6 – Language Dictionary
66
[SOURce:]CURRent:MODE
[SOURce:]CURRent:LIMit:MODE
These commands determine what happens to the output current and current limit during a triggered event.
FIXed
The output current and output current limit is unaffected when a trigger occurs.
STEP
The output current is set by the CURR:TRIG value when a trigger occurs.
The current limit is set by the CURR:LIM:TRIG value when a trigger occurs.
Command Syntax
[SOURce:]CURRent:MODE <mode>, (@<channel list>)
[SOURce:]CURRent:LIMit:MODE <mode>, (@<channel list>)
Parameters
FIXed | STEP
*RST Value
FIXed
Examples
CURR:MODE FIX,(@1) CURR:LIM:MODE FIX,(@1)
Query Syntax
[SOURce:]CURRent:MODE? (@<channel list>)
[SOURce:]CURRent:LIMit:MODE? (@<channel list>)
Returned Parameters
<CRD>
[SOURce:]DELay
This command sets the delay when [SOUR:]DEL:MODE is set to MANUAL. If an output is changed and
a subsequent measurement is requested, the measurement will be delayed to allow the output to settle.
Command Syntax
[SOURce:]DELay[:TIMe] <Nrf+>, (@<channel list>)
Parameters
0 to 1000 (seconds)
*RST Value
0
Examples
DEL .001,(@1)
Query Syntax
[SOURce:]DELay[:TIMe]? (@<channel list>)
Returned Parameters
<NR3>
[SOURce:]DELay:MODE
This command selects the source delay mode.
AUTO
The dc source selects an appropriate delay for the present output voltage or current
MANual
The delay programmed by [SOURce:] will be used as the delay.
Command Syntax
[SOURce:]DELay:MODE <mode>, (@<channel list>)
Parameters
AUTO | MANual
*RST Value
AUTO
Examples
DEL:MODE AUTO,(@1)
Query Syntax
[SOURce:]DELay:MODE? (@<channel list>)
Returned Parameters
<CRD>