Display Rate
Function Times/Second
ACV 7
ACV + dB 7
DCV 7
AC + DC V 2
Ω/nS 14
Diode 14
Capacitance 4 (< 100 µF)
DCI 7
ACI 7
AC + DC I 2
Temperature 6
Frequency 1 (>10 Hz)
Duty cycle/Pulse width 0.5 (>10 Hz)
Decibel (Db) Calculation
dB Base Reference Default Reference
1 mΩ (dBm) 1-9999 Ω 50 Ω
1 V (dBV) 1 V 1 V
Manual and Interval Data Logging
Logging
Type
Maximum Data Points
[5]
U1251B U1252B U1253B
MANUAL 100 100 100
INTERVAL 200 200 1000
[1] All frequency counters are susceptible to errors. Shielding inputs from external noise pickup is critical to minimize measurement errors. For
non-square wave signals, add 5 counts.
[2] Output impedance: 3.5 kΩ maximum.
[3] The positive or negative pulse width must be greater than 50 µs for adjustment of the duty cycle or pulse width under different frequencies.
Otherwise, the accuracy and range will be different from the specifications defined.
[4] For signal frequencies greater than 1 kHz, an addition of 0.1% per kHz is added to the accuracy.
[5] For data logging to PC, maximum number of data points is dependent on available hard disk space.
[6] For the U1253B model.
Input Impedance
Table A
Function Range U1251B U1252B/U1253B
DC Voltage 50 mV to 1000 mV 10 MΩ 10 MΩ
5 V to 1000 V 10 MΩ (nominal), with 10 MΩ in
parallel at dual display
10 MΩ (nominal), with 10 MΩ in
parallel at dual display
AC Voltage 50 mV to 1000 mV 10 MΩ in parallel with < 100 pF 10 MΩ in parallel with < 100 pF
5 V to 1000 V
AC + DC Voltage 50 mV to 1000 mV N/A 10 MΩ
5 V to 1000 V 10 MΩ (nominal) in parallel with
10 MΩ, < 100 pF