13
Settling time
Manual filter, 10-dB decreasing power step (not across switching points)
1
0.045
0.042
2
0.09
0.05
4
0.17
0.09
8
0.34
0.17
16
0.66
0.34
32
1.3
0.66
64
2.6
1.3
128
5.2
2.6
256
10.4
5.2
1024
41.9
20.9
No. of averages
Settling time (s)
Normal mode
x2 mode
512
20.9
10.4
Auto filter, default resolution, 10-dB decreasing power step
In Fast mode (using free run trigger), at 10-dB decreasing power step
Settling time = 25 ms*
* When a power step crosses through the sensor’s auto-range switching point, add 25 ms.
Specifications (continued)
x2
mode
Normal
mode
Sensor
dynamic
range
Typical
settling
times
Min dBm
Low Power Path
High Power Path
Max dBm
45 ms
82 ms
1.3 s
1.5 s
450 ms
1.6 s
20 s
24 s
24 s
45 ms
90 ms
2.6 s
2.7 s
460 ms
2.8 s
39 s
42 s
42 s
+10 dBm
+2 dBm
–4 dBm
–10 dBm
–20 dBm
–30 dBm
–40 dBm
–50 dBm
+20 dBm
+12 dBm
+6 dBm
0 dBm
–10 dBm
–20 dBm
–30 dBm
–40 dBm
+40 dBm
+32 dBm
+26 dBm
+20 dBm
+10 dBm
0 dBm
–10 dBm
–20 dBm
U2000/1/2/4A U2000/1/2H U2000/1B