Chapter 1 Specifications
Supplemental Characteristics
18
Output Voltage Overshoot
During turn-on or turn-off of ac power, output plus overshoot will not exceed 1 V if
the output control is set to less than 1 V. If the output control is set to
1 V or higher, there is no overshoot.
Programming Language
SCPI (Standard Commands for Programmable Instruments)
State Storage Memory
Three (3) user-configurable stored states
Recommended Calibration Interval
1 year
Output Terminal Isolation (maximum, from chassis ground)
±60 Vdc when connecting shorting conductors without insulation to the
(+) output to the (+) sense and the (-) output and the (-) sense terminals.
±240 Vdc when connecting insulated shorting conductors to the (+) output
to the (+) sense and the (-) output and the (-) sense terminals.
AC Input Ratings (selectable via rear panel selector)
std 115 Vac ± 10%, 47 to 63 Hz
opt 0E3 230 Vac ± 10%, 47 to 63 Hz
opt 0E9 100 Vac ± 10%, 47 to 63 Hz
Maximum Input Power
700 VA with full load
Cooling
Fan cooled
Operating Temperature
0 to 40 °C for full rated output. At higher temperatures, the output current is derated
linearly to 50% at 55 °C maximum temperature.