Agilent Technologies U3606B Guía de usuario Pagina 71

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 205
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 70
Operation and Features 2
Measuring Frequency/Pulse Width/Duty Cycle (Voltage Path)
U3606B User’s Guide 55
Frequency/pulse width/duty cycle measurement (voltage path)
summary
Table 2 - 9 Frequency/pulse width/duty cycle measurement (voltage
path) summary
Item Description
Available ranges
100.000 mV, 1.00000 V, 10.0000 V, 100.000 V, 750.00 V —
range is based on the voltage level of the signal, not
frequency
Measurement
method
Reciprocal counting technique
Signal level 10% of range to full scale input on all ranges
Input protection 750 V
rms
on all ranges
U3606-90054_UG_EN.book Page 55 Friday, July 11, 2014 5:12 PM
Vista de pagina 70
1 2 ... 66 67 68 69 70 71 72 73 74 75 76 ... 204 205

Comentarios a estos manuales

Sin comentarios