N2780/81/82/83 User’s and Service Guide 43
test procedure. While the test system is warming up,
clean the magnetic contacts on the probe jaw.
2 Press the DEMAG button on the probe terminator.
3 Configure the DMM to measure AC Volts.
4 Clamp the probe around the coil and ensure that the
probe is as perpendicular as possible to the coils of wire.
5 Set up the signal generator as follows:
Wave shape: Sine
Frequency: 50 Hz
01 01 COM COM
SENSE
!
Signal Generator
DMM
Probe
Power
Supply
Current Probe
Test Coil
Adapter
medium_standard.book Page 43 Friday, September 10, 2010 2:45 PM