2-port TRL/LRM Calibration Operating Procedure
1. Setting stimulus conditions
Select the ENA’s active channel to the one for which you want to perform calibration and set the stimulus
conditions of the channel by manual operation. Typical setup items includes:
- Start and stop frequencies
- Sweep type
- Number of points
- Source power level
- IF Bandwidth
- Averaging factor and Averaging ON/OFF
For information on the setting procedure, refer to Chapter 3 “Setting Measurement Conditions” in the
E5070B/E5071B User's Guide.
2. Starting VBA macro
Load and run the TRL/LRM Calibration VBA Macro.
Press [Macro Setup], Load&Run, and select TRL_LRM_cal.
(Alternatively, you can press [Macro Setup], Load Project, load “TRL_LRM_cal.vba “ at D:/VBA, and
press
[Macro Run]
.)
3. Turning off system correction
When the message shown in Figure 1 appears, press OK to turn off the system error correction.
NOTES
- This message does not appear when the system error correction has already been turned off.
- System error correction of all channels is turned off.
- System error correction can be turned on by [Preset].
Figure 1. Turning off system error correction
If you press OK, the TRL/LRM Calibration main control panel appears as shown in Figure 2.