3
System Configuration
The basic 8510C measurement system configura-
tion needed for test port flatness correction is
shown in Figure 1. A single channel power meter
such as the Agilent 437B, E4418A, or E4418B is
required, along with a compatible sensor, for per-
forming the power flatness calibration. Power
meters with the 437B command set must be used
for the power level flatness correction.
Figure 1. Basic Agilent 85107B measurement system con-
figuration for test port flatness correction
The Agilent 8511B frequency converter test
set may be used with the test port power flatness-
correction feature. The range of the test port
power is determined by the available power of
the 8360 source and the insertion loss of the com-
ponents added to the measurement system.
8517B test set
8485A
power sensor
E4418B
power meter
83651B
synthesized sweeper
8510 system bus
85107B Measurement System
8510C network analyzer