4416A/E4417A Service Guide A-7Worst case error on 15.8V ma

Pagina 31 - Environmental Characteristics

Measurement Uncertainty Analysis – Instrument Accuracy TestMeasurement IntroductionA-8 Agilent E4416A/E4417A Service GuideBy using the DMM Uncertainti

Pagina 32 - Dimensions

Measurement Uncertainty Analysis – Instrument Accuracy TestMeasurement IntroductionAgilent E4416A/E4417A Service Guide A-9Combined Standard Uncertaint

Pagina 33 - Non-Volatile Memory

Measurement Uncertainty Analysis – Instrument Accuracy TestMeasurement IntroductionA-10 Agilent E4416A/E4417A Service Guide

Pagina 34

Measurement Uncertainty Analysis – Power Reference Level TestAgilent E4416A/E4417A Service Guide B-1Appendix BMeasurement Uncertainty Analysis – Power

Pagina 35 - Performance Tests

Table of ContentsPageAgilent E4418B/E4419B Service Guide Contents-1Specifications...

Pagina 36

Measurement Uncertainty Analysis – Power Reference Level TestMeasurement IntroductionB-2 Agilent E4416A/E4417A Service GuideMeasurement IntroductionIn

Pagina 37 - Equipment Required

Measurement Uncertainty Analysis – Power Reference Level TestMeasurement IntroductionAgilent E4416A/E4417A Service Guide B-3where,Ps is the reflection

Pagina 38 - Test Procedures

Measurement Uncertainty Analysis – Power Reference Level TestMeasurement IntroductionB-4 Agilent E4416A/E4417A Service Guide:The equation above includ

Pagina 39 - Zero Test

Measurement Uncertainty Analysis – Power Reference Level TestMeasurement IntroductionAgilent E4416A/E4417A Service Guide B-5A normal distribution is a

Pagina 40 - Procedure

Measurement Uncertainty Analysis – Power Reference Level TestMeasurement IntroductionB-6 Agilent E4416A/E4417A Service GuideConsider the model below:A

Pagina 41

Measurement Uncertainty Analysis – Power Reference Level TestMeasurement IntroductionAgilent E4416A/E4417A Service Guide B-7So u(Ms) from the equation

Pagina 42 - Test Setup

Measurement Uncertainty Analysis – Power Reference Level TestMeasurement IntroductionB-8 Agilent E4416A/E4417A Service GuideResistanceThe uncertainty

Pagina 43

Measurement Uncertainty Analysis – Power Reference Level TestMeasurement IntroductionAgilent E4416A/E4417A Service Guide B-9Table B-1: Measurement Re

Pagina 44

Measurement Uncertainty Analysis – Power Reference Level TestMeasurement IntroductionB-10 Agilent E4416A/E4417A Service GuideThe readings in Table B-1

Pagina 45 - Equipment

Measurement Uncertainty Analysis – Power Reference Level TestMeasurement IntroductionAgilent E4416A/E4417A Service Guide B-11Using the above values gi

Pagina 46

Contents-2 Agilent E4418B/E4419B Service GuideEquipment ... 2-5Test Setup..

Pagina 47

Measurement Uncertainty Analysis – Power Reference Level TestMeasurement IntroductionB-12 Agilent E4416A/E4417A Service GuideCombined Standard Uncerta

Pagina 48

IndexIndex-1 Agilent E4418B/E4419B Service GuideAA1 power supply/Battery Charger assembly 4-3A2 processor assembly 4-4A3 front panel assembly 4-7A4 in

Pagina 49 - Power Reference Level Test

Index-2 Agilent E4418B/E4419B Service Guidefront panel repair 5-6GGP-IBconnector characteristic 1-14Hhumidity characteristic 1-15IIEC 1010-1 1-viiinst

Pagina 50

Agilent E4418B/E4419B Service Guide Index-3replaceable parts 5-2returning for service 1-xiRmt I/O 1-14RS232connector characteristics 1-14RS422connecto

Pagina 51

Index-4 Agilent E4418B/E4419B Service Guide

Pagina 52

Agilent E4418B/E4419B Service Guide Contents-3Procedure... 3-11Theory of

Pagina 53

Contents-4 Agilent E4418B/E4419B Service GuideDefine the Measurement Equation...A-3Sensitivity Coefficients Ci:...

Pagina 55

SpecificationsIntroduction1-2 Agilent E4418B/E4419B Service GuideIntroductionThis chapter details the power meter’s specifications and supplemental ch

Pagina 56

SpecificationsPower Meter SpecificationsAgilent E4418B/E4419B Service Guide 1-3Power Meter SpecificationsMeterFrequency Range100 kHz to 110 GHz, power

Pagina 57

ii Agilent E4418B/E4419B Service Guide© Copyright 1998, 2003Agilent TechnologiesAll rights reserved. Reproduction, adaptation, or translation without

Pagina 58

SpecificationsPower Meter Specifications1-4 Agilent E4418B/E4419B Service GuideAccuracyInstrumentationAbsolute: ±0.02 dB (Logarithmic) or ±0.5% (Linea

Pagina 59 - Adjustments

SpecificationsPower Meter SpecificationsAgilent E4418B/E4419B Service Guide 1-51 mW Power Reference1• Power Output: 1.00 mW (0.0 dBm). • Accuracy: (fo

Pagina 60

SpecificationsPower Meter Supplemental Characteristics1-6 Agilent E4418B/E4419B Service GuidePower Meter Supplemental CharacteristicsMeasurement Speed

Pagina 61 - Post-Repair Adjustments

SpecificationsPower Meter Supplemental CharacteristicsAgilent E4418B/E4419B Service Guide 1-7Measurement NoisePower sensor dependent (refer to Table 1

Pagina 62

SpecificationsPower Meter Supplemental Characteristics1-8 Agilent E4418B/E4419B Service GuideTable 1-3: Power Sensor Specifications±Power Sensor Zero

Pagina 63

SpecificationsPower Meter Supplemental CharacteristicsAgilent E4418B/E4419B Service Guide 1-9Settling Time0 to 99% settled readings over the GP-IB.For

Pagina 64

SpecificationsPower Meter Supplemental Characteristics1-10 Agilent E4418B/E4419B Service GuideFor Agilent E-series power sensorsIn FAST mode, within t

Pagina 65

SpecificationsPower Meter Supplemental CharacteristicsAgilent E4418B/E4419B Service Guide 1-11Auto filter, default resolution, 10 dB decreasing power

Pagina 66

SpecificationsPower Meter Supplemental Characteristics1-12 Agilent E4418B/E4419B Service GuidePower Sensor SpecificationsDefinitionsZero SetIn any pow

Pagina 67

SpecificationsBattery Option 001 Operational CharacteristicsAgilent E4418B/E4419B Service Guide 1-13Battery Option 001 Operational CharacteristicsThe

Pagina 68 - Measurement Uncertainty

Legal InformationAgilent E4418B/E4419B Service Guide iiiLegal InformationNoticeInformation contained in this document is subject to change without not

Pagina 69

SpecificationsGeneral Characteristics1-14 Agilent E4418B/E4419B Service GuideGeneral CharacteristicsRear Panel ConnectorsRecorder Output(s)Analog 0-1

Pagina 70

SpecificationsEnvironmental CharacteristicsAgilent E4418B/E4419B Service Guide 1-15Environmental CharacteristicsGeneral ConditionsComplies with the re

Pagina 71 - Theory of Operation

SpecificationsGeneral1-16 Agilent E4418B/E4419B Service GuideGeneralDimensionsThe following dimensions exclude front and rear panel protrusions:212.6

Pagina 72

SpecificationsGeneralAgilent E4418B/E4419B Service Guide 1-17Remote ProgrammingInterfaceGP-IB interface operates to IEEE 488.2. RS-232 and RS-422 inte

Pagina 73

SpecificationsGeneral1-18 Agilent E4418B/E4419B Service Guide

Pagina 75

Performance TestsIntroduction2-2 Agilent E4418B/E4419B Service GuideIntroductionThe procedures in this chapter test the power meter’s electrical perfo

Pagina 76

Performance TestsIntroductionAgilent E4418B/E4419B Service Guide 2-3Equipment RequiredTable 2-1 lists all the equipment required for the adjustments a

Pagina 77 - A3 Front Panel Assembly

Performance TestsIntroduction2-4 Agilent E4418B/E4419B Service GuidePerformance Test RecordResults of the performance tests may be tabulated in Table

Pagina 78 - A4 Comms Assembly

Performance TestsZero TestAgilent E4418B/E4419B Service Guide 2-5Zero Test SpecificationDescriptionAfter the power meter is initially calibrated and z

Pagina 79 - A5 Daughter Assembly

Legal Informationiv Agilent E4418B/E4419B Service GuideLimitation of Warranty The foregoing warranty shall not apply to defects resulting from imprope

Pagina 80 - A6 Measurement Assembly

Performance TestsZero Test2-6 Agilent E4418B/E4419B Service GuideProcedureThe following procedure should be performed for the Zero Test. Note The pr

Pagina 81

Performance TestsInstrument Accuracy TestAgilent E4418B/E4419B Service Guide 2-7Instrument Accuracy Test SpecificationDescriptionThe power meter accur

Pagina 82

Performance TestsInstrument Accuracy Test2-8 Agilent E4418B/E4419B Service GuideTest SetupFigure 2-2: Instrument Accuracy Test SetupProcedureThe foll

Pagina 83

Performance TestsInstrument Accuracy TestAgilent E4418B/E4419B Service Guide 2-9Note When switching the range calibrator to STANDBY, allow enough ti

Pagina 84

Performance TestsInstrument Accuracy Test2-10 Agilent E4418B/E4419B Service GuideTable 2-3: Instrument Accuracy Results (For Indication Only)Note T

Pagina 85

Performance TestsInstrument Accuracy TestAgilent E4418B/E4419B Service Guide 2-11Test Procedure 2Note This test procedure makes use of a 11683A Rang

Pagina 86

Performance TestsInstrument Accuracy Test2-12 Agilent E4418B/E4419B Service GuideProcedureThe following procedure should be performed for the Instrume

Pagina 87 - Assembly Part Numbers

Performance TestsInstrument Accuracy TestAgilent E4418B/E4419B Service Guide 2-13Note When switching the range calibrator to STANDBY, allow enough t

Pagina 88

Performance TestsInstrument Accuracy Test2-14 Agilent E4418B/E4419B Service Guide14. Set the DC calibrator output to the values shown in Table 2-4. Fo

Pagina 89 - Power Supply Assembly Types

Performance TestsPower Reference Level TestAgilent E4418B/E4419B Service Guide 2-15Power Reference Level Test DescriptionThis test verifies the power

Pagina 90 - Front Panel Connector Repair

Equipment OperationAgilent E4418B/E4419B Service Guide vEquipment OperationWarnings and CautionsThis guide uses warnings and cautions to denote hazard

Pagina 91 - Firmware Upgrades

Performance TestsPower Reference Level Test2-16 Agilent E4418B/E4419B Service GuideIf an equivalent measurement system is employed for post-factory te

Pagina 92

Performance TestsPower Reference Level TestAgilent E4418B/E4419B Service Guide 2-17Equipment• Test power meter: Agilent 432A.• Thermistor mount: Agile

Pagina 93

Performance TestsPower Reference Level Test2-18 Agilent E4418B/E4419B Service Guide4. Preset the power meter:Press , then on the power meter.Set th

Pagina 94

Performance TestsAgilent E-Series Power Sensor Interface TestAgilent E4418B/E4419B Service Guide 2-19Agilent E-Series Power Sensor Interface TestThe A

Pagina 95

Performance TestsAgilent E-Series Power Sensor Functional Test2-20 Agilent E4418B/E4419B Service GuideAgilent E-Series Power Sensor Functional TestDes

Pagina 96

Performance TestsAgilent E-Series Power Sensor Functional TestAgilent E4418B/E4419B Service Guide 2-212. Press then .3. On the power meter press ,

Pagina 97 - Removing the A7 Fan Assembly

Performance TestsAgilent E-Series Power Sensor Functional Test2-22 Agilent E4418B/E4419B Service GuideTable 2-6: Functional Test ResultTestChannel A

Pagina 98

Performance TestsPerformance Test RecordAgilent E4418B/E4419B Service Guide 2-23Performance Test RecordModel Agilent E4418B/E4419B Power MeterTested b

Pagina 99

Performance TestsPerformance Test Record2-24 Agilent E4418B/E4419B Service GuideOutput Standing Wave Ratio Test1N/A___________ ___________1.061, 21.08

Pagina 101

Equipment Operationvi Agilent E4418B/E4419B Service GuideGeneral Safety ConsiderationsWARNING Before this instrument is switched on, make sure it ha

Pagina 102 - Assembly

AdjustmentsIntroduction3-2 Agilent E4418B/E4419B Service GuideIntroduction This chapter describes adjustments and checks which ensure proper performan

Pagina 103 - (Options 002 and 003)

AdjustmentsIntroductionAgilent E4418B/E4419B Service Guide 3-3Equipment RequiredThe adjustment procedures include a list of recommended test equipment

Pagina 104

AdjustmentsPower Reference Oscillator Frequency Adjustment3-4 Agilent E4418B/E4419B Service GuidePower Reference Oscillator Frequency AdjustmentNote

Pagina 105 - (Option 003)

AdjustmentsPower Reference Oscillator Frequency AdjustmentAgilent E4418B/E4419B Service Guide 3-53. Turn the power reference on by pressing , on th

Pagina 106

AdjustmentsPower Reference Oscillator Level Adjustment3-6 Agilent E4418B/E4419B Service GuidePower Reference Oscillator Level AdjustmentDescriptionThi

Pagina 107

AdjustmentsPower Reference Oscillator Level AdjustmentAgilent E4418B/E4419B Service Guide 3-7Note The power meter may be returned to the nearest Agi

Pagina 108

AdjustmentsPower Reference Oscillator Level Adjustment3-8 Agilent E4418B/E4419B Service GuideBy substituting equation 3 into equation 1 and manipulati

Pagina 109 - Troubleshooting

AdjustmentsPower Reference Oscillator Level AdjustmentAgilent E4418B/E4419B Service Guide 3-93. Connect the equipment as shown in Figure 3-3. The lead

Pagina 110 - Introduction

AdjustmentsPower Reference Oscillator Level Adjustment3-10 Agilent E4418B/E4419B Service Guide14.Remove the power meter’s cover and adjust A2R90 until

Pagina 111

AdjustmentsDisplay Brightness and Contrast AdjustmentAgilent E4418B/E4419B Service Guide 3-11Display Brightness and Contrast AdjustmentIntroductionThe

Pagina 112

Equipment OperationAgilent E4418B/E4419B Service Guide viiIEC 1010-1 ComplianceThis instrument has been designed and tested in accordance with IEC Pub

Pagina 113

AdjustmentsDisplay Brightness and Contrast Adjustment3-12 Agilent E4418B/E4419B Service Guide

Pagina 114

4Theory of Operation

Pagina 115

Theory of OperationIntroduction4-2 Agilent E4418B/E4419B Service GuideIntroductionThis chapter describes how each of the power meter’s assemblies oper

Pagina 116

Theory of OperationA1 Power Supply/Battery ChargerAgilent E4418B/E4419B Service Guide 4-3A1 Power Supply/Battery ChargerThe A1 power supply/battery ch

Pagina 117

Theory of OperationA2 Processor Assembly4-4 Agilent E4418B/E4419B Service GuideA2 Processor AssemblyThe processor assembly contains the microcontrolle

Pagina 118

Theory of OperationA2 Processor AssemblyAgilent E4418B/E4419B Service Guide 4-5The LCD controller on the A3 front panel assembly is configured as a me

Pagina 119 - Instrument Accuracy Test

Theory of OperationA2 Processor Assembly4-6 Agilent E4418B/E4419B Service GuideThe microcontroller circuits that control all the above functions, and

Pagina 120 - Measurement Introduction

Theory of OperationA3 Front Panel AssemblyAgilent E4418B/E4419B Service Guide 4-7A3 Front Panel AssemblyThe front panel assembly is made up of a liqui

Pagina 121 - Define the Measurand

Theory of OperationA4 Comms Assembly4-8 Agilent E4418B/E4419B Service GuideA4 Comms AssemblyThe comms assembly contains the circuitry required for rem

Pagina 122

Theory of OperationA5 Daughter AssemblyAgilent E4418B/E4419B Service Guide 4-9A5 Daughter AssemblyThe A5 daughter assembly is loaded vertically into t

Pagina 123

List of Related Publicationsviii Agilent E4418B/E4419B Service GuideList of Related PublicationsThe Agilent E4418B and Agilent E4419B User’s Guides ar

Pagina 124

Theory of OperationA6 Measurement Assembly4-10 Agilent E4418B/E4419B Service GuideA6 Measurement AssemblyThere is one measurement assembly in the Agil

Pagina 125

Theory of OperationA6 Measurement AssemblyAgilent E4418B/E4419B Service Guide 4-11The DSP chip controls the logic which sets the chopper driver voltag

Pagina 126 - Sensitivity Coefficients Ci:

Theory of OperationA8 Rechargeable Battery Assembly4-12 Agilent E4418B/E4419B Service GuideA8 Rechargeable Battery Assembly The A8 Rechargeable Batter

Pagina 127 - Expanded Uncertainty kuc(y):

Theory of OperationA8 Rechargeable Battery AssemblyAgilent E4418B/E4419B Service Guide 4-13Remove this page and replace with pullout.fm.

Pagina 128

Theory of OperationA8 Rechargeable Battery Assembly4-14 Agilent E4418B/E4419B Service Guide

Pagina 130

Replaceable PartsIntroduction5-2 Agilent E4418B/E4419B Service GuideIntroductionThis chapter contains details of some of the higher level components a

Pagina 131

Replaceable PartsAssembly Part NumbersAgilent E4418B/E4419B Service Guide 5-3Assembly Part NumbersReference DesignatorDescriptionAgilent Part NumberA1

Pagina 132 - Uncertainty Equation

Replaceable PartsAssembly Part Numbers5-4 Agilent E4418B/E4419B Service GuideA304 Front Panel Assembly for:Agilent E4419B Option 003New E4419-61002Ref

Pagina 133 - Mismatch

Replaceable PartsAssembly Part NumbersAgilent E4418B/E4419B Service Guide 5-5Power Supply Assembly TypesBefore ordering an A1 Power Supply Assembly, c

Pagina 134

Sales and Service OfficesAgilent E4418B/E4419B Service Guide ixSales and Service OfficesFor more information about Agilent Technologies test and measu

Pagina 135 - Calibration factor

Replaceable PartsFront Panel Connector Repair5-6 Agilent E4418B/E4419B Service GuideFront Panel Connector RepairThe front panel assembly is an exchang

Pagina 136 - Repeatability

Replaceable PartsFirmware UpgradesAgilent E4418B/E4419B Service Guide 5-7Firmware UpgradesThe power meter’s firmware can be upgraded using a PC. The c

Pagina 137

Replaceable PartsAssembly and Disassembly Guidelines5-8 Agilent E4418B/E4419B Service GuideAssembly and Disassembly GuidelinesThe guidelines in this s

Pagina 138 - Sensitivity Coefficients (Ci)

Replaceable PartsAssembly and Disassembly GuidelinesAgilent E4418B/E4419B Service Guide 5-9Removing the A1 Power Supply Assembly1. Remove the power su

Pagina 139

Replaceable PartsAssembly and Disassembly Guidelines5-10 Agilent E4418B/E4419B Service GuideRemoving the A2 Processor Assembly1. Remove the A5 daughte

Pagina 140 - Expanded Uncertainty

Replaceable PartsAssembly and Disassembly GuidelinesAgilent E4418B/E4419B Service Guide 5-11Removing the A4 Comms Assembly1. Remove the A2 processor,

Pagina 141

Replaceable PartsAssembly and Disassembly Guidelines5-12 Agilent E4418B/E4419B Service GuideRemoving the A5 Daughter or A6 Measurement Assemblies1. Di

Pagina 142

Replaceable PartsAssembly and Disassembly GuidelinesAgilent E4418B/E4419B Service Guide 5-13Removing the A7 Fan Assembly1. Remove the A1 power supply

Pagina 143

Replaceable PartsAssembly and Disassembly Guidelines5-14 Agilent E4418B/E4419B Service GuideRemoving the Power Meter RAM Battery (BT1)1. Remove the A2

Pagina 144

Replaceable PartsAssembly and Disassembly GuidelinesAgilent E4418B/E4419B Service Guide 5-15Removing the A3 Front Panel Assembly1. Disconnect the foll

Comentarios a estos manuales