2-16
8753ET Specifications and Characteristics
Instrument Specifications
Table 2-14 Test Port Input
8753ET Test Port Input
Dynamic Accuracy (Characteristic)
For the reflection port, accuracy of the test port input power reading relative to the reference
input power level.
Input: reflection port
For test port powers > 50 dBm and < 0 dBm, magnitude dynamic accuracy is
0.02 dB + 0.001 dB/dB from the reference power, phase dynamic accuracy is
0.132 deg + 0.0066 deg/dB from the reference power.
For frequencies < 2 MHz, the accuracy is degraded 0.1 dB above 5 dBm test port power.