Agilent Technologies 81642A Manual de usuario Pagina 34

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 40
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 33
www.agilent.com/comms/lightwave
Systems & Polarization Analysis
Agilent 86038B Photonic Dispersion and Loss Analyzer (preliminary)
Group Delay and Differential Group Delay Measurement
Relative Group Delay Repeatability
<0 dB to –10 dB level ± 20 fs
<–10 dB to –20 dB level (characteristic) ± 150 fs
<–20 dB to–30 dB level (characteristic) ± 500 fs
<–30 dB to –40 dB level (characteristic) ± 5 ps
Relative Group Delay Uncertainty (<0 dB to –10 dB level) ± 50 fs
Differential Group Delay Repeatability ± 50 fs
Differential Group Delay Uncertainty (<0 dB to –10 dB level) ± 90 fs
PMD Uncertainty ± 0.03 ps + 7%
2nd-order uncertainty (typ) PCD based on DGD uncertainty
Group Delay Time Resolution 1 fs
Modulation Frequency Range 5 MHz to 2.5 GHz
Group Delay Loss Range 50 dB
Measurements performed at the same temperature as the normalization temperature ± 0.5C. Performance measured using a 2.2 meter thermally isolated SMF patch cord.
Modulation frequency = 2 GHz. IFBW = 70 Hz, 1 nm wavelength step size. Repeatability is defined as the worst (plus or minus) standard deviation over the TLS wavelength range from 10 sweeps.
Length Measurement
Length Uncertainty (typ) ± 0.2mm +5x10-6L for L<56km.
Amplitude Measurement
Polarization Dependent Loss Accuracy ± 0.15typ ( 0.03 dB with Option #400)
System dynamic range (characteristic) 50 dB
Gain Loss Uncertainty ± 0.1 dB typ. ( 0.02 dB with Option #400)
Wavelength Measurement
Wavelength Range
with Agilent 81600B-200 Tunable Laser Source 1440 nm to 1640 nm
with Agilent 81600B-160 or 81640B 1495 nm to 1640 nm
with Agilent 81600B-150 1450 nm to 1590 nm
with Agilent 81600B-140 1370 nm to 1495 nm
with Agilent 81600B-130 1260 nm to 1375 nm
with Agilent 81640A 1510 nm to 1640 nm
Minimum Wavelength Step Size 0.1 pm
Absolute Wavelength Accuracy
a,b
Stepped mode with Agilent 86122A (typ.) ± 1 pm
Swept mode without Agilent 86122A and with 81600B or 81640B <5pm
Stepped mode without Agilent 86122A and with 81640A ± 15 pm
Relative Wavelength Accuracy
a,c
Stepped mode without Agilent 86122A and with 81600B or 81640B ± 5 pm
Stepped mode without Agilent 86122A and with 81600B or 81640B (characteristic) ± 2 pm
Stepped mode without Agilent 86122A and with 81640A ± 7 pm
Stepped mode with Agilent 86122A and with 81640A (characteristic) ± 3 pm
a
Valid for one month and within a ±4.4 K temperature range after automatic wavelength zeroing. Measured with wavelength meter based on wavelength in vacuum.
b
For details, refer to tunable laser's absolute wavelength accuracy specification.
c
For details, refer to tunable laser's relative wavelength accuracy specification.
Optical Fiber Chromatic Dispersion Measurement
CD accuracy 0.1ps/nm +0.3%CD
Zero Dispersion Wavelength Accuracy (characteristic)
a
± 150 pm
Zero Dispersion Wavelength Repeatability (characteristic)
a
±9 pm
Accuracy of dispersion slope at the zero dispersion wavelength (characteristic)
a
± 25 fs/nm2
Repeatability of dispersion slope at the zero dispersion wavelength (characteristic)
a
± 3 fs/nm2
a
Derived from GD specification.
General Information
Assembled Dimensions: (H x W x D) 55.5cm x 43.5cm x 55.5cm
Net Weight Standard system: 54 kg
34
Vista de pagina 33
1 2 ... 29 30 31 32 33 34 35 36 37 38 39 40

Comentarios a estos manuales

Sin comentarios