Specifications and Characteristics 6
U3401A User’s and Service Guide 89
Supplemental Measurement Specifications
Display update rate
Measurement Specifications
Table 6 - 8 Full scale display update rate
Display Count
50000
Table 6 - 9 Measurement specifications
DC voltage
Measurement method
Sigma Delta A-to-D converter
Input resistance
10 MΩ ± range (typical)
Maximum input voltage
1000 VDC or PEAK AC on all ranges
Input protection
1000 V on all ranges
Response time
Approximate 1.0 second when the displayed reading reaches 99.9% DC value of the tested
input signal at the same range.