Agilent Technologies U1252A Manual de usuario Pagina 7

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 10
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 6
7
U1252A FREQUENCY SENSITIVITY DURING VOLTAGE MEASUREMENT
MODE RANGE ACCURACY AT FULL SCALE
DC Coupling 0.01% ~ 99.99% 0.3% per kHz + 0.3%
FREQUENCY SENSITIVITY AND TRIGGER LEVEL FOR U1252A
INPUT RANGE MINIMUM SENSITIVITY
(R.M.S. Sine-Wave)
TRIGGER LEVEL FOR DC COUPLING
(Maximum input for specifi ed
accuracy = 10 x Range or 1000 V)
20 Hz ~ 200 kHz >200 kHz ~ 500 kHz < 100 kHz >100 kHz ~ 500 kHz
50.000 mV 10 mV 25 mV 10 mV 25 mV
500.00 mV 70 mV 150 mV 70 mV 150 mV
1000.0 mV 120 mV 300 mV 120 mV 300 mV
5.0000 V 0.3 V 1.2 V 0.6 V 1.5 V
50.000 V 3 V 5 V 6 V 15 V
500.00 V 30 V < 100 kHz N/A 60 V N/A
1000.0 V 50 V < 100 kHz N/A 120 V N/A
DUTY CYCLE
[1]
PULSE WIDTH
[1]
[1] Positive or negative pulse width must be greater than 10 ms and the range of duty cycle should be considered. The range of pulse width is determined by the frequency
of the signal.
MODE RANGE ACCURACY AT FULL SCALE
500 ms 0.01 ms 0.2% + 3
2000 ms 0.1 ms 0.2% + 3
U1251A/U1252A FREQUENCY SENSITIVITY DURING CURRENT MEASUREMENT
INPUT RANGE MINIMUM SENSITIVITY (R.M.S. Sine-Wave)
20 Hz ~ 20 kHz
500.00 µA 100 µA
5000.0 µA 250 µA
50.000 mA 10 mA
440.00 mA 25 mA
5.0000 A 1 A
10.000 A 2.5 A
PEAK HOLD (Capturing changes)
SIGNAL WIDTH ACCURACY FOR DC mV/VOLTAGE/CURRENT
Single event > 1 ms 2% + 400 for all ranges
Repetitive > 250 µs 2% + 1000 for all ranges
U1252A FREQUENCY COUNTER SPECIFICATIONS
Divide 1 (secondary display “-1-”)
RANGE RESOLUTION ACCURACY ± (% of reading + No. of Least
Signifi cant Digit)
SENSITIVITY MIN. INPUT FREQUENCY
99.999 Hz 0.001 Hz 0.02% + 3
100 mV R.M.S.
0.5 Hz
999.99 Hz 0.01 Hz
0.002% + 3,
< 2 MHz
9.9999 kHz 0.0001 kHz
99.999 kHz 0.001 kHz
999.99 kHz 0.01 kHz
200 mV R.M.S.
9.9999 MHz 0.0001 MHz
Divide 100 (secondary display “-100-”)
RANGE RESOLUTION ACCURACY ± (% of reading + No. of Least
Signifi cant Digit)
SENSITIVITY MIN. INPUT FREQUENCY
9.9999 MHz 0.0001 MHz
0.002% + 5,
< 20 MHz
400 mV R.M.S.
1 MHz
99.999 MHz 0.001 MHz 600 mV R.M.S.
Vista de pagina 6