3
THERMOCOUPLE TYPE RANGE RESOLUTION ACCURACY ± (% of Reading + Offset Error)
K (for U1241A, U1241B,
U1242A and U1242B)
–40 to 1000 ºC/
–48 to 1832 ºF
0.1 ºC/
0.1 ºF
1% + 1 ºC/
1% + 1.8 ºF
J (for U1242A and
U1242B only)
–40 to 1000 ºC/
–48 to 1832 ºF
0.1 ºC/
0.1 ºF
1% + 1 ºC/
1% + 1.8 ºF
TEMPERATURE SPECIFICATIONS
CAPACITANCE SPECIFICATIONS
RANGE RESOLUTION ACCURACY ± (% of Reading + No. of Least Significant Digit)
1000.0 nF 0.1 nF
1.2% + 410.000 µF 0.001 µF
100.00 µF 0.01 µF
1000.0 µF 0.1 µF
2% + 4
10.000 mF 0.001 mF
HARMONIC RATIO SPECIFICATIONS
RANGE FREQUENCY VOLTAGE
0.0% to 99.9% 40 Hz to 500 Hz 100 mVAC to 1000 VAC
RANGE RESOLUTION ACCURACY MINIMUM INPUT FREQUENCY
100.00 Hz 0.01 Hz
0.03%+3 1 Hz
1000.0 Hz 0.1 Hz
10.000 kHz 0.001 kHz
100.00 kHz 0.01 kHz
1000.0 kH
z
[1]
0.1 kHz
FREQUENCY SPECIFICATIONS
[1] Effective frequency measurement of up to 200 kH; refer to frequency sensitivity table below for details.
SWITCH COUNTER DEFINITION
SWITCH CONDITION
[1[2]
CIRCUIT SWITCH DISPLAY
[3]
SWITCH THRESHOLD
Low Level Normally Close Lo < 370 Ohms
Intermittent
[4]
Close to Open Number of switch count Low to high transition
High Level Normally Open Hi > 430 Ohms
Intermittent
[5]
Open to Close Number of switch count High to low transition
[1] Detects intermittent Opens or Closes lasting for at least 250 μsec.
[2] Test current of 0.5 mA with maximum open circuit voltage of 2.8 V is used.
[3] Maximum count reading: 199.99 M (display “OL” when achieving 2 x 10
8
and thereafter).
[4] Count only low to high transition for initial switch condition of Low.
[5] Count only high to low transition for initial switch condition of High.
Comentarios a estos manuales