Agilent Technologies U2000 Series Manual de servicio Pagina 95

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 99
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 94
Service 4
U2000 Series Operating and Service Guide 79
Attenuator Disassembly and Reassembly Procedure for
U2000B and U2001B
Disassembly Procedure
Tools required for disassembly:
Table 4-12 Attenuator Disassembly Procedure
Tools Purpose Torque value
¾” torque wrench To loosen the attenuator 80 lbs.in
½” wrench To prevent rotation N/A
NOTE
The attenuator for U2000B and U2001B must not be disassembled under any
circumstances except during annual calibration. Removing the attenuator for U2000B
and U2001B will void the calibration.
1 Loosen the connector using the
torque wrench.
2 After that clean the connector’s
threads with IPA. Ensure that
the crystallized loctite is
cleaned properly.
Vista de pagina 94
1 2 ... 90 91 92 93 94 95 96 97 98 99

Comentarios a estos manuales

Sin comentarios