81642A Tunable Laser
Agilent 81642A
Wavelength range 1510nm to 1640mn
Wavelength resolution 0.1 pm, 12.5 MHz at 1550 nm
Mode-hop free tuning range Full wavelength range
Absolute wavelength accuracy
[1] [2]
±0.015nm
Relative wavelength accuracy
[1] [2]
±7 pm, typ. ±3 pm
Wavelength repeatability
[2]
±1 pm, typ. ±0.5 pm
Wavelength stability
[2]
< ±1 pm (typ., 24 h at const. temp.)
Tuning speed (typ. for a 1/10/100 nm step) 400ms/600ms/2.8s
Linewidth (typ.), coherence control off
Effective linewidth (typ.), coherence control on
100 kHz
> 50 MHz (1520-1620 nm, at maximum flat output power)
Output power (continuous power during tuning)
[3]
for #003
≥ +7 dBm peak typ.
≥ +6 dBm (1560 -1610 nm)
≥ +4 dBm (1530 -1610 nm)
≥ +2 dBm (1520 -1620 nm)
≥ - 3 dBm (1510 -1640 nm)
reduced by 1.5dB
Minimum output power
[3]
with option #003
-3 dBm
-4.5 dBm (60-dBm in attenuation mode)
Power stability
[8]
±0.01 dB, 1 hour, typ. ±0.03 dB, 24 hours
Power repeatability (typ.) ±0.01 dB
Power linearity
[3]
±0.3 dB
Power flatness versus wavelength
[3]
±0.3 dB, typ. ±0.15 dB
Side-mode suppression ratio (typ.)
[4] [7]
≥ 40 dB (1530 – 1610 nm)
Signal to source spontaneous emission ratio
[5] [7]
≥45 dB/nm (1530 –1620 nm)
≥40 dB/nm (1520 – 1620 nm)
≥35 dB/nm (1510 –1640 nm)
Signal to total source spontaneous emission ratio (typ.)
[6] [7]
≥ 27 dB
(typ.,1530 – 1610 nm)
Relative intensity noise (RIN, typ.)
[7]
-145 dB/Hz (1530 – 1610 nm)
[1]
Valid for one month and within a ±5 K temperature range after automatic wavelength zeroing.
[2]
At CW operation. Measured with wavelength meter based on wavelength in vacuum.
[3]
Option#003: built-in optical attenuator.
[4]
Measured by heterodyning method.
[5]
Measured with optical spectrum analyzer at 1 nm resolution bandwidth.
[6]
Measured with optical spectrum analyzer.
[7]
Output power as specified per wavelength range.
[8]
Warm up time: 1 hour
Comentarios a estos manuales